Noise Measurement by Percentage of Effective Linear Attenuation Coefficient of Water in CT Image of AAPM CT Performance Phantom  

Jong-Eon, Kim (Department of Radiological Science, Kaya University)
Sang-Hun, Lee (Department of Radiological Science, Kaya University)
Publication Information
Journal of the Korean Society of Radiology / v.16, no.6, 2022 , pp. 771-778 More about this Journal
Abstract
The purpose of this study is to present a method of measuring noise by the percentage of effective line attenuation coefficient of water that can be used for quality control of CT image noise using AAPM CT performance phantom in clinical practice. In the CT images obtained by scanning the AAPM CT performance phantom with a 120 kVp CT X-ray beam, the mean CT number was measured for each pin and water in the CT number linearity insert part. The effective energy was determined as the photon energy with the largest correlation coefficient from the correlation coefficients of the linear regression analysis of the measured mean CT number for each pin and water and the linear attenuation coefficient for each photon energy. And for water and acrylic, the contrast scale was calculated as 0.000188 cm-1 · HU-1 from the measured mean CT number and effective line attenuation coefficient. Using the calculated contrast scale, the effective line attenuation coefficient of water, and the standard deviation measured in the water of the alignment pin part of the AAPM CT performance phantom, The noise measurement value by the percentage of effective line attenuation coefficient of water obtained 0.31 ~ 0.52% in the range of 100 ~ 300 mAs.
Keywords
Noise; CT number; Linear attenuation coefficient; Correlation coefficient; Linearity;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 AAPM report No. 1, "Phantoms for Performance Evaluation and Quality Assurance of CT Scanners", American Association of Physicists in Medicine, pp. 1-23, 1977.
2 M. E. Phelps, E. J. Hoffman, M. M. Ter-Pogossian, "Attenuation Coefficients of Various Body Tissues, Fluids and Lessions at Photon Energies of 18 to 136 keV", Radiology, Vol. 117, No. 3, pp. 573-583, 1975. https://doi.org/10.1148/117.3.573   DOI
3 P. S. Rao, E. C. Gregg, "Attenuation of Monoenergetic Gamma Rays in Tissues", American Journal of Roentgenology, Vol. 123, No. 3, pp. 631-637, 1975. https://doi.org/10.2214/ajr.123.3.631   DOI
4 AAPM report No. 39, "Specification and Acceptance Testing of Computed Tomography Scanners", American Association of Physicists in Medicine, pp. 1-95, 1993.
5 http://www.doza.ru/docs/med/aapm_ct.pdf.
6 https://www.nist.gov/pml/xcom-photon-cross-sections-database.
7 M. R. Millner, W. H. Payne, R. G. Waggener, W. D. McDavid, M. J. Dennis, V. J. Sank, "Determination of Effective Energies in CT Calibration", Medical Physics, Vol. 5, No. 6, pp. 543-545, 1978. http://dx.doi.org/10.1118/1.594488   DOI
8 J. E. Kim, "Determination of Effective Energy of CT X-ray Beams", Journal of the Korean Society of Radiology, Vol. 13, No. 4, pp. 517-522, 2019. https://doi.org/10.7742/jksr.2019.13.4.517   DOI
9 J. E. Kim, S. H. Lee, "Derivation of the Effective Energy Calculation Formula of the X-ray Beam Generated by the CT Simulator", Journal of the Korean Society of Radiology, Vol. 15, No. 6, pp. 869-875, 2021. https://doi.org/10.7742/jksr.2021.15.6.869   DOI
10 P. F. Judy, "Comparison of Equivalent photon Energy Calibration Methods on Computed Tomography", Medical Physics, Vol. 7, No. 6, pp. 685-691, 1980. http://dx.doi.org/10.1118/1.594722   DOI
11 S. Okayama, T. Soeda, Y. Takami, R. Kawakami, S. Somekawa, S. Uemura, Y. Saito, "The Influence of Effective Energy on Computed Tomography Number Depends on Tissue Characteristics in Monoenergetic Cardiac Imaging", Radiology Research and Practice, Vol. 2012, pp. 1-7, 2012. https://doi.org/10.1155/2012/150980   DOI
12 J. Nosil, K. I. Pearce, R. A. Stein, "Linearity and Contrast Scale Control in Computed Tomography", Medical Physics, Vol. 16, No. 1, pp. 110-113, 1989. https://doi.org/10.1118/1.596396   DOI
13 J. E. Barnes, "Characteristics and Control of Contrast in CT", Radiographics, Vol. 12, No. 4, pp. 825-837, 1992. https://doi.org/10.1148/radiographics.12.4.1636042   DOI
14 C. D. Njiki, J. N. Manyol, Y. E. Yigbedeck, D. W. Abouou, B. C. Yimele, J. F. Sabouang, "Assessment of Image Quality Parameters for Computed Tomography in the City of Yaounde", Open Journal of Radiology, Vol. 8, No. 1, pp. 37-44, 2018. https://doi.org/10.4236/ojrad.2018.81005   DOI
15 M. J. Medinaa, F. J. Funesa, A. Palaub, V. M. Sanz, A. R. Osorioa, M. Namiasb, "Acceptance testing of a CT scanner with a knowledge-based iterative reconstruction algorithm", Brazilian Journal of Radiation Sciences, Vol. 7, No. 1, pp. 1-17, 2019. https://doi.org/10.15392/bjrs.v7i1A.456   DOI