The purpose of this study is to derive a formula for calculating the effective energy of an X-ray beam generated by a CT simulator. Under 90, 120, and 140 kVp X-ray beams, the CT number calibration insert part of the AAPM CT performance phantom was scanned 5 times with a CT simulator. The CT numbers of polyethylene, polystyrene, water, nylon, polycarbonate, and acrylic were measured for each CT slice image. The average value of CT number measured under a single tube voltage and the linear attenuation coefficients corresponding to each photon energy calculated from the data of the National Institute of Standards and Technology were linearly fitted. Among the obtained correlation coefficients, the photon energy having the maximum value was determined as the effective energy. In this way, the effective energy of the X-ray beam generated at each tube voltage was determined. By linearly fitting the determined effective energies(y) and tube voltages(x), y=0.33026x+30.80263 as an effective energy calculation formula was induced.
Keywords
CT number; Tube voltage; Effective energy; CT simulator;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 J. E. Kim, S. h. Lee, "Determination of the Effective Energy of an X-ray Beam Uusing Optically Stimulated Luminescent nanoDot Dosimeters", Journal of Korean Society of Radiology, Vol. 9, No. 6, pp. 375-379, 2015. https://doi.org/10.7742/jksr.2015.9.6.375   DOI
2 Users Manual, "Nuclear Associates 76-410-4130 and 76-411 AAPM CT Performance Phantom", Fluke corporation. 2005.
3 M. R. Millner, W. H. Payne, R. G. Waggener, W. D. McDavid, M. J. Dennis, V. J. Sank, "Determination of Effective Energies in CT Calibration", Medical Physics, Vol. 5, No. 6, pp. 543-545, 1978. https://doi.org/10.1118/1.594488   DOI
4 AAPM report No. 1, "Phantoms for Performance Evaluation and Quality Assurance of CT Scanners", American Association of Physicists in Medicine, pp. 1-23, 1977.
5 J. E. Kim, "Measurement of CT Numbers for Effective Atomic Number and Physical Density of Compound", Journal of the Korean Society of Radiology, Vol. 15, No. 2, pp. 125-130, 2021. https://doi.org/10.7742/jksr.2021.15.2.125   DOI
6 https://www.physics.nist.gov/PhysRefData/Xcom/html/xcom1.html.
7 M. Hiraoka, M. Mitsumori, K. Okajima, Y. Nagata, M. Takahashi, M. Nakata, M. Abe, "Use of a CT Simulator in Radiotherapy Treatment Planning for Breast Conserving Therapy", Radiotherapy and Oncology, Vol. 33, No. 1, pp. 48-55, 1994. https://doi.org/10.1016/0167-8140(94)90085-x   DOI
8 J. E. Kim, "Determination of Effective Energy of CT X-ray Beams", Journal of the Korean Society of Radiology, Vol. 13, No. 4, pp. 517-522, 2019. https://doi.org/10.7742/jksr.2019.13.4.517   DOI
9 F. M. KHAN, The Physics of Radiation Therapy, 3rd Ed., Wolters Wluwer Co., New York, pp. 97-102, 2003.
10 National Institute of Standards and Technology, "NIST Measurement Services: Calibration of X-ray and Gamma-ray Measuring Instruments", NIST Special Publication 250-58, pp. 1-96, 2001.
11 C. B. Saw, A. Loper, K. Komanduri, T. Combine, S. Huq, C. Scicutella, "Determination of CT-to-Density Conversion Relationship for Image-Based Treatment Planning Systems", Medical Dosimetry, Vol. 30, No. 3, pp. 145-148, 2005. http://doi.org/10.1016/j.meddos.2005.05.001   DOI
12 S. C. Chen, W. L. Jong, A. Z. Harun, "Evaluation of X-ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem 78 models", Malaysian Journal of Medical Sciences, Vol. 19, No. 3, pp. 22-28, 2012.