Browse > Article

A Study on the Development of a Human Middle Ear Implant  

Yoo, Seung-Hyun (아주대학교 기계공학부)
Kim, Hak-Kyun (한국타이어 중앙연구소)
Kim, Jong-Bum (GM대우 기술연구소)
Song, Joon-Ho (자트코코리아 엔지니어링 기술연구소)
Oh, Dae-Jin (아주대학교 대학원 기계공학과)
Abstract
This paper describes the design of the ossicular replacement prosthesis for the people who have problem of hearing because of middle ear disease. Also the simulation results of the designed product for verification by finite element analysis software (ABAQUS) are presented. New model is applied to middle ear FE analysis which was generated in previous study. The full replacements of ossicular prosthesis for ossicles (malleus, incus and stapes) are made of Hydroxyapatite and Titanium. Although other existing prosthesis models consider only sound amplification effect, current type has damper system which is operating on the audible frequencies. High frequency sound transmitted to inner ear can be reduced and the prosthesis and inner ear can be prevented from damage.
Keywords
Middle Ear; Finite Element Method; Ossicular Replacement Prosthesis; TRIZ;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Zwislocki, J., "Analysis of Some Auditory Characteristics. Handbook of Mathematical Psychology", edited by Luce, R.D., Bush, R.R. and Galanter, E., Wiley, New York, 1965
2 Goldfire Manual, Invention Machine Inc., 2005
3 Huber, A., Ball, G, Asai, M. and Goode, R., "The Vibration Pattern of the Tympanic Membrane After Placement of a Total Ossicular Replacement Prosthesis", Proc. International Workshop on Middle Ear Mechanics in Research and Otosurgery (Dresden, Germany), pp. 219-222, 1997
4 Yoo, S. H., Park, K. H., Moon, S. K., Kim, W. S. and Bae, J. H., "Evaluation of Dynamic Behavior of the Human Middle Ear with Nonhomogeneity by Finite Element Method", Key Engineering Materials, Vol. 270-273, pp. 2067-2072, 2004   DOI
5 신문균 외, 인체 해부학, 현문사, 1998
6 Bathe, K. J., Finite Element Procedures, Prentice Hall, 1996
7 Koike, T. and Wada, H., "Modeling of the Human Middle Ear using the Finite Element Method", J Acoust. Soc. Am., Vol. 111, pp. 1306-1317, 2002   DOI   ScienceOn
8 Gyo, K., Aritomo, H. and Goode, R. J., "Measurement of the Ossicular Vibration Ratio in Human Temporal Bones by Use of a Video Measuring System", Acta. Otolaryngology, Vol. 103, pp. 87-95, 1987   DOI   ScienceOn
9 Terninko, J., Zusman, A. and Zlotin, B., "체계적인 이노베이션", 유승현 외 역, 한국 트리즈 협회, 2003
10 Ferris, P. and Prendergast, P. J., "Middle-ear Dynamics Before and After Ossicular Replacement", J Biomechanics Vol. 33, pp. 581-590, 2000   DOI   ScienceOn
11 Funnell, W. R. 1. and Laszlo, C. A., "Modeling of the Cat Eardrum as a Thin Shell Using the Finite Element Method". J Acoust. Soc. Am., Vol. 63, pp. 1461-1467, 1978   DOI   ScienceOn
12 ABAQUS User's Manual, Hibbit, Karsson & Sorensen Inc., 2004
13 Taylor, W. R., Roland, E., Ploeg, H., Hertig, D., Klabunde, R., Warner, M. D., Hobatho, M. C., Rakotomanana, L. and Clift, S. E., "Determination of Orthotropic Bone Elastic Constants Using FEA and Modal analysis", J Biomechanics, Vol. 35, pp. 767-773, 2002   DOI   ScienceOn