REMOVAL TORQUE OF BICORTICALLY STABILIZED RBM(RESORBABLE BLAST MEDIA) PIN IMPLANTS IN RABBIT TIBIA  

Kim, Kwon-Sik (Major in Dentistry, Department of Medical Science, Korea University)
Suh, Kyu-Won (Department of Prosthodontics, Korea University Medical Center, Korea University)
Lee, Richard Sung-Bok (Department of Prosthodontics, School of Dentistry, Kyung-Hee University)
Ryu, Jae-Jun (Department of Prosthodontics, Korea University Medical Center Korea University)
Publication Information
The Journal of Korean Academy of Prosthodontics / v.44, no.6, 2006 , pp. 722-733 More about this Journal
Abstract
Statement of problem. The use of small diameter implants having less than 3 mm in diameter were restricted because of lack of bonding strength to bone. Purpose. The purpose of this study was to observe how much resorbable blast media pin implants increase the binding force to the bone compared to machined transitional pin implants by measuring removal torque, and whether they can be used as final implants for replacement of small diameter teeth. Material and method. Fifteen rabbits were used in this study. Two kinds of implants (resorbable blast media pin implants and machined transitional pin implants) were inserted in each tibia bicortically. After healing time of 2, 4 and 8 weeks, the removal torque values were recorded and the rabbits were sacrificed for histological analysis. Linear finite element method analyses were conducted to compare bicortical fixation with monocortical fixation. Result and conclusion. Within the limitation of this in vivo study, the following conclusions were drawn: 1) The removal torque value of RBM pin implants showed statistically significant increase compared to machined pin implants at 2, 4, and 8 weeks respectively (p<0.05). 2) The removal torque value of RBM pin implants at 2, 4, and 8 weeks was increased statistically significantly with time (p<0.05). 3) Bicortical fixation showed better stress distribution compared with monocortical fixation in a linear finite element method analysis. 4) RBM pin implants are not recommended as transitional implants because they showed a lot of bone fracture in histologic specimens.
Keywords
RBM(Resorbable blast media); Pin implant; Removal torque; Rabbit. Histological Analysis;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Schnitman PA, Wohrle PS, Rubenstein JE. Immediate fixed interim prostheses supported by two-stage threaded implants: Methodology and results. J Oral Implantol 1990;16:96-105
2 Sagara M, Akagawa Y, Nikai H, Tsuru H. The effects of early occlusal loading on one-stage titanium alloy implants in beagle dogs: A pilot study. J Prosthet Dent 1993;69:281-288   DOI   ScienceOn
3 Lee SB. Immediate temporary prosthesis and loading by using transitional pin implant. J Kyung Hee Medical. Vol.16 No2. 2000:126-135
4 Petrungaro PS. Fixed temporization and boneaugmented ridge stabilization with transitional implants. Pract Periodontics Aesthet Dent 1997;9:1071- 1078; quiz 1080
5 Wojcik MS, Pokorny PH. Use of a one-stage narrow- diameter implant to replace incisors. J Mich Dent Assoc 2006;88:42, 44-9
6 Ivanoff CJ, Sennerby L, Lekholm U. Influence of mono- and bicortical anchorage on theintegration of titanium implants. A study in the rabbit tibia. Int J Oral Maxillofac Surg 1996;25:229-235   DOI   ScienceOn
7 Pierrisnard L, Renouard F, Renault P, Barquins M. Influence of implant length and bicortical anchorage on implant stress distribution. Clin Implant Dent Relat Res 2003;5:254-62   DOI   ScienceOn
8 Gapski R, Wang HL, Mascarenhas P, Lang NP. Critical review of immediate implant loading. Clin Oral Implants Res 2003;14:515-527   DOI   ScienceOn
9 Comfort MB, Chu FC, Chai J, Wat PY, Chow TW. A 5-yearprospective study on small diameter screw-shaped oral implants. J Oral Rehabil 2005 ;32:341-5   DOI   ScienceOn
10 Hallman M. A prospective study of treatment of severely resorbed maxillae with narrow nonsubmerged implants: results after 1 year of loading. Int J Oral Maxillofac Implants 2001;16:731-6
11 Laskin DM. Nonsurgical management of bilateral mandibular fractures associated with dental implants: report of a case. Int J Oral Maxillofac Implants 2003;18:739-44
12 Albrektsson T, Dahl E, Enbom L, Engevall S, Engquist B, Eriksson AR, Feldmann G, Freiberg N, Glantz PO, Kjellman O. and et, al. Osseointegrated oral implants. A Swedish multicenter study of 8139 consecutively inserted Nobelpharma implants. J Periodontol 1988; 59:287-96   DOI
13 Li D, Ferguson SJ, Beutler T, Cochran DL, Sittig C, Hirt HP, Buser D. Biomechanical comparison of the sandblasted and acid-etched and the machined and acid-etched titanium surface for dental implants. J Biomed Mater Res 2002;60:325-32   DOI   ScienceOn
