A radiation spectroscopy detector using a spherical scintillator was designed, and a system was developed to track the position of a radiation source using several detectors. The position tracking algorithm was designed based on the theory that the number of radiations decreases according to the inverse square law of distance, and the position of the radiation source was calculated by measuring the number of radiations generated from the radiation sources at various positions. The radiation generated from the radiation source is detected by different coefficients in each detector, and the difference between these detected coefficients varies in proportion to the inverse square of the distance. Geant4 Application for Tomographic Emission (GATE) simulation was performed to verify and evaluate the performance of the designed radiation source position tracking system, and radiation generated from radiation sources placed at different positions was counted with each detector. The number of measured radiations was tracked through the radiation source position tracking algorithm, and the error between the actual radiation source position and the position calculated by the algorithm was evaluated. The error between the position of the actual radiation source and the calculated position was measured as an average of 0.11% on the X-axis and 0.37% on the Y-axis, and it was verified that the position can be measured very accurately.
Keywords
Radiation spectroscopy detector; Spherical scintillator; Inverse square low of distance; Radiation source position tracking;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 H. I. Kim, C. H. Baek, S. J. An, S. W. Kwak, Y. H. Chung, "Gamma camera with a two-layer diverging-slat collimator for radioisotope monitoring," Nuclear Instruments and Methods in Physics Research A, Vol. 698, pp. 90-93, 2013. http://dx.doi.org/10.1016/j.nima.2012.09.049   DOI
2 C. H. Baek, S. J. Lee, Y. Choi and Y. H. Chung, "Optimization of Large-Angle Pinhole Collimator for Environmental Monitoring System," IEEE Transations on Nuclear Science, Vol. 57, No. 3, pp. 1404-1408, 2010. http://dx.doi.org/10.1109/TNS.2009.2039877   DOI
3 D. Cester, G. Nebbia, L. Stevanato, G. Viesti, F. Neri, S. Petrucci, S. Selmi, C. Tintori, P. Peerani, A. Tomanin, "Special nuclear material detection with a mobile multi-detector system," Nuclear Instruments and Methods in Physics Research A, Vol. 663, pp. 55-63, 2012. http://dx.doi.org/10.1016/j.nima.2011.10.011   DOI
4 M. Gmar, O. Gal, C. Le Goaller, O. P. Ivanov, V. N. Potapov, V. E. Stepanov, F. Laine, and F. Lamadie, "Development of Coded-Aperture Imaging With a Compact Gamma Camera," IEEE Transactions on Nuclear Science, Vol. 51, No. 4, pp. 1682-1687, 2004. http://dx.doi.org/10.1109/TNS.2004.832608   DOI
5 C. H. Baek, H. I. Kim, J. Y. Hwang, S. J. An, K. H. Kim, S-W. Kwak, Y. H. Chung, "Large-angle pinhole gamma camera with depth-of-interaction detector for contamination monitoring," Nuclear Instruments and Methods in Physics Research A, Vol. 648, pp. S111-S115, 2011. http://dx.doi.org/10.1016/j.nima.2010.12.080   DOI
6 S. J. Lee, C. H. Baek, "Positioning Method for Radiation Source Using Radiation Spectroscopy Detectors," Journal of the Korean Physical Society, Vol. 77, No. 7, pp. 557-560, 2020. http://dx.doi.org/10.3938/jkps.77.557   DOI
7 K. P. Ziock, W. W. Craig, L. Fabris, R. C. Lanza, S. Gallagher, B. K. P. Horn, and N. W. Madden, "Large Area Imaging Detector for Long-Range, Passive Detection of Fissile Material," IEEE Transactions on Nuclear Science, Vol. 51, No. 5, pp. 2238-2244, 2004. http://dx.doi.org/10.1109/TNS.2004.836117   DOI
8 H. I. Kim, S. J. An, S. W. Kwak, Y. H. Chung, "Development of an All-in-one Gamma Camera/CCD System for Safeguard Verification," Journal of the Korean Physical Society, Vol. 65, No. 12, pp. 2013-2016, 2014. http://dx.doi.org/10.3938/jkps.65.2013   DOI
9 W. J. Jo, H. I. Kim, S. J. An, C. Y. Lee, H. K. Song, H. S. Shin, S. K. Ahn, S. H. Park, Y. H. Chung, "Feasibility Study of a Gamma Camera for Monitoring Nuclear Materials in the PRIDE Facility," Journal of the Korean Physical Society, Vol. 64, No. 9, pp. 1293-1297, 2014. http://dx.doi.org/10.3938/jkps.64.1293   DOI
10 Y. S. Kim, J. H. Kim, H. S. Lee and C. H. Kim, "Position-Sensitive NaI(Tl) Detector Module for Large-Area Compton Camera," Journal of the Korean Physical Society, Vol. 72, No. 1, pp. 26-32, 2018. http://dx.doi.org/10.3938/jkps.72.26   DOI
11 G. Santin, D. Strul, D. Lazaro, L. Simon, M. Krieguer, M. V. Martins, V. Breton and C. Morel, "GATE: a Geant4-based simulation platform for PET and SPECT integrating movement and time management," IEEE Transactions on Nuclear Science, Vol. 50, No. 5, pp. 1516-1521, 2003. http://dx.doi.org/10.1109/TNS.2003.817974   DOI
12 S. Jan, G. Santin, D. Strul, S. Staelens, et al., "GATE: a simulation toolkit for PET and SPECT," Physics in Medicine and Biology, Vol. 49, No. 19, pp. 4543-4561, 2004. http://dx.doi.org/10.1088/0031-9155/49/19/007   DOI