A Analysis of Effectiveness of Aluminium Filter in the added Compound Filtration by Detective Quantum Efficiency and Image Quality Evaluation  

Kim, Sang-Hyeon (서울대학교병원 영상의학과)
Kim, Yun-Min (삼성서울병원 영상의학과)
Kwon, Kyoung-Tae (동남보건대학교 방사선학과)
Ma, Sang-Chull (신한대학교 방사선학과)
Han, Dong-Gyoon (을지대학교 방사선학과)
Publication Information
Abstract
This study analysed the effectiveness of aluminium(Al) filter in the added compound filtration for the removal of characteristic radiation from high atomic number material by DQE and image evaluation. 1mm Al was applied to each 0.1, 0.2, 0.3 mm copper and befere and after use were evaluated. Beam quality and DQE were tested by IEC regulations and image quality was evaluated by PSNR, MAE, MSE, CNR, SNR and qualitative analysis was performed by 7 items for resolution and contrast from chest x-ray criteria of national cancer checkup. MTF 10 and 50% were the same by 4.6, 2.54 cycle/mm and NPS, DQE, PSNR MAE, MSE, CNR, SNR and qualitative analysis were all the same or slightly better when Al was not used. PSNR is over 30dB and all significant and at the qualitative analysis, the p-value of t-test was over 0.05. The DQE and image quality evaluation have little difference between before and after use of Al filter and it is effective to use the Al filter for the reduction of skin dose by removal of characteristic radiation.
Keywords
Aluminium Filter; DQE; Image Evaluation; Characteristic Radiation; Skin Dose;
Citations & Related Records
Times Cited By KSCI : 7  (Citation Analysis)
연도 인용수 순위
1 W. I. Cho, Y. K. Kim, and G. D. Lee, "Change of dose exposure and improvement of image quality by additional filtration in mammography", Journal of Radiation Protection, Vol.38, No.2, pp.78-90, 2013.   DOI   ScienceOn
2 G. S. Shin, J. H. Choi, Y. H. Kim, J. M. Kim, and C. G. Kim, "Patient dose in mammography", Journal of radiological science and technology, Vol.28, No.4, pp.293-299, 2005.
3 S. J. Moon, Y. G. Kim, and S. G. Lee, "Reduction of patient dose exposure and improvement of image quality by use of additional filtration in digital radiology", Korean J Digit Imaging Med, pp.19-25, 2010.
4 Richard R. Carlton, Arlene M, Adler Principles of Radiographic Imaging, 4th ed, pp.156-158, 2006.
5 S. S. Park and G. J. Kim, Introduction to the physics of Diagnostic Radiology, Dae Hak Seo Lim, Seoul, pp.135-141, 1985.
6 C. H. Lee and C. S. Lim, "A Study on Added Filters for Reduction of Radiation Exposure Dose in Skull A-P Projection", Journal of the Korea Academia-Industrial cooperation Society, Vol.12, No.7, pp.3117-3122, 2011.   DOI   ScienceOn
7 P. R. Granfors and R. Aufrichting, DQE(f) of an amorphous-silicon flat-panel x-ray detector: detector parameter influences and measurement methodology physics of medical imaging, SPIE, 3997, 2000.
8 G. J. Lee, M. K. Kim, J. W. Lee, and H. C, Kim, "Research for the Environmental optimization of Dose and Image quality in Digital Radiography", Journal of the Institute of Electronics Engineers of Korea, Vol.50, No.2, 2013.
9 Liu Bin, Wang Yuanyuan, and Wang Weiqi, "Spectrogram enhancement algorithm: a soft thresholding-based approach", Ultrasound in Medicine & Biology, Vol.25, No.5, pp.839-846, 1999.   DOI   ScienceOn
10 H. S. Kim, J. H. Jeong, and J. W. Lee, "Research on Image Quality and Effective dose by Exposure Index Variation", Journal of the Korean Society of Radiology, Vol.7, No.1, pp.63-69, 2013.   DOI   ScienceOn
11 Okka W. Hamer, "Contrast-Detail Phantom Study for X-ray Spectrum Optimization Regarding Chest Radiography Using a Cesium Iodide-Amorphous Silicon Flat-Panel Detector", Invest Radiol, Vol.39, pp.610-618, 2004.   DOI   ScienceOn
12 IEC 62220-1. Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: determination of the detective quantum efficiency, Geneva, 2003.
13 IEC 61267. Medical diagnostic X-ray equipment Radiation conditions for use in the determination of characteristics, Geneva, 1994.
14 H. Illers, E. Buhr, and C. Hoeschen, "Measurement of the detective quantum efficiency(DQE) of digital X-ray detectors according to the novel standard IEC 62220-1", Raiat. Prot. Dosimetry, Vol.114, pp.39-44, 2005.   DOI   ScienceOn
15 E. Samei and M. J. Flynn, "An experimental comparison of detector performance for direct and indirect digital radiography systems", Med Phy, Vol.30, No.4, pp.608-622, 2003.   DOI
16 S. G. Shin, "Decreased of Patient Dose by Built-in Filter in Pelvis A-P Projection", Journal of Korea Contents Association, Vol.12, No.8, pp.233-239, 2012.   DOI
17 S. Yang, J. B. Han, N. G. Choi, and S. G. Lee, "The Review of Exposure Index in Digital Radiography and Image Quality", Jounrnal of Radiation Protection, Vol.38, No.1, pp.29-36, 2013.   DOI   ScienceOn
18 M. Uffmann and S. Cornelia, "Digital Radiography: Balance between image quality and required radiation dose", Eur. J. Radil, Vol.72, pp.202-208, 2009.   DOI   ScienceOn