Browse > Article
http://dx.doi.org/10.5392/JKCA.2017.17.09.069

Evaluate the Activation of Linear Accelerator Components and Shielding Wall through Simulation  

Lee, Dong-Yeon (동남권원자력의학원 방사선종양학과)
Park, Eun-Tae (인제대학교 부산백병원 방사선종양학과)
Kim, Jung-Hoon (부산가톨릭대학교 보건과학대학 방사선학과)
Publication Information
Abstract
This study evaluated the activation of the shielding wall and the components around the accelerator by using the medical linear accelerator. We performed simulations for energy values of 20 MV with the operating time ranging from day 1 to 30 years, and linear accelerator head and shielding wall concrete were also evaluated. The results showed that neutrons in large quantities were analyzed using high energy around thetarget point where photons were formed. Based on the activation analysis with these results, radioactivity increased with an increase in operation time and activated nuclides usually start saturating in10 years. Furthermore, the general types of nuclides formed owingto the activation were Co-60, W-181, 185, 187, Na-24, Ca-45, Mn-54, 56, and Fe-55, 59.
Keywords
Linear Accelerator; Activation; Radioactivity; MCNPX; Simulation;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 R. A. Forrest, "FISPACT-2007: User manual," UKAEA FUS, Vol.534, 2007.
2 International atomic Energy Agency, Classification of Radioactive Waste, General Safety Guide No. GSG-1, 2009.
3 A. Mesbahi, M. Fix, M. Allahverdi, E. Grein, and H. Garaati, "Monte Carlo calculation of Varian 2300C/D Linac photon beam Characteristics: a comparison between MCNP4C, GEANT3 and measurements," Applied Radiation and Isotopes, Vol.62, No.3, pp.469-477, 2005.   DOI
4 A. Baumgartner, A. Steurer, and F. Maringer, "Simulation of photon energy spectra from Varian 2100C and 2300C/D Linacs: Simplified estimates with PENELOPE Monte Carlo models," Applied Radiation and Isotope, Vol.67, No.11, pp.2007-2012, 2009.   DOI
5 W. L. Huang, Q. F. Li, and Y. Z. Lin, "Calculation of photoneutrons produced in the targets of electron linear accelerators for radiography and radiotherapy applications," Nuclear Instruments and Methods in Physics Research, Vol.B, No.229, pp.339-347, 2005.
6 이동연, 박은태, "몬테칼로법을 이용한 의료용 선형가속기 차폐벽의 방사화 특성 분석," 한국콘텐츠학회논문지, 제16권, 제10호, pp.758-765, 2016.   DOI
7 Helmut W. Fischer, Ben E. Tabot, and Bjorn Poppe, "Activation processes in a medical linear accelerator and spatial distribution of activation products," Phys. Med. Biol. Vol.51, pp.N461- N466, 2006.   DOI
8 배성철, 김준호, 이철수, "선형가속기 10 MV 광자선에서 산란판(Beam Spoiler) 사용 시 표면선량 변화," 대한방사선치료학회지, 제18권, 제1호, pp.21-28, 2006.
9 한국방사선진흥협회, 2015년도 방사선 및 방사성 동위원소 이용실태 조사, 미래창조과학부, 2017
10 이정옥, 정동혁, 강상수, "24 MV 의료용 선형가속기의 중성자 발생에 관한 연구," 의학물리학회, 제16권, 제2호, pp.97-103, 2005.
11 National Council on Radiation Protection and Measurement, Neutron Contamination from Medical Electron Accelerator, NCRP Report No.79, 1984.
12 박은태, "MCNPX를 이용한 방사선 치료실의 광중성자 선량평가," 한국콘텐츠학회논문지, 제15권, 제6호, pp.283-289, 2015.   DOI
13 O. Chibani and M. A. CMC, "Photonuclear dose calculations for high-energy photon beams from Simens and Varian linacs," Med. Phys., Vol.30, No.8, pp.1990-2000, 2003.   DOI
14 H. W. Fischer, B. E. Tabot, and B. Poppe, "Activation processes in a medical linear accelerator and spatial distribution of activation products," Phys. Med. Biol., Vol.51, pp.N461-N466, 2006.   DOI
15 B. Julian, Simulation of neutron production at a medical linear accelerator, Institute of Experimental Physics University of Hamburg, 2007.
16 J. A. Rawlinson, M. K. Islam, and D. M. Galbraith, "Dose to radiation therapists from activation at high-energy accelerators used for conventional and intensity-modulated radiation therapy," Med. Phys., Vol.29, No.4, pp.598-608, 2002.   DOI
17 Y. Z. Wang, M. D. C. Evans, and E. B. Podgorsak, "Characteristics of induced activity from medical accelerators," Med. Phys., Vol.32, No.9, pp.2899-2910, 2005.   DOI
18 International atomic Energy Agency, Radiation Protection in the Design of Radoitherapy Facilities, Safety Report Series No.47, 2006.
19 Korea Atomic Energy Research Institute, Shielding Technology for High Energy Radiation Production Facility, 2004.