Measurement of the Skin Dose of Patient Using the Optically Stimulated Luminescent Dosimeter at Diagnostic Radiography  

Kim, Jong-Eon (가야대학교 방사선학과)
Im, In-Chul (동의대학교 방사선학과)
Min, Byung-In (인제대학교 응용공학부)
Publication Information
Abstract
The purpose of this study is an measurement of the skin dose of a patient by using the OSLD(optically stimulated luminescent dosimeter) under several irradiation conditions of the X-ray beam for diagnostic radiography. The measurements of skin dose were performed for head, chest, and pelvis. And test of reproducibility was carried out at the chest. As a result, we obtained the skin dose at forehead of head to be 1.30 mSv. The skin doses at xiphoid process, breast and apex of the lung of the chest were acquired 0.92, 0.52 and 0.70 mSv, respectively. And we obtained the skin doses at the left pelvis and the right pelvis to be 2.78 and 3.08 mSv, respectively. As for reproducibility, a coefficient of variation was 0.033. The skin doses were exhibited the values corresponding from 1/100 to 1/17 of the dose limit of the public(50 mSv) at the deterministic effect. In order to make accurate measurements of the skin doses for each tube voltage, the measured values have to multiply by the displayed values of reader by a correction factor. The energy response of the OSLD with the tube voltage will be studied in the near future.
Keywords
Radiography; Optically Stimulated Luminescent Dosimeter; Skin Dose; Patient;
Citations & Related Records
연도 인용수 순위
  • Reference
1 K. Matthews and P. C. Brennan, "The application of diagnostic reference levels: general principles and a irish perspective," Radiography, Vol.15, pp.171-178, 2009.   DOI   ScienceOn
2 P. Engel-Hills, "Radiation protection in medical imaging," Radiography, Vol.12, pp.153-160, 2006.   DOI   ScienceOn
3 I. O. Olarinoye, and I. Sharifat, "A protocol for setting dose reference level for medical radiography in nigeria: a review," Bajopas, Vol.3, No.1, pp.138-141, 2010.
4 N. Küçük, A. Koliç, G. Kemikler, L. Özkan, and K. Engin, "Analyses of surface dose from high energy photon beams for different clinical setup parameters," Turk J Med Sci, Vol.32, pp.211-215, 2002.
5 S. Devic, J. Seuntjens, W. Abdel-Rahman, M. Evans, M. Olivares, and E. B. Podgorsak, "Accurate skin dose measurements using radiochromic film in clinical applications," Med. Phys., Vol.33, No.4, pp.1116-1124, 2006.   DOI   ScienceOn
6 R. Hanify and M. Salasky, "Single point detector(dot) type testing summary report," Landaur Inc, pp.1-11, 2006.
7 microStar user manual, Landauer Inc., 2009.
8 C. S. Reft, "The energy dependence and dose reponse of a commercial optically stimulated luminescent detector for kilovolatge photon, megavoltage photon, electron, proton, and carbon beams," Med. Phys., Vol.36, No.5, pp.1690-1699, 2009.   DOI