A Study on the Quality Control of Transvaginal Ultrasound Transducer using ATS-539 Ultrasound Phantom  

Park, Ji Hye (Department of Radiology, Asan Medical Center)
Heo, Yeong Cheol (Department of Radiological Science, Graduate School, Eulji University)
Kim, Yon min (Department of Radiological Science, Wonkwang health Science University)
Han, Dong Kyoon (Department of Radiological Science, Graduate School, Eulji University)
Publication Information
Journal of the Korean Society of Radiology / v.15, no.4, 2021 , pp. 463-472 More about this Journal
Abstract
Demand for examinations using transvaginal transducer with high frequencies is increasing to observe pelvic organs in gynecological ultrasound tests. However, the quality control of the replacement probe in clinical trials is not properly implemented and the evaluation criteria have not been established. Therefore, 58 transvaginal transducers and 20 convex transducers were applied to the ATS-539 ultrasound phantom for 20 ultrasound devices currently in clinical use to obtain their respective images and measure them quantitatively and qualitatively. For quantitative measurements, vertical measurement, horizontal measurement, and focal zone and qualitative measurements, dead zone, axial·lateral resolution, sensitivity, functional resolution, gray scale·dynamic range were performed. Quantitative statistical analysis showed significant differences between the two transducers in the lateral measurement and local area (p<0.05). qualitative comparative analysis showed differences in sensitivity and functional resolution. This occurs due to the difference in frequency between transducers and the transducer's injection geometry. Based on the above experiments, the tolerance for horizontal measurement is raised to 10% (±8 mm), the tolerance for sensitivity is observed up to 6 cm deep, which is 12 cm deep,which is the level of the third quartile (75%). The permissible range of functional resolution is up to 6 (12 cm), 6 (12 cm), 11 (11 cm), 9 (9 cm), 6 (6 cm) target, which is the level of the third quartile (75%). It is considered reasonable to adjust the depth of targets in gray scale·dynamic range to measure at a depth of 2 cm, which is 50% of the depth of 4 cm. As above, the criteria for evaluating the quality of transvaginal transducer for use in the past have been proposed and it is expected that this study will be used as a basic data for the production of phantom exclusively for transvaginal transducer in the future.
Keywords
Transvaginal transducer; Ultrasound; Ultrasonic quality control; ATS-539 Ultrasound Phantom;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Mitchell M Goodsitt, Scott Wittc, David L. Hykes, James M. Kofler, Jr. Real-time B-mode ultrasound quality control test proceduresa Report of AAPM Ultrasound Task Group No.1. Medical Physics, Vol. 25, No. 8, 1998
2 ATS Laboratories. Clinical quality assurance phantoms: Multipurpose phantom model 539 Bridgeport, CT: ATS Laboratories Inc. 2000
3 Ministry of Health and Welfare. "Standards for reporting targets and identification code of medical equipment status". Ministry of Health and Welfare Notice No. 125, 2011
4 Browne JE, Watson AJ, Gibson NM, Dudley NJ, Elliott AT. Objective measurements of image quality. Ultrasound in Medicine and Biology. Vol. 30, No. 2, pp. 229-237, 2004   DOI
5 Y. Huh. Technical and Industrial Trends of Ultrasonic Medical Devices. Korea Evaluation Institute Of Industrial Technology. Vol. 6, pp. 49-50, 2011
6 E Merz 1, D Miric-Tesanic, F Bahlmann, G Weber, S Wellek. Sonographic size of uterus and ovaries in pre- and postmenopausal women, Ultrasound Obstet Gynecol. Vol. 7, pp. 38-42, 1996   DOI
7 C. H. Kim, S. N. Lee, E. K Kim. Use of Diagnostic Ultrasonography in Primary Care. Korean J Fam Pract. Vol. 2, No. 2, pp. 94-97, 2012
8 The Royal Colege of Radiologists. standard for the provision of an ultrasound service. London. 2014
9 S. E. Jung. Principles of quality management in medical imaging. Journal of the Korean Medical Association. Vol. 58, No. 12, pp. 1112-1118, 2015   DOI
10 P. N. Kim, J. W. Lim, H. C. Kim, Y. C. Yoon, D. J. Sung, J. S. Kim, J. C. Kim. Quality Assessment of Ultrasonographic Equipment Using an ATS-539 Multipurpose Phantom. Journal of the Korean Society of Radiology. Vol. 58, No. 5, pp. 533-541, 2008   DOI
11 S. H. Chung, H. J. Lee, H. S. Kim, J. Y. Oh. Health Insurance Benefit Criteria and Quality Assurance Policies of Diagnostic Ultrasound Service in Other Countries. Health Policy and Management. Vol. 24, No. 2, pp. 109-1191, 2014   DOI
12 Ministry of Health and Welfare. Details on the Application Criteria and Methods of Care Benefits」 Ministry of Health and Welfare Notice No. 2020-15, 2020
13 Ministry of Health and Welfare. Korean National Health Accounts In 2017
14 K. H. Kim, J. M. Chae, S. M. Kim, D. K. Lee, K. Y. Kim. A Study on the Evaluation of Ultrasonic Suitability. Health Insurance Review and Assessment Service. 2018
15 T. H. Im, D. G. Na. Annual report No. 1 2004-2006. Seoul, Korean Institute for Accreditation of Medical Imaging. pp. 1-28. pp. 75-105, 2007
16 J. H. Han. Analysis of a performance that use of TE phantom in ultrasound systems. The Korean Society of Medical Sonographers. Vol. 5, pp. 41-50, 2010
17 J. Y. Yu. Performance of Ultrasonographic Equipment in General Hospitals: Quality Assessment using an ATS-539 Multipurpose Phantom.. Jeollabuk-do, Graduate School of Chonbuk National University, 2012