Change of PAE according to Detector Measurement Method  

Im, In-Chul (가야대학교 방사선학과, 방사선사고지원단(U-REST))
Publication Information
Abstract
The aim of this study is to investigate PAE, as the result of the test of kVp accuracy, according to detector measurement method. Based on the indicated value of 70kVp, each distance between a focus and a kVp meter was 100cm, 80cm and 60cm and the angle of X-ray tube was set on $5^{\circ},\;10^{\circ},\;15^{\circ},\;20^{\circ},\;25^{\circ},\;30^{\circ}$. Each indicated value, 60kVp, 70kVp, 80kVp, 90kVp and 100 kVp, was used compare Small focus with Large focus. As a result, PAE on the side of cathode was higher than it on the side of anode in the case of 100cm and PAE on the side of anode was higher in the case of 80cm and 60cm. The coefficient rate was stable both the side of cathode and anode in the case of 100cm and it was fluctuated in the case of 80cm and 60cm. PAE in the case of Small focus was higher than Large focus and it was disproportionate to an indicated value. Error rate was in inverse proportion to the indicated value.
Keywords
Percent Average Error; Tube Voltage; X-tube; Focus;