Browse > Article
http://dx.doi.org/10.7742/jksr.2019.13.6.895

Assessment of Uterine Internal Temperature according to the Time of Convex Probe Injection using a Self-made Uterine Model Phantom  

Lee, Hyun-Kyung (Department of Radiological science, Hallym Polytechnic University)
Heo, Yeong-Cheol (Department of Radiological science, College of Health sciences, Eulji University)
Publication Information
Journal of the Korean Society of Radiology / v.13, no.6, 2019 , pp. 895-900 More about this Journal
Abstract
Ultrasound is known to be harmless to the human body and is widely used in obstetrics and gynecology to confirm the diagnosis and development status of fetus. Diagnosis Although long - term use of ultrasound may cause changes in body temperature, studies on the uterine temperature changes due to ultrasound have been lacking. The purpose of this study was to investigate the change of temperature according to ultrasonic scanning time using a self - produced uterine model phantom. Ultrasound equipment and a 4MHz convex probe were used to construct the uterine model phantom similar to the human uterus using acrylic and pig uterus, which are tissue equivalents. Three probe type thermometers were installed to measure the inside of the acrylic water tank, the uterus, and the atmospheric temperature. The temperature of the uterine phantom was ascertained by measuring the temperature of the subject for 6 hours, 361 times. In this study, the possibility of human body temperature elevation due to ultrasound could be confirmed and this study will be used as the basic data of ultrasonic heat absorption study.
Keywords
Ultrasound; Scan Time; Uterine Model Phantom; Heat Absorption index; convex probe;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 S. J. Shin, B. J. Jeong, "Principle and Comprehension of Ultrasound Imaging," Journal of the Korean Orthopedic Association, Vol. 48, pp. 325-333, 2013.   DOI
2 C. J. Lee, Y. J. Koo, M. J. Kang, "A Study on Development of Guideline to establish Safety and Performance Evaluation for Portable Ultrasound Diagnostic Device," FDC Legal research, Vol. 11, pp. 73-83, 2016.
3 S. H. Kim, S. L. Im, H. R. Kang, W. H. Choi, Y. R. Kim, A. R. Kim, "Basic research for strengthening pregnancy and childbirth support," Office of National Health Insurance Service, 2015.
4 H. S. Ko, "Significance of multidisciplinary counseling on prospective parents with fetus of congenital disease," Korean Journal of Obstetrics and Gynecology, Vol. 53, pp. 700-706, 2010.   DOI
5 S. H. Cheon, B. H. Choi, J. S. Noh, D. W. Han, M. Y. Park, "Poster Presentation: Maternal Fetal Medicine: Ultrasound Surveillance as a Prenatal Test in Korea," The Korean Association of Obstetricians and Gynecologists, Vol. 94, pp. 118-118, 2008.
6 M. L. Martinez, M. J. Gracia, L. F. Caidas, "High maternal fever during gestation," American Journal of Medicine, Vol. 98, pp. 201-203, 2010.
7 J. Y. Bae, S. Y. Hong, "Guideline for exercise in pregnancy," Journal of Korean Medical Association, Vol. 59, pp. 514-520, 2016.   DOI
8 J. S. Park, "Is Diagnostic Ultrasound Harmful to the Fetus," Journal of the Korean Medical Association, Vol. 51, pp. 823-830, 2008.   DOI
9 Stephen Bly, Michiel C Van den Hof, "Obstetric Ultrasound Biological Effects and Safety," Journal of Obstetric and Gynaecology Canada, Vol. 27, pp. 572-575, 2005.   DOI
10 P. J. Ross, J. Wu, P. L. Aida, "Interspecies Chimerism with Mammalian Pluripotent Stem Cells," Cell, Vol. 168, pp. 473-486, 2017.   DOI
11 M. H. Jeong "Study on the Examination of Deformity in Embryo Using Precision Ultrasonic Diagnosis Equipment," The Graduate school of life science Catholic University of Pusan, 2010.
12 Y. J. Jo, "A clinical pilot study of influence of temperature increase at rectum by ultrasonic energy to patients with constipation," The Graduate School Yonsei University, 2015.
13 C. K. Choi, H. S. Lim, "Effect of the Bean Sprouts Growth by Scanning Frequency of Diagnostic Ultrasound Probe Type and Mode," Journal of Biomedical Engineering Research, Vol. 36, pp. 150-154, 2015.   DOI
14 J. S. Halle, C. R. Scoville, D. G. Greathous, "Ultra sound's effect on the conduction latency of the superficial radial nerve in man," physical therapy, Vol. 61, pp. 345-350, 1981.   DOI
15 S. J. Eun "Temperature changes in water according to 4-D imensional ultrasonography and Comparison of MI and TI in each mode," Department of Radiological Science, The Graduate School Catholic University of Daegu, 2016.
16 J. W. Kim, H. S. Lim, "A Study On Temperature Changes In Thigh According To Diagnostic Ultrasonography Scan Time," Journal of the Korean Society of Radiology, Vol. 6, pp. 201-206, 2012.   DOI
17 J. H. Jung, J. S. Kim, K. L. Ha, M. J. Kim, "Tissue Mimicking Phantom for Visualization of Temperature Elevation Caused by Ultrasound," The Journal of the Acoustical Society of Korea, Vol. 33, pp. 291-299, 2014.   DOI
18 Y. T. Kim, J. J. Moon, Y. H. Yun, Y. C. Pu, "Thermal Sensitivity of the Bean Curd by Ultrasonic Irradiation," The Journal of the Acoustical Society of Korea, Vol. 23, pp. 503-513, 2004.