Browse > Article
http://dx.doi.org/10.5392/JKCA.2015.15.06.297

A Study on the Thyroid Dose High-Energy Radiation Therapy of Lung Cancer  

Yang, Oh-Nam (목포과학대학교 방사선과/한서대학교 보건의료학과)
Lim, Cheong-Hwan (한서대학교 보건의료학과)
Publication Information
Abstract
High-energy medical linear accelerator on the dose to the thyroid cancer during radiotherapy were evaluated using optical stimulation luminescence dosimeters(OSLD) using. Scattered's influence in the case of 3D-CRT 25.4 mSv, 28.8 mSv, 31.3 mSv, 26.5 mSv, 27.4 mSv 5 times with an average 27.9 mSv, in the IMRT 46.8 mSv, 43.2 mSv, 42.3 mSv, 41.5 mSv, 44.1 mSv to five times the average of 43.6 was the result of mSv. In the case of light neutron dosimetry results 3D-CRT 3 mSv, 3 mSv, 3.4 mSv, 3.5 mSv, 3.1 mSv to five times the average 3.2 mSv, in the IMRT 5.1 mSv, 4.8 mSv, 4.2 mSv, 4.8 mSv, 4.9 mSv, to five times the average of 4.7 was the result of mSv. Both parties and the light scattered neutrons were significantly appreciated compared to IMRT 3D-CRT. Treatment of cancer using radiation workers, as in this study, and that a significant amount of scattered rays in the adjacent normal tissues during radiation therapy using energy assessment to influence by fully aware of this information is necessary for the exposure reduction efforts the feed.
Keywords
Lung Cancer; High-energy; Thyroid; Linear Accelerator; Photoneutrons;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 대한방사선치료학회 장비통계, 2011.
2 통계청, 암 통계, 2008.
3 장남준, 석진용, 원희수 등, "두경부암 세기변조 방사선치료 계획 시 부분적 고에너지 광자선 사용에 따른 치료계획 평가", 대한방사선치료학회지, 제21권, 제1호, pp.1-8, 2013.
4 H. Ing, W. R. Nelson, and R. A. Shore, "Unwanted photon and neutron radiation resulting from collimated photon beams interacting with the body of rad iotherapy patients," Med. Phys. Vol.9, pp.27-33, 1982.   DOI   ScienceOn
5 정제호, 김홍석, 박윤환, "의료용 선형가속기에서 생성되는 광중성자에 의한 선량 평가", 한국방사성폐기물학회지, 제2004권, 제1호, pp.296-296, 2004.
6 R. Nath, E. R. Epp, J. S. Langhlin, W. P. Swanson, and V. P. Bond, "Neutrons from high-energy X-ray medical accelerators. Anestimate of risk to the radiotherapy patient," Med. Phys., Vol.11, pp.231-241, 1984.   DOI   ScienceOn
7 박병석, 안종호, 권동열, "10 MV 이상 고에너지 치료 시 발생되는 광중성자의 영향", 대한방사선치료학회지, 제25권, 제1호, pp.9-14, 2013.
8 이화중, 김대영, 김영택, "10MV 이상 고에너지 사용시 wedge filler의 방사화가 작업환경에 미치는 영향평가", 대한방사선치료학회지, 제16권, 제2호, pp.69-79, 2004.
9 Omar Chibani, and Chang-Ming Charlie Ma., "Photonuclear does calcula ions for high energy photon beams from Simens and Varian lineacs," Medical Physics, Vol.30, No.8, pp.1990-2000, 2003.   DOI   ScienceOn
10 김대섭, 김유현, "방사선치료 장치 및 관련시설에서의 산란중성자에 관한 연구", 대한방사선치료학회지, 제17권, 제2호, pp.141-145, 2005.
11 ICRP(74), "1995 Conversion coefficients for use in radiological protection a gainst external radiation," International Commission on Radiological Protection, ICRP Publication 74.
12 G. Bauman and R. B. Rumble, "Intensityodulated Radiotherapy in the Treatment of Prostate Cancer," Clinical oncology : a journal of the Royal College of Radiologists, Vol.24, No.7, pp.461-473, 2012.   DOI
13 P. M. McCloskey, S. Atallah, and L. Coate, "Comparison of 3D Conformal Radiation Therapy (3D CRT) and Intensity Modulated Radiation Therapy (IMRT) in Stage III Non-small Cell Lung Cancer (NSCLC)," International journal of radiation oncology, biology, physics, Vol.84, No.3, suppl, pp.S597-S598, 2012.
14 S. Sharma, J. T. Whaley, and W. Zou, "Incidental Nodal Irradiation in Stage III Lung Cancer Treated With Involved-Field Radiation: Comparison Between 3D CRT and IMRT," International journal of radiation oncology, biology, physics, Vol.84, No.3, suppl, p.S564, 2012.
15 정제호, 의료용 선형가속기에서 생성되는 광중성자에 의한 선량평가, 한양대학교 대학원, 원자력공학과 학위논문집, pp.1-56, 2005.
16 M. H. Lee, "The Effect of Scattering Dose on the Thyroid During Mammography," Journal of the Korean institute of electronic material engineers, Vol.23, No.10, pp.826-830, 2010.   DOI   ScienceOn
17 이혜림, 김현영, 최형욱 이혜미, 임창선, 파노라마촬영 시 조직등가물질을 이용한 갑상선보호대의 갑상선 피폭선량 감소효과", Journal of the Korea Academia-Industrial cooperation Society, Vol.13, No.5, pp.2278-2284, 2012.   DOI   ScienceOn
18 ICRP(1991), "1990 Recommendations of the Internaitonal Commission on Radiological Protection," Internaional Commission on Radiological Protection, ICRP Publicaton 60.
19 ICRP(1996), "1995 Conversion coefficients for use in radiological protect ion a gainst external radiation," International Commission on Radiological Protection, ICRP Publication 74.