Comparisons of Image Quality and Entrance Surface Doses according to Care Dose 4D + Care kV in Chest CT  

Kang, Eun-Bo (Department of Radiological Science, Dong-Eui Institute of Technology)
Publication Information
Journal of the Korean Society of Radiology / v.16, no.1, 2022 , pp. 45-51 More about this Journal
Abstract
This study compared DLP values along with phantom entrance surface doses and the image quality of chest CT scans made using a Care Dose 4D+Care kV System, scans that are made using only the Care Dose 4D function, and scans that are made with changes made by applying 80 kVp, 100 kVp, 120 kVp, and 140 kVp to the Care Dose 4D and tube voltage to search for methods to maintain the highest image quality with minimal patient doses. It was shown that DLP values decreased 6.727% when scans were taken with Chest Care Dose 4D + Care kV semi 100 and 6.481% when scans were taken with Chest Care Dose 4D + Care kV. With Chest Non as a standard, skin surface doses decreased 16.519% when scans were taken with Chest Care Dose 4D + Care kV semi 100 and 15.705% when scans were taken with Chest Care Dose 4D + Care kV. With comparisons of image quality, when comparisons were made with Chest Non, comparisons made of SNR values and CNR values in all scanning conditions including Care Dose 4D + Care kV showed that there were no significant differences at P>0.05. Imaging using Chest Care Dose 4D + Care kV in chest CT showed that exposure doses decreased similarly to result values gained from the best conditions through manual adjustments of kV and mAS, and there were no significant differences in image SNR and CNR. If the Chest Care Dose 4D + Care kV function is used, image quality is maintained and patient exposure to radiation can be reduced.
Keywords
Care Dose 4D; Care kV; SNR; CNR;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 http://www.mfds.go.kr.
2 J. Valentin, "Low-dose extrapolation of radiation-related cancer risk", Annals of the ICRP, Vol. 35, No. 4, pp. 1-140, 2005. https://doi.org/10.1016/j.icrp.2005.11.002   DOI
3 ICRP, 2007a. Biological and epidemiological information on health risks attributable to ionising radiation: a summary of judgements for the purposes of radiological protection of humans. Annex A to 2007 Recommendations.
4 ICRP 105, "ICRP Publication 105. Radiation protection in medicine", Vol. 37, No. 6, pp. 1-63, 2007. https://doi.org/10.1016/j.icrp.2008.08.001   DOI
5 Y. G. Kwak, C. Y. Kim, S. P. Jeong, "Research on Dose Reduction During Computed Tomography Scanning by CARE kV System and Bismuth", The Journal of the Korea Contents Association, Vol. 14, No. 8, pp. 233-242, 2014. https://doi.org/10.5392/JKCA.2014.14.08.233   DOI
6 UN Scientific Committee on the Effects of Atomic Radiation, Sources and Effects of Ionising Radiation. United Nations Scientific Committee on the Effects of Atomic Radiation Report to the General Assembly with Scientific Annexes, United Nations, New York, Vol. 1, p. 1-89, 2000.
7 http://opendata.hira.or.kr/op/opc/olapInfraEquipmentStatInfo.do
8 The Catholic University of Korea, "Development of guidelines for justification process and optimization of clinical protocol in radiation exposure from CT scan", Radiation Safety Division, National Institute of Food & Drug Safety Evaluation, 1475006214, 2011. https://scienceon.kisti.re.kr/commons/util/originalView.do?cn=TRKO201200007179&dbt=TRKO&rn=
9 K. P. Kim, "Assessment of Radiation Exposure of Korean Population by medical radiation", Korea Disease Control and Prevention Agency, 2020.https://www.kdca.go.kr/board/board.es?mid=a20305050000&bid=0003
10 ICRP, Managing Patient Dose in Computed Tomography, ICRP Publication 87. Ann. ICRP, Vol. 30, No. 4, 2000.
11 J. Valentin, "Managing patient dose in multi-detector computed tomography(MDCT). ICRP Publication 102", Annals of the ICRP, Vol. 37, No. 1, pp. 1-79, 2007. https://doi.org/10.1016/j.icrp.2007.09.001   DOI
12 NAS/NRC, Health Risks from Exposure to Low Levels of Ionising Radiation: BEIR VII Phase 2. Board on Radiation Effects Research, National Research Council of the National Academies, Washington, D.C. 2006.
13 J. M. Seok, W. J. Jeon, Y. J. Park, J. Lee, "Dose Reduction and Image Quality Assessment of the CareDose 4D Technique on Abdomen Liver Computed Tomography", Journal of the Korean Society of Radiology, Vol. 11, No. 3, pp. 109-115, 2017. https://doi.org/10.7742/jksr.2017.11.3.109   DOI