14 Nazarian A. Mini dental implants: immediate gratification for patient and provider. Dent Today. 2005 Oct;24(10):110, 112
15 Cho SA, Park KT. The removal torque of titanium screw inserted in rabbit tibia treated by dual acid etching. Biomaterials 2003;24:3611-7   DOI   ScienceOn
16 Zinsli B, Sagesser T, Mericske E, Mericske-Stern R. Clinical evaluation of small-diameter ITI implants: a prospective study. Int J Oral Maxillofac Implants. 2004 Jan-Feb;19(1):92-9
17 Friberg B, Sennerby L, Linden B, Grondahl K, Lekholm U. Stability measurements of one-stage Branemark implants during healing in mandibles. A clinical resonance frequency analysis study. Int J Oral Maxillofac Surg 1999;28:266-72   DOI   ScienceOn
18 Tjellstrom, Jacobsson M, Albrektsson T. A Removal torque of osseointegrated craniofacial implants: A clinical study. Int J Oral Maxillofac Implants 1988;3:287-289
19 Van Oosterwyck H, Duyck J, Vander Sloten J, Van der Perre G, De Cooman M, Lievens S, Puers R, Naert I. The influence of bone mechanical properties and implant fixation upon bone loading around oral implants. Clin Oral Implants Res 1998;9:407-418   DOI   ScienceOn
20 Davarpanah M, Martinez H, Tecucianu JF, Celletti R, Lazzara R. Small-diameter implants: indications and contraindications. J Esthet Dent 2000;12:186-94   DOI
21 Carlsson L, Rostlund T, Albrektsson B, Albrektsson T. Removal torques for polished and rough titanium implants. Int J Oral Maxillofac Implants 1988;3:21- 24
22 Piattelli M, Scarano A, Paolantonio M, Iezzi G, Petrone G, Piattelli A. Bone response to machined and resorbable blast material titanium implants: An experimental study in rabbits. J Oral Implantol 2002;28:2-8   DOI   ScienceOn
23 Goodacre CJ, Kan JY, Rungcharassaeng K. Clinical complications of osseointegrated implants. J Prosthet Dent 1999;81:537-52   DOI   ScienceOn
24 Froum S, Emtiaz S, Bloom MJ, Scolnick J, Tarnow DP. The use of transitional implants for immediate fixed temporary prostheses in case of implant restoration. Pract Periodontics Aesthet Dent 1998;10:737-746; quiz 748
25 Sennerby L, Dasmah A, Larsson B, Iverhed M. Bone tissue responses to surface-modified zirconia implants: A histomorphometric and removal torque study in the rabbit. Clin Implant Dent Relat Res 2005;7 Suppl 1:S13-20   DOI
26 Barber HD, Seckinger RJ. The role of the small-diameter dental implant: a preliminary report on the Miniplant system. Compendium 1994;15:1390, 1392
27 Adell R. Tissue integrated prostheses in clinical dentistry. Int Dent J 1985;35:259-65
28 Simon H, Caputo AA. Removal torque of immediately loaded transitional endosseous implants in human subjects. Int J Oral Maxillofac Implants 2002;17:839-845
29 Mazor Z, Steigmann M, Leshem R, Peleg M. Miniimplants to reconstruct missing teeth in severe ridge deficiency and small interdental space: a 5-year case series. Implant Dent. 2004 Dec;13(4):336-41   DOI
30 Branemark P, Hansson B, Adell R, Breine U, Linstrom J, Hallen O et al. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10 years period. Scand J Plast Reconstr Surg 1977;16:1-132
31 Murata T, Yamashita Y, Kurokawa H, Takahashi T. Dental rehabilitation using an implant-supported overdenture after repair of a fracture in a severely resorbed edentulous mandible: a case report. Int J Oral Maxillofac Implants 2004;19:749-52
32 Gonshor A, Goveia G, Sotirakis E. A prospective, multicenter, 4-year study of the ACE surgical resorbable blast media implant. J Oral Implantol 2003;29:174-180   DOI   ScienceOn
33 Johansson C, Albrektsson T. Integration of screw implants in the rabbit: A 1-year follow-up of removal torque of titanium implants. Int J Oral Maxillofac Implants 1987;2:69-75
34 Sanz A, Oyarzun A, Farias D, Diaz I. Experimental study of bone response to a new surface treatment of endosseous titanium implants. Implant Dent 2001;10:126-31   DOI   ScienceOn
35 Cehreli MC, Akca K. Narrow-diameter implants as terminal support for occlusal three-unit FPDs: a biomechanical analysis. Int J Periodontics Restorative Dent 2004;24:513-9
36 English C, Bahat O, Langer B, Sheets CG. What are the clinical limitations of wide-diameter (4 mm or greater) root-form endosseous implants- Int J Oral Maxillofac Implants 2000;15:293-6
37 Wennerberg A, Albrektsson T, Andersson B, Krol JJ. A histomorphometric and removal torque study of screw-shaped titanium implants with three different surface topographies. Clin Oral Implants Res 1995;6:24-30   DOI   ScienceOn
38 Davies JE. Mechanisms of endosseous integration. Int J Prosthodont 1998;11:391-401