A Study of the CT MAR using Single-Source and Dual-Source Devices: Practical Comparison using Animal Phantom Fabrication  

Goo, EunHoe (Department of Radiological Science, Cheonju University)
Publication Information
Journal of the Korean Society of Radiology / v.14, no.7, 2020 , pp. 1003-1011 More about this Journal
Abstract
This study aims to compare and evaluate the image differences between single and dual sources in applying a technique to reduce metal artifacts using dual energy CT. Discovery CT 256 (GE, USA) as a single source device and Somatom Definition Flash (Siemens Health Care, Forchheim, Germany) as a dual source device. The self-made phantom (pigs with medical titanium screws inserted) was quantitative and qualitatively evaluated under the same conditions by varying the dose under the same conditions using a dual energy CT. The evaluation method was compared by measuring SNR for metal artifacts (scattering, stripe) generated by metal inserts, divided around bones and around tissues. There was a difference in images in the method of reducing metal artifacts between single-source and dual-source devices. In a single source device, the linearized prosthesis by metal implantation showed a greater decrease than the image obtained from a double source device, and the surrounding tissue was well observed without interference from the artifact. In dual-source devices, scattering and stripe artifacts caused by metal inserts decreased more than on a single source device, and signals from adjacent tissues surrounding the metal implant were well observed without diminishing. If the examination is conducted separately between single source and dual source devices depending on whether the area to which the patient is intended to be viewed during the examination is adjacent to the metal insert or the total tissue surrounding the metal insert, it is believed that diagnostic helpful images can be obtained.
Keywords
Dual energy; Single source; Dual source; Metal artifact; SNR;
Citations & Related Records
연도 인용수 순위
  • Reference
1 J.Hsieh, R. C.Molthen, C. A. Dawson, R. H. Johnson, "An iterative approach to the beam hardening correction in cone beam CT", Medical physics, Vol. 27, No. 1, pp. 23-29, 2000. http://www.ncbi.nlm.nih.gov/pubmed/   DOI
2 T. D. Roth, K. A. Buckwalter, Choplin, R. H, "Musculoskeletal computed tomography: current technology and clinical applications", In Seminars in roentgenology, Vol. 48, No. 2, pp. 126-139, 2013. http://dx.doi.org/10.1053/j.ro.2012.11.009   DOI
3 D. Ropers, J. Rixe, K. Anders, A. Küttner, U. Baum, W. Bautz, S. Achenbach, "Usefulness of Multidetector Row Spiral Computed Tomography With 64- × 0.6-mm Collimation and 330-ms Rotation for the Noninvasive Detection of Significant Coronary Artery Stenoses", The American journal of cardiology, Vol. 97, No. 3, pp. 343-348, 2006. http://dx.doi.org/10.1016/j.amjcard.2005.08.050   DOI
4 T. Kaewlek, D. Koolpiruck, S. Thongvigitmanee, M. Mongkolsuk, S. Thammakittiphan, S. O. Tritrakarn, P. Chiewvit, "Metal artifact reduction and image quality evaluation of lumbar spine CT images using metal sinogram segmentation", Journal of X-ray Science and Technology, Vol. 23, No. 6, pp. 649-666, 2015. http://dx.doi.org/10.3233/XST-150518   DOI
5 S. Zhao, D. D. Robeltson, G. Wang, B. Whiting, K. T. Bae, "X-ray CT metal artifact reduction using wavelets: an application for imaging total hip prostheses", IEEE transactions on medical imaging, Vol. 19, No. 12, pp. 1238-1247, 2000. http://dx.doi.org/10.1109/42.897816   DOI
6 M. Kidoh, T. Nakaura, S. Nakamura, S. Tokuyasu, H. Osakabe, K. Harada, Y. Yamashita, "Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR)", Clinical radiology, Vol. 69, 1, pp. 11-16, 2014. http://dx.doi.org/10.1016/j.crad.2013.08.008   DOI
7 T. R. Johnson, B. Krauss, M. Sedlmair, M. Grasruck, H. Bruder, D. Morhard, T. Flohr, "Material differentiation by dual energy CT: initial experience", European radiology, Vol. 17, No. 6, pp. 1510-1517, 2007. http://dx.doi.org/10.1007/s00330-006-0517-6   DOI
8 R. K. Kaza, J. F. Platt, R. H. Cohan, E. M. Caoili, M. M. Al-Hawary, A. Wasnik, "Dual-energy CT with single-and dual-source scanners: current applications in evaluating the genitourinary tract", Radiographics, Vol. 32, No. 2, pp. 353-369, 2012. http://dx.doi.org/10.1148/rg.322115065   DOI
9 M. S. May, M. Wiesmueller, R. Heiss, M. Brand, J. Bruegel, M. Uder, W. Wuest, "Comparison of dual-and single-source dual-energy CT in head and neck imaging", European radiology, Vol. 29, No. 8, pp. 4207-4214, 2019. https://doi.org/10.1007/s00330-018-5762-y   DOI
10 F. Secchi, C. N. De Cecco, J. V. Spearman, J. R. Silverman, U. Ebersberger, F. Sardanelli, U. J. Schoepf, "Monoenergetic extrapolation of cardiac dual energy CT for artifact reduction", Acta radiologica, Vol. 56, No. 4, pp. 413-418, 2015. http://dx.doi.org/10.1177/0284185114527867   DOI
11 C. Zhou, Y. E. Zhao, S. Luo, H. Shi, L. Zheng, L. J. Zhang, G. Lu, "Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures", Academic radiology, Vol. 18, No. 10, pp. 1252-1257, 2011. http://dx.doi.org/10.1016/j.acra.2011.05.009   DOI
12 R. Donnino, J. E. Jacobs, J. V. Doshi, E. M. Hecht, D. C. Kim, J. S. Babb, M. B. Srichai, "Dual-source versus single-source cardiac CT angiography: comparison of diagnostic image quality", American Journal of Roentgenology, Vol. 192, No. 4, pp. 1051-1056, 2009. https://academic.naver.com/article.naver?doc_id=555911888   DOI
13 M. Wang, H. T. Qi, X. M. Wang, T. Wang, J. H. Chen, C. Liu, "Dose performance and image quality: dual source CT versus single source CT in cardiac CT angiography", European journal of radiology, Vol. 72, No. 3, pp. 396-400, 2009. http://dx.doi.org/10.1016/j.ejrad.2008.08.010   DOI
14 G. T. Herman, "Correction for beam hardening in computed tomography", Physics in Medicine Biology, Vol. 24, No. 1, pp. 81-106. 1979. http://dx.doi.org/10.1088/0031-9155/24/1/008   DOI
15 J. Philippe, S. Daniel, W. Stefan, B. Arno, M. Peter, "Comparison of Metal Artifact Reduction in Dual- and Single-Source CT: A Vertebral Phantom Study", American journal of roentgenology, Vol. 211, No. 6, pp. 1298-1305, 2018. http://dx.doi.org/10.2214/AJR.17.19397   DOI
16 G. H. Glover, "Compton scatter effects in CT reconstructions", Medical physics, Vol. 9, No. 6, pp. 860-867, 1982. https://doi.org/10.1118/1.595197   DOI
17 I. I. Goldman, "Intensity effects in Compton scattering", Soviet Physics Journal of Experimental and Theoretical Physics, Vol. 19, No. 4, pp. 954-957, 1964. http://dx.doi.org/10.1016/0031-9163(64)90728-0   DOI