A Study on Photon Characteristics Generated from Target of Electron Linear Accelerator for Container Security Inspection using MCNP6 Code  

Lee, Chang-Ho (Department of Emergency Management, Inje University)
Kim, Jang-Oh (Department of Emergency Management, Inje University)
Lee, Yoon-Ji (Department of Emergency Management, Inje University)
Jeon, Chan-hee (Department of Emergency Management, Inje University)
Lee, Ji-Eun (Department of Emergency Management, Inje University)
Min, Byung-In (Department of Nuclear Applied Engineering, Inje University)
Publication Information
Journal of the Korean Society of Radiology / v.14, no.3, 2020 , pp. 193-201 More about this Journal
Abstract
The purpose of this study is to evaluate the photon characteristics according to the material and thickness of the electrons incidented through a linear accelerator. The computer simulation design is a linear accelerator target consisting of a 2 mm thick tungsten single material and a 1.8 mm and 2.3 mm thick tungsten and copper composite material. In the research method, First, the behavior of primary particles in the target was evaluated by electron fluence and electron energy deposition. Second, photons occurring within the target were evaluated by photon fluence. Finally, the photon angle-energy distribution at a distance of 1 m from the target was evaluated by photon fluence. As a result, first, electrons, which are primary particles, were not released out of the target for electron fluence and energy deposition in the target of a single material and a composite material. Then, electrons were linearly attenuated negatively according to the target thickness. Second, it was found that the composite material target had a higher photon generation than the single material target. This confirmed that the material composition and thickness influences photon production. Finally, photon fluence according to the angular distribution required for shielding analysis was calculated. These results confirmed that the photon generation rate differed depending on the material and thickness of the linear accelerator target. Therefore, this study is necessary for designing and operating a linear accelerator use facility for container security screening that is being introduced in the country. In addition, it is thought that it can be used as basic data for radiation protection.
Keywords
Container security inspection; MCNP6; Linear accelerator; Target; Photon characteristics;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Ministry of Oceans and Fisheries press release, "Development of the World's First 3-D High-Speed Container Inspection System", Port Management Division, pp. 1-2, 27. Feb. 2017.
2 G. S. Jo, "Large Container Radiation Inspection Technology and Domestic Status", Radioisotope Journal, Vol. 21, No. 2, pp.26-32, 2006.
3 Nuclear Safety Act, 2020.
4 S. M. Lee, "A Study on Evaluation of Dose Rate of Shielding Facility for Air cargo Security System using Monte Carlo Method", Hanyang University, pp. 10-17, 2019
5 Q. Gao, H. Zha, H. Chen, J. Shi, "Design and optimization of the target in electron linear accelerator", 4th International Particle Accelerator Conference, pp. 3663-3665, 2013.
6 F. A. A. Ajaj, N. M. H. Ghassal, "An MCNP-based model of a medical linear accelerator x-ray photon beam", Australasian Physical & Engineering Sciences in Medicine, Vol. 26, No. 3, pp. 140-144, 2003. https://doi.org/10.1007/BF03178784   DOI
7 M. Mauricio, B. Daniel, O. Emico, M. Helio, B. Marcio, M. Mario, "Application of the Monte Carlo in nuclear physics using the GEANT4toolkit", AIP conference Proceedings, Vol. 1139, No. 1, pp.51-56, 2009. https://doi.org/10.1063/1.3157829
8 S. I. Bae, S. H. Shin, "Calculation of Neutron Energy Distribution from the Components of Proton Therapy Accelerator Using MCNPX", Journal of the Korean Society of Radiology, Vol. 13, No. 7, pp. 917-924, 2019.
9 Denise B. Pelowiz, MCNP6 USER'S MANUAL Version 1.0, LA-CP-13-00634, Rev.0, 2013.
10 W. R. Nelson, "X-ray Production Yields for Linatrons 200A, 1000A and 3000A", WRN-98-1, pp. 1-3, 1998.
11 Linatron-M9&M9A Modular high-energy X-ray Source, Varian Medical systems, 2012.
12 James E. Clayton, Henderson, "X-ray Radiation Source with Low Neutron Emissions for Radiation Scanning", Patent Application Publication, US 2009/0041197A1, pp.1-2, 2008.
13 Z. Wang, H. Zha, J. Shi, H. Chen, "Thermal and Stress Analysis of an X-ray Target for 6 MeV Medical Linear Accelerators", 9th International Particle Accelerator Conference, pp. 572-574, 2018. https://doi.org/10.18429/JACoW-IPAC2018-MOPML070
14 Nuclear Data Center at KAERI "http://atom,kaeri.re.kr"
15 McConn, Ronald J, Gesh, Christopher J, Pagh, Richard T, Rucker, Robert A, & Williams, III, Robert, Compendium of Material Composition Data for Radiation Transport Modeling, United States, pp. 116-318, 2011. https://doi.org/10.2172/1023125
16 C. W. Lee, "Evaluation of Radiation Shielding in the Center for Proton Accelerators", Summer Workshop of The Korean Association for Radiation Protection", 2010.
17 Xu Hai-Bo, "Monte Carlo simulation for bremsstrahlung and photoneutron yields in high-energy x-ray radiography", Chinese Physics, B, Vol. 19, No. 6. 2010. http://dx.doi.org/10.1088/1674-1056/19/6/062901
18 I. Park, K. H. Kim, S. C. Oh, J. Y. Song, O, H. Kwon, K. P. Kim, "Evaluation of Radiation Dose to Patients in Intraoral Dental Radiography Using Monte Carlo Method", Journal of Radiation Industry, Vol. 10, No. 3, pp. 139-144, 2016.
19 WCO, Guidelines for the Procurement and Deployment of Scanning/NII Equipment, 2018
20 P. W. Nam, S. H. Han, "A Study on the Current Situations and Some Problems of Container Security Initiative(CSI) in Korea", Journal of Industrial Economics and Business, Vol. 17, No. 6, pp. 2651-2672, 2004.