Abstract
The purpose of this study is to evaluate exposure dose and absorbability according to differences in the content of iodine in contrast media used in interventional procedure and angiography. There were 7 levels of iodine content as 240, 270, 300, 320, 350, 370, and 400 I $mg/m{\ell}$. The higher iodine content was, the more elevated exposure dose and absorbability were. The output dose was about 8.5% from iodine content with 350I $mg/m{\ell}$ higher than that with 270 I $mg/m{\ell}$ in TACE. These results mean that contrast media can have an effect on patient's exposure dose and high iodine content contrast media results in elevation of patient's exposure dose.
Keywords
Contrast Media; Iodine; AEC;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 신성규, "골반촬영 시 내장된 필터를 이용한 환자 선량 감소", 한국콘텐츠학회논문지, 제12권, 제8호, pp.233-239, 2012.
2 마상철, Vascular and Interventional Radiology, 대학서림, 2007.
3 E. Vano, L. Arranz, and J. M. Sastre, "1998a. Dosimetric and radiation protection considerations based on some cases of patient skin injuries in interventional cardiology," Brit. J. Radiol, Vol.71, pp.510-516, 1998.   DOI
4 ICRP, Recommendations of the International Commission on Radiological Protection, ICRP Publication 26, Pergamon Press, Oxford, New york, 1997.
5 유명규, "중재적 시술 시 Detector의 높이와 영상의 확대에 따른 환자의 흡수선량 변화에 관한 연구", 대한인터벤션영상기술학회지, 제13권, 제1호, pp.122-132, 2010.
6 이종석, "흉부 및 복부에서 AEC 적용에 따른 MDCT의 선량 감소 효과", 한국콘텐츠학회논문지, 제9권, 제3호, pp.225-231, 2009.   과학기술학회마을   DOI   ScienceOn
7 Roberto lezzi, "High concentration contrast media and radiation dose reduction in CT vascular imaging," Diagnostic Imaging Europe, Vol.5, pp.1-17, 2011.
8 이재기, 방사선 방호를 위해 표준한국인 설정 연구, 한양대학교, 2004.
9 ICRP Publication 103 : Recommendations of the International Commission on Radiological Protection, Oxford, 2007.
10 정진욱, 인터벤션영상의학, 대한인터벤션영상의학회, 2007.
11 http://www.healthcare.philips.com/pwc_hc/main/shared/Assets/Documents/DoseWise/allura_xper_cardiac_system_imp_auto_dose_rate_ctrl.pdf
12 도영수, 혈관조영 X-선 장치의 안전관리 기준 개발, 식품의약품안전청 용역연구개발사업, 2003.