Comparison on the Error Rates of Calibration Modes in Intervention  

Kong, Chang gi (Department of Radiology, Cheom dan Hospital)
Ryu, Young hwan (Department of Radiological Science, Donggang University)
Publication Information
Journal of the Korean Society of Radiology / v.14, no.5, 2020 , pp. 619-626 More about this Journal
Abstract
This study aimed to identify the error rates in Catheter Calibration Mode, Auto Calibration Mode, and Segment Calibration Mode among many calibration modes as a quantitative evaluation tool used for predicting the diameter and length of balloon or stent in percutaneous intravascular balloon dilatation or stent insertion. Our experiment was conducted with Copper Wire of 2 mm × 80 mm (diameter × length) manufactured elaborately for quantitative evaluation in calibration and Metal Ball of 5, 10, 15, 30, and 40 mm and Acryl Phantom of 25 mm, 50 mm, 75mm, 100 mm, 125 mm, 150mm, 175 mm, and 200 mm. At each height, subtraction images were acquired with a cineangiograph and Stenosis Analysis Tool as a software provided by the equipment company was used for measurement. To evaluate the error rates in Catheter Calibration Mode, Copper Wire was put on each acryl phantom before shooting. Copper Wire of 2 mm in diameter was set as a diameter for catheter, and Copper Wire of 8 mm in length was measured with Multi-segments. As a result, the error rates appeared at 1.13 ~ 5.63%. To evaluate the error rates in Auto Calibration Mode, the height of acryl was entered at each height of acryl phantom and the length of 8 mm Copper Wire was measured with Multi-segments and as a result, the error rates appeared at 0 ~ 0.26%. To evaluate the error rates in Segment Calibration Mode, each metal ball on the floor of table was calibrated and the length of 8 mm Copper Wire on each acryl phantom was measured and the length of 8 mm Copper Wire depending on the changes of acryl phantom height was measured with Mutli-segments and as a result, the error rates appeared at 1.05 ~ 19.04%. And in the experiment on OID changes in Auto Calibration Mode, the height of acryl phantom was fixed at 100mm and OID only changed within the range of 450 mm ~ 600 mm and as a result, the error rates appeared at 0.13 ~ 0.38%. In conclusion, it was found that entering the height values in Auto Calibration Mode, among these Calibration Modes for evaluating quantitative vascular dimensions provided by the software was the calibration method with the least error rates and it is thus considered that for calibration using a metal ball or other objects, putting them in the same height as that of treatment sites before calibrating is the method that can reduce the error rates the most.
Keywords
Catheter Calibration; Auto Calibration; Segment Calibration;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Y. M. Han, K. Y. Kim, B. H. Chung, H. C. Yu, K. H. Kim, H. P. Hwang, "Rupture, Breakdown, and Pulmonary Artery Embolism of a Balloon Catheter Tip during Percutaneous Transluminal Angioplasty of Arteriovenous Fistula", Vascular specialist international, Vol. 35, No. 4, pp. 245-250, 2019. http://dx.doi.org/10.5758/vsi.2019.35.4.245   DOI
2 Johan H C Reiber, Shengxian Tu, Joan C Tuinenburg, Gerhard Koning, Johannes P Janssen, Jouke Dijkstra, "QCA, IVUS and OCT in interventional cardiology in 2011", Cardiovascular diagnosis and therapy, Vol. 1, No. 1, pp. 57-70, 2011. http://dx.doi.org/10.3978/j.issn.2223-3652.2011.09.03   DOI
3 Y. H. Kim, M. K. Hong, J. W. Kim, S. K. Lee, C. W. Lee, S. S. Cheong, K. J. Choi, D. H. Kang, J. K. Song, J. J. Kim, S. W. Park, S. J. Park "Comparison between Intravascular Ultrasound and Quantitative Coronary Angiographic Measurements in Coronary Artery Stenting", Korean Circulation Journal, Vol. 27, No. 12, pp. 1265-1271, 1997. http://dx.doi.org/10.4070/kcj.1997.27.12.1265   DOI
4 J. B. Lee, J. B. Seo, D. H. Kim, K. S. Kim, S. B. Kim, "comparison of Accuracyaccording to calibration method when measuring vessel", Journal of Korean Society of Cardio-Vascular Interventional Technology, Vol. 10, No. 1, pp. 92-96, 2007.
5 J. B. Seo, J. B. Lee, D. H. Kim, K. S. Kim, S. B. Kim, "A study on the usefulness of TOD (Table Object Distance) Calibration using CT Image when measuring Vessel diameter", Journal of Korean Society of Cardio-Vascular Interventional Technology, Vol. 11, No. 1, pp. 205-209, 2008.
6 S. G. Son, S. Y. Kim, W. H. Jo, M. S. Ahn, "Analysis of error rate according to calibration metal ball position when measuring vessel diameter", Journal of Korean Society of Cardio-Vascular Interventional Technology, Vol. 18, No. 1, pp. 155-164, 2015.
7 T. J. Kim, J. M. Goo, K. W. Lee, H. J. Lee, K. G. Kim, J. G. Im, "Quantitative Analysis of Airway Walls Using CT Software", J Korean Radiol Soc., Vol. 59, No. 6, pp. 395-403, 2008. http://dx.doi.org/10.3348/jkrs.2008.59.6.395   DOI
8 Y. K. Cho, S. H. Hur, S. H. Hwang, M. J. Kim, S. H. Lee, C. W. Nam, Y. S. Lee, D. W. Hyun, S. W. Han, K. S. Kim, Y. N. Kim, K. B . Kim, "A Comparison of the Intravascular Ultrasound Findings before and afterCoronary Stent Implantation in Patients with Small Vessel Disease(2.5-2.75 mm) by Quantitative Coronary Angiography", Korean Circulation Journal, Vol. 35, No. 2, pp. 123-130, 2005. http://dx.doi.org/10.4070/kcj.2005.35.2.123   DOI
9 M. S. Hyon, Masahiro Terashima, Michael V. McConnell, "Evaluation of the Coronary-Size Nitinol Stent patency with Magnetic Resonance Angiography", Journal of Soonchunhyang Medical Science, Vol. 12, No. 2, pp. 233-240, 2006.
10 http://opendata.hira.or.kr/op/opc/olapDiagBhvInfo.do
11 K. Y. Kim, S. H. Hur, Y. W. Cho, "Optimal Stent Expansion by Nominal Pressure Balloon Inflation: an Intravascular Ultrasound Study", Korean Circulation Journal, Vol. 32, No. 8, pp. 666-673, 2002. http://dx.doi.org/10.4070/kcj.2002.32.8.666   DOI
12 D. F. Fortin, L. A. Spero, J. T. Cusma, L. Santoro, R. Burgess, T. M. Bashore, "Pitfalls in the determination of absolute dimensions using angiographic catheters as calibration devices in quantitative angiography", The American Journal of Cardiology, Vol. 68, No. 11, pp. 1176-1182, 1991. https://doi.org/10.1016/0002-9149(91)90190-V   DOI