Abstract
Purpose: The purpose of this study was to compare the stress distribution characteristics of four different abutment connections on SS-$III^{(R)}$ fixture under occlusal loading, using 3-dimensional finite element method. Materials and methods: The fixture of SS-$III^{(R)}$ (Osstem, Korea) with 4 mm diameter and 11.5 mm length and 4 types of abutments were analyzed; Solid, Com-Octa, ComOcta Gold, and Octa abutment. The models were placed in the area of first molar in the mandible. The 4 loading conditions were; (1) the vertical loading of 100 N on the central fossa, (2) the vertical loading of 100 N on the buccal cusp, (3) the $30^{\circ}$ inclined loading of 100 N to lingual side on the central fossa, and (4) the $30^{\circ}$ inclined loading of 100 N to the lingual side on the buccal cusp. The 3G.Author program was used, the von-Mises stress was calculated and the stress contours were plotted on each part of the implant systems and the surrounding bone structures. Results: Regardless of abutment types and loading conditions, higher stress concentration was observed at the cortical bone. In cancellous bone, the highest stress was observed at apical portion and the maximum stress occurred at the implant neck. The higher internal stress was observed in the fixtures than in the bone. The lowest stress was observed at loading condition 1 and the stress concentration was also lower than any other loading conditions. Conclusion: Within the limitation of the result of this study, it seems that the abutment connection type does not affect much on the stress distribution of bone structure.
Keywords
Three-dimensional FEA; SS-$III^{(R)}$ implant; Implant abutment; Stress distribution;
Citations & Related Records
Times Cited By KSCI : 5  (Citation Analysis)
연도 인용수 순위
1 Khraisat A, Stegaroiu R, Nomura S, Miyakawa O. Fatigue resistance of two implant/abutment joint designs. J Prosthet Dent 2002;88:604-10.   DOI   ScienceOn
2 Branemark PI, Zarb GA, Albrektsson T, Rosen H. Tissue-integrated prostheses. Osseointegration in clinical dentistry. Plast Reconstr Surg 1986;77:496-7.   DOI
3 Han SU, Park HO, Yang HS. Stress analysis of supporting tissues and implants according to implant fixture shapes and implant-abutment connections. J Korean Acad Prosthodont 2004;42:226-37.   과학기술학회마을
4 Andersson B, O¨dman P, Boss A, Jo ¨rne′us L. Mechanical testing of superstructures on the CeraOne abutment in the Branemark system. Int J Oral Maxillofac Implants 1994;9:665-72.
5 Kwon JH, Choi MH, Kim YL, Cho HW. Three-dimensional finite element stress analysis of single implant restoration using different fixture and abutment screw diameters. J Korean Acad Prosthodont 2005;43:105-19.   과학기술학회마을
6 Sutter F, Weber HP, Sorensen J, Belser U. The new restorative concept of the ITI Dental Implant System: design and engineering. Int J Periodontics Restorative Dent 1993;13:409-31.
7 Park WH, Lee YS. Three dimensional finite element stress analysis of implant prosthesis according to the different fixture locations and angulations. J Korean Acad Prosthodont 2005;43:61-77.   과학기술학회마을
8 Matsushita Y, Kitoh M, Mizuta K, Ikeda H, Suetsugu T. Two dimensional FEM analysis of hydroxyapatite implants: diameter effects on stress distribution. J Oral Implantol 1990;16:6-11.
9 Lum LB, Osier JF. Load transfer from endosteal implants to supporting bone: an analysis using statics. Part one: Horizontal loading. J Oral Implantol 1992;18:343-8.
10 Chung KM, Chung CH, Jeong SM. Finite element stress analysis of implant prosthesis according to platform width of fixture. J Korean Acad Prosthodont 2003;41:674-88.   과학기술학회마을
11 Sutter F, Schroeder A, Buser DA. The new concept of ITI hollowcylinder and hollow-screw implants: Part 1. Engineering and design. Int J Oral Maxillofac Implants 1988;3:161-72.
12 Moon BH, Yang JH. A study on the stress analysis of three root form implants with finite element analysis. J Korean Acad Prosthodont 1993;31:129-50.
13 Richter EJ. In vivo vertical forces on implants. Int J Oral Maxillofac Implants 1995;10:99-108.
14 Rangert BR, Sullivan RM, Jemt TM. Load factor control for implants in the posterior partially edentulous segment. Int J Oral Maxillofac Implants 1997;12:360-70.
15 Norton MR. An in vitro evaluation of the strength of a 1-piece and 2-piece conical abutment joint in implant design. Clin Oral Implants Res 2000;11:458-64.   DOI   ScienceOn
16 Ahn JK, Kay KS, Chung CH. Finite element stress analysis of implant prosthesis with internal connection between the implant and the abutment. J Korean Acad Prosthodont 2004;42:356-72.
17 Rangert B, Krogh PH, Langer B, Van Roekel N. Bending overload and implant fracture: A retrospective clinical analysis. Int J Oral Maxillofac Implants 1995;10:326-34.
18 Norton MR. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997;8:290-8.   DOI   ScienceOn
19 Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81:159-66.   DOI   ScienceOn
20 Naert IE, Duyck JA, Hosny MM, van Steenberghe D. Freestanding and tooth-implant connected prostheses in the treatment of partially edentulous patients. Part 1: An up to 15-years clinical evaluation. Clin Oral Implants Res 2001;12:237-44.   DOI   ScienceOn
21 Lee JM, Kim YS, Kim CH, Kim YH. 3-D FEA of three different single tooth abutments : Cement-retained vs Screw-retained. J Korean Acad Prosthodont 1999;37:269-88.
22 Shin HS, Chun HJ, Han CH, Lee SH. Three-dimensional stress analysis of implant systems in the mandibular bone with various abutment types and loading conditions. J Korean Acad Prosthodont 2003;41:617-25.   과학기술학회마을