Browse > Article
http://dx.doi.org/10.5392/JKCA.2014.14.03.346

Development and Evaluation of the Reader for Blood Typing using Column Agglutination Method  

Kim, Jin (한림대학교 정보전자공과대학 컴퓨터공학과)
Jeon, Wooram (한림대학교 정보전자공과대학 컴퓨터공학과)
Park, Seoung Woo (강원대학교 의학전문대학원 신경외과)
Lee, Chang Youl (강원대학교 의학전문대학원 진단검사의학교실)
Lee, Da-Hyeon (강원대학교 의학전문대학원 진단검사의학교실)
Choi, In-Taek (강원대학교 의학전문대학원 진단검사의학교실)
Kim, Ju Yeon (강원대학교 의학전문대학원 진단검사의학교실)
Suh, In Bum (강원대학교 의학전문대학원 진단검사의학교실)
Publication Information
Abstract
Accurate blood typing tests are essential for safe blood transfusion. Recently many automated test equipments have been introduced to reduce errors and increase the efficiency of the test. However, those equipments being high in price, it is difficult to introduce automated test equipment for every hospital. In this study, we developed a reader for blood typing using column agglutination test. In the process, the results, read out by the image processing, are stored and reaffirmed. To evaluate the reader, 148 samples for ABO and RhD blood typing tests and 154 samples for unexpected antibody test were used. The positive and negative intensity of the reading and the reading of the reaction were 100% in agreement with the result of traditional manual method. If additional verification is completed, this reader can be efficiently and economically used in small-and medium-sized hospitals.
Keywords
Blood Type; Column Agglutination; Reader; Unexpected Antibody;
Citations & Related Records
연도 인용수 순위
  • Reference
1 M. U. Heim, M. Bock, H. J. Kolb, M. Schleuning, and W. Mempel, "Intravenous anti-D gammaglobulin for the prevention of rhesus isoimmunization caused by platelet transfusions in patients with malignant diseases," Vox Sang, Vol.62, No.3, pp.165-168, 1992.   DOI   ScienceOn
2 P. Gurevich, S. Erina, S. Gershon, and I. Zusman, "The role of the fetal immune system in the pathogenesis of RhD-hemolytic disease of newborns," Hum Antibodies, Vol.9, No.2, pp.76-89, 1997.
3 D. R. Anderson, J. Wiseman, J. MacLeod, E. Burton, and E. Zayed, "Evaluation of polyethylene terephthalate for ABO and Rh typing and alloantibody screening," Transfusion, Vol.40, No.6, pp.669-672, 2000.   DOI   ScienceOn
4 J. R. Storry, L. Castilho, G. Daniels, W. A. Flegel, G. Garratty, and C. L. Francis, "International Society of Blood Transfusion Working Party on red cell immunogenetics and blood group terminology: Berlin report," Vox Sang, Vol.101, No.1, pp.77-82, 2011.   DOI   ScienceOn
5 한규섭, 박명희, 김상인, 수혈의학 3판, 고려의학, 2006.
6 Y. Lapierre, D. Rigal, J. Adam, D. Josef, F. Meyer, and S. Greber, "The gel test: a new way to detect red cell antigen-antibody reactions," Transfusion, Vol.30, pp.109-113, 1990.   DOI   ScienceOn
7 W. Malomgre and B. Neumeister, "Recent and future trends in blood group typing," Anal Bioanal Chem, Vol.393, pp.1443-1451, 2009.   DOI   ScienceOn
8 K. J. Reis, R. Chachowski, A. Cupido, D. Davies, J. Jakway, and T. M. Setcavage, "Column agglutination technology: the antiglobulin test," Transfusion, Vol.33, No.8, pp.639-643, 1993.   DOI   ScienceOn
9 M. Bajpai, R. Kaur, and E. Gupta, "Automation in immunohematology," Asian J Transfus Sci, Vol.6, No.2, pp.140-144, 2012.   DOI   ScienceOn
10 H. Schoenfeld, K. Bulling, C. von Heymann, B. Neuner, U. Kalus, and H. Kiesewetter, "Evaluation of immunohematologic routine methods using the new erythrocyte-magnetized technology on the QWALYS 2 system," Transfusion, Vol.49, No.7, pp.1347-1352, 2009.   DOI   ScienceOn
11 G. R. Choi, B. K. Min, H. N. Kim, C. H. Cho, Y. H. Kim, and J. Y. Choi, "Evaluation of the automated instrument QWALYS-3 for unexpected antibody screening," Korean J Blood Transfus, Vol.22, pp.38-45, 2011.
12 Y. H. Ko, J. M. Kim, S. H. Koo, J. Lim, Y. C. Park, and K. C. Kwon, "Evaluation of the automated QWALYS-3 system for ABO and RhD grouping and unexpected antibody screening," Korean J Blood Transfus, Vol.22, pp.144-150, 2011.