Browse > Article

EFFECTS OF THE ER,CR:YSGG LASER ON BONE BED PREPARATION WITH VARIOUS LASER TIPS  

Kim, Seong-Kyun (Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University)
Heo, Seong-Joo (Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University)
Koak, Jai-Young (Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University)
Hong, Seong-Doo (Department of Oral Pathology, School of Dentistry, Seoul National University)
Lee, Shin-Jae (Department of Orthodontics, School of Dentistry, Seoul National University)
Lee, Joo-Hee (Department of Prosthodontics, Asan Medical Center, College of Medicine, University of Ulsan)
Publication Information
The Journal of Korean Academy of Prosthodontics / v.46, no.3, 2008 , pp. 255-260 More about this Journal
Abstract
STATEMENT OF PROBLEM: Preparation of implant beds with lasers is considered a safe and reliable method, but the accuracy of this technique has not been examined. PURPOSE: The purpose of this study was to evaluate the accuracy and effectiveness of implant bed preparation using an Er,Cr:YSGG laser. MATERIAL AND METHODS: An Er,Cr:YSGG laser was applied to pig rib bone. The laser was employed at a 5.75 W power setting, 30 Hz/sec pulse repetition, and 70 ${\mu}s$ pulse duration with 50 % water and 60% air spray. According to laser tips the groups were divided as follows; Group 1: paralleled - shaped sapphire tip (0.6 mm${\Phi}$), Group 2: paralleled - shaped zirconia tip (0.6 mm${\Phi}$), Group 3: tapered sapphire tip (0.4 mm${\Phi}$). The Er,Cr:YSGG laser tip was separated by 1 mm from the bone and applied for 15 seconds in a non-contact mode. After the application, the bone was sectioned for specimens. Histologic measurements were determined by computerized morphometry. The length of the prepared bone surface was measured and the width of the entrance was measured. The results were analyzed with one-way ANOVA (P<0.05). RESULTS: The prepared length of group 3 was longer than that of group 2. The prepared bone width was larger than the width of the laser tip in every group. Additional bone removal was observed adjacent to the prepared area and displayed an irregular surface. CONCLUSION & DISCUSSION: Different cutting effects were observed according to the laser tip, emphasizing the importance of proper tip selection in the clinical setting. This preliminary study supported the existence of hydrokinetic effects.
Keywords
Er.Cr:YSGG laser; Laser tip; Bone cutting length; Bone cutting width; Shape of bone cutting;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Aoki A, Yoshino T, Akiyama F, Miura M, Kinoshita A, Oda S, Watanabe H, Ishikawa I. Comparative study of Er:YAG laser and rotating bur for bone ablation: SEM and longterm histological examinations. Lasers in Dentistry 2003;389-91
2 Riziou I, Kimmel A. Atomized fluid particles for electromagnertically induced cutting. US patent 5,741,247. 1998
3 Sasaki KM, Aoki A, Ichinose S, Ishikawa I. Ultrastructural analysis of bone tissue irradiated by Er:YAG laser. Lasers Surg Med 2002;31:322-32   DOI   ScienceOn
4 Rizoiu IM, DeShazer LG. New laser-matter interaction concept to enhance hard tissue cutting efficiency. Laser-Tissue Interaction 1994;Vol.2134A:309-17
5 Wang X, Zhang C, Matsumoto K. In vivo study of the healing processes that occur in the jaws of rabbits following perforation by an Er,Cr:YSGG laser. Lasers Med Sci 2005;20:21-7   DOI
6 Eriksson A, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury. A vital microscoping study in the rabbit. J Prosthet Dent 1983; 50:101-7   DOI   ScienceOn
7 Kimura Y, Yu DG, Fujita A, Yamashita A, Murakami Y, Matsumoto K. Effects of erbium, chromium:YSGG laser irradiation on canine mandibular bone. J Periodontol 2001;72:1178-82   DOI   ScienceOn
8 el-Montaser M, Devlin H, Dickinson MR, Sloan P, Lloyd RE. Osseointegration of titanium implant in erbium- YAG laser-prepared bone. Implant Dent 1999;8:79-85   DOI   PUBMED   ScienceOn
9 van As G. Erbium lasers in dentistry. Dent Clin N Am 2004;48:1017-59   DOI   ScienceOn
10 Eyrich G. Hard-tissue drilling and cutting with a 9.6 um $CO_2$ laser. Med Habililationsschrift, Zurich. 2004
11 Ivanenko M, Sader R, Afilal S, Werner M, Hartstock M, von Hanisch C, Milz S, Erhardt W, Zeilhofer H-F, Hering P. In vivo animal trials with a scanning $CO_2$ laser osteotome. Lasers Surg Med 2005;37:144-8   DOI   ScienceOn
12 Eversole LR, Rizoiu IM. Preliminary investigations on the utility of an erbium, chromium:YSGG laser. J Calif Dent Assoc 1995;23:41-7   PUBMED
13 Walsh LJ. The current status of laser applications in dentistry. Aust Dent J 2003;48:146-55   DOI   ScienceOn
14 Kesler G, Romanos G, Koren R. Use of Er:YAG laser to improve osseointegration of titanium alloy implant-a comparison of bone healing. Int J Oral Maxillofac Implants 2006;21:375-9
15 Schwarz F, Olivier W, Herten M, Sager M, Chaker A, Becker J. Influence of implant bed preparation using an Er:YAG laser on the osseointegration of titanium implants: a histomorphometrical study in dogs. J Oral Rehabil 2007;34:273-81   DOI   ScienceOn
16 Salina S, Maiorana C, Iezzi G, Colombo A, Fontana F, Piattelli A. Histological evaluation, in rabbit tibiae, of osseointegration of mini-implants in sites prepared with Er:YAG laser versus sites prepared with traditional burs. J Long Term Eff Med Implants 2006;16:145-56   DOI   PUBMED   ScienceOn
17 Wang X, Ishizaki NT, Suzuki N, Kimura Y, Matsumoto K. Morpholgoical changes of bovine mandibular bone irradiated by Er,Cr:YSGG laser: an in vitro study. J Clin Laser Med Surg 2002;20:245-50   DOI   ScienceOn
18 Hossain M, Nakamura Y, Yamada Y, Kimura Y, Matsumoto, N, Masumoto, K. Effects of Er,Cr:YSGG laser irradiation in human enamel and dentin: ablation and morphological studies. Journal of Clinical Laser Medicine and Surgery, 1999b;17:155   PUBMED
19 Aoki A, Sasaki KM, Watanabe H, Ishikawa I. Lasers in nonsurgical periodontal therapy. Periodontol 2000 2004;36:59-97   DOI   ScienceOn
20 Sasaki KM, Aoki A, Ichinose S, Yoshino T, Yamada S, Ishikawa I. Scanning electron microscopy and Fourier transformed infrared spectroscopy analysis of bone removal using Er:YAG and $CO_2$ lasers. J Periodontol 2002;73:643-52   DOI   ScienceOn