Browse > Article
http://dx.doi.org/10.5392/JKCA.2013.13.02.308

Analysis of Specific Absorption Rate in Magnetic Resonance Imaging  

Han, Jae-Bok (전남대학교 전자컴퓨터공학과)
Hong, Seong-Hun (전남대학교 전자컴퓨터공학과)
Choi, Nam-Gil (동신대학교 방사선학과)
Seong, Ho-Jin (조선대학교 원자력공학과)
Publication Information
Abstract
In this paper, we measured specific absorption rate (SAR) using characteristic variables such as flip angle, repetition time (TR) and echo time (TE) at magnetic resonance imaging. The subject was applied to same scan technique from body weight 10 kg to 90 kg, were measured for the average SAR and the peak SAR values according to the change of parameter. SAR with different body weight levels was not seen a significant change at TE but it increased in the larger flip angle and the shorter TR. SAR value was within the limits of human head acceptable standard and SNR in segmental body weights was not proportional to the increase of body weights. In conclusion, this study can be helpful for diagnosis by using appropriate parameters which obtained the various contrast and SNR.
Keywords
Magnetic Resonance Imaging; Repetition Time; Specific Absorption Rate;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 F. R. Korosec and C. A. Mistretta "MR angiography: basic principles and theory," Magn Reson Imaging Clin N am, Vol.6, pp.223-256, 1998.
2 FCC, OET Bulletin 65, Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency Electromagnetic fields, Edition 97-01, released August 25, 1997.
3 CENELEC, Considerations for Human Exposure to EMFs from Mobile Telecommunication Equipment (MTE) in the Frequency Range 30 MHz - 6 GHz, European Committee for Electrotechnical Standardisation SECRETARIAT SE 211/B WAMTE Fec. 1997.
4 J. Hennig and K. Scheffler, "Hyperechoes in RARE(TSE, FSE)-sequences," Pro. Intl. Soc. Mag Reson Med, Nol.9. pp.1769, 2001.
5 H. Bosmans, G. Marchal, G. Luktio, Time-of -flight MR angiography of the brain: comparison of acquistion techniques in healthy volunteers," Am J Roentgenol, Vo1.164, pp.161-167, 1995.
6 S. J. Go, "Metal Artifact Caused by Magnetic Field Strength and Sequence on T1WI-MRI," J. of Contents Association, Vol.10, No.9, pp.302-308, 2010.   DOI   ScienceOn
7 L. S. Babiars, J. M. Romero, E. K. Murphy, B. Brobeck, P. W. Schaefer, and R. G. Gonzlez, "Contrast enhanced MR angiography is not more accurate than unenhanced 2D time-of-flight MR angiography for determining internal cartoid artery stenosis," AJNR Am J Neuroradial, Vol.30, pp.760-768, 2009.
8 J. Alvarez Linera, J. Benito-Len, J. EscribanoJ, J. Campollo, and R. Gesto, "Prospective evaluation of cartoid artery stenosis: elliptic centric contrast-enhanced MR angiography and spiral CT angiography compared with digital subtraction angiography," AJNR Am J Neuroradial, Vol.24, pp.1012-1019, 2003.
9 K. S. Lee, H. Sim, J. H. Moon, and J. C. Oh, "Implementation about measurement of the head SAR and variable parameter according to operation control mode in brain MR study with 1.5Tesla," The Transactions of the Korean Institute of Electrical Engineers, Vol.4, pp58-60, 2007.
10 A. K. Lee, H. D. Choi, and J. S. Kim, "Analysis of SAR in a Human Head for a Celluar Phone," Journal of Electromagnetic Engineering and Science, Vol.9, No.6, pp.776-787, 1998.
11 American National Standard-Safety levels with respect to human exposure to radio frequency electromagnetic fields, 3 kHz to 300 GHz, ANSI/IEEE C95.1-1992.