Design of Two Layer Depth-encoding Detector Module with SiPM for PET  

Lee, Seung-Jae (Department of Radiological Science, Dongseo University)
Publication Information
Journal of the Korean Society of Radiology / v.13, no.3, 2019 , pp. 319-324 More about this Journal
Abstract
A depth-encoding detector module with silicon photomultipliers(SiPMs) using two layers of scintillation crystal array was designed, and the position measurement capability was verified using DETECT2000. The depth of interaction of the crystal pixels with the gamma rays was tracked through the image acquired with the combination of surface treatment of the crystal pixels and reflectors. The bottom layer was treated as a reflector except for the optically coupled surfaces, and the crystals of top layer were optically coupled each other except for the outer surfaces so that the light sharing was made easier than the bottom layer. Flood images were obtained through the combination of specular reflectors and random reflectors, grounded and polished surfaces of crystal pixels, and the positions at which layer images were generated were measured and analyzed. The images were reconstructed using the Anger algorithm, whose the SiPM signals were reduced as the 16-channels to 4-channels. In the combination of the grounded surface and all reflectors, the depth positions were discriminated into two layers, whereas it was impossible to separate the two layers in the all polished surface combinations. Therefore, using the combination of grounded surface crystal pixels and reflectors could improve the spatial resolution at the outside of the field of view by measuring the depth position in preclinical positron emission tomography.
Keywords
SiPM; depth of interaction; reflector; DETECT2000;
Citations & Related Records
연도 인용수 순위
  • Reference
1 M. Streun, G. Brandenburg, H. Larue, H. Saleh, E. Zimmermann, K. Ziemons, and H. Halling, "Pulse Shape Discrimination of LSO and LuYAP Scintillators for Depth of Interaction Detection in PET," IEEE Transactions on Nuclear Science, Vol. 50, No. 3, pp. 344-347, 2003.   DOI
2 Y. Shao, R. W. Silverman, R. Farrell, L. Cirignano, R. Grazioso, K. S. Shah, G. Visser, M. Clajus, T. O. Tumer, S. R. Cherry, "Design Studies of A High Resolution PET Detector Using APD Arrays," IEEE Transactions on Nuclear Science, Vol. 47, No. 3, pp. 1051-1057, 2000.   DOI
3 Y. Yang, P