Although magnetic resonance imaging without linear hardening of CT is recognized as a method of obtaining high contrast of tissue and excellent resolution image in brain disease and head and neck examination, magnetic susceptibility artifact is generated in case of metal implants in the oral cavity, which is an obstacle to image diagnosis. Therefore, an effort was made in this thesis to find a method to reduce artifacts caused by dental implants and prosthesis in MRI. Implant-induced artifacts in magnetic resonance imaging showed that the signal size increased with shorter TE in GE technique and was inconsistent with water temperature change. In SE technique as well, the signal size of water was generally higher than that of air, but the signal to noise ratio (SNR) was not different by air and temperature. In EPI technique, images with fewer artifacts were obtained quantitatively and qualitatively when there was more water than air, and the signal to noise ratio was measured the highest, especially at water temperatures of 20° and 30°. In conclusion, when examining using the EPI technique rather than the SE or the GE technique, obtaining brain diffusion using a 20° and 30° water bag reduces the magnetic susceptibility artifacts caused by implants and prosthesis, suggesting that it may provide images with high diagnostic value.
Keywords
Magnetic Resonance Imaging; Susceptibility artifact; Implant; Prostheses; Echo planar imaging; EPIA Study on the Artifact Reduction Method of Magnetic;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 F. G. Shellock, S. Morisoli, E. Kanal, "MR procedures and biomedical implants, materials, and devices. : 1993 updates," Journal of Radiology, Vol. 189, pp. 587-599, 1993.   DOI
2 Z. Starcuk, K. Bartusek, H. Hubalkova, T. Bachorec, J. Starcukova, P. Krupa, "Evaluation of MRI artifacts caused by metallic dental implants and classification of the dental materials in use," Measurement Science Review, Vol. 6, Section 2, 2006.
3 Jana Starcukova, Zenon Starcuk Jr, Hana Hubalkova. "Magnetic susceptibility and elecrical conductivity of metallic dental materials and their impact on MR imaging artifacts," Dental materials, Vol. 24, No. 6, pp. 715-723, 2008.   DOI
4 K. B. Pauly, G. E. Gold, Correction of slice profile distortions from metallic devices, Proceedings of the 14th Annual Meeting of ISMRM, Seattle, pp. 2380, 2006.
5 Forbes K. P., Pipe J. G., Karis J. P., Heiserman J. E., "Improved image quality and detection acute cerebral infarction with PROPELLER diffusion-weighted MR imaging," Journal of Radiology, Vol. 225. pp. 551-555, 2002.   DOI
6 M. Dujovny, N. Kossovsky, Ram, R Valdivia, JS Suk, et al. "Aneurysm clip motion during magnetic resonance imaging(In vivo experimental study with metallurgical factor analysis)," Journal of Neurosurgery, Vol. 17. pp. 543-548, 1985.   DOI
7 Frayne R., Goodyear B. G., Dickhoff P., Lauzon M. L., Sevic R. J., "Magnetic Resonance Imaging at 3.0 Tesla: challenges and advantages in clinical neurological imaging," Journal of Investigative Radiology, Vol. 38, pp. 385-402, 2003.
8 Bammer, R "Basic principles of diffusion-weighted imaging," European journal of radiology, Vol. 45, No.3, pp.169-184, 2003.   DOI
9 Josephs, O, Deichmann, R, Turner, R "Trajectory measurement and generalized reconstruction in rectilinear EPI," Journal of Neuroimage, Vol. 11, No.5, S543, 2000.   DOI
10 Woo Jeon Choi, Dong Hyun Kim. "Reduction of Artifacts in Magnetic Resonance Imaging with diamagnetic Substance," Journal of Society Radiology, Vol. 13, No. 4, pp. 581-588, 2019.
11 E. Mohamed, EI. Ghar. Abou, EI. Assmy. Ahmed, "Bladder Cancer: Diagnosis with Diffusion-weighted MR Imaging in Patients with Gross Hematuria," Journal of Radiology, Vol. 251, No. 2, pp. 415-421, 2009.   DOI
12 T. Masakatsu, C. S. Richard, Z. Mauricio, E J. Jorge, A. Ersan, P. Ertan, S. Kazuro, "Quantitative and qualitative comparison of 3.0 T and 1.5 T MR imaging of the liver in patients with diffuse parenchymal liver disease," European journal of Radiology, Vol. 72. No. 2. pp. 314-320, 2009.   DOI
13 X. Q. Xu, H. Hu, H. Liu, J. F. Wu, P. Cao, H. B. Shi, F. Y. Wu, "Benign and malignant orbital lymphoproliferative disorders: Differentiating using multiparametric MRI at 3.0T," Journal of Magnetic Resonance Imaging, Vol. 45. No. 1. pp. 167-176, 2017.   DOI
14 R. L. Becker, J. F. Norfray, G. P. Teitelbaum, W. G. Bradley, L. Wacaser, et al., "MR imaging in patients with intracranial aneurysm clips," AJNR, Vol 9, No. 5, pp.885-889, 1988.
15 Chul-Ho Sohn, In-Soo Kim, "PROPELLER (periodically rotated overlapping parallel lines with enhanced reconstruction) and EPI Diffusion-weighted MR Imaging at 3.0 T: Pontine Magnetic Susceptibility Artifacts Depend on Pneumatization of the Sphenoid Sinus," Journal of Korean Society Radiology, Vol. 55, pp. 321-326, 2006.   DOI
16 J. H. Lee, Y. S. Yu, J. S. Lee, I. C. Im, "Application of the Orally Inserted Guiding Device for the Improved Motion Artifacts of the Cervical Spine MRI," Journal of Korean Society Radiology, Vol. 8. No. 6, 2014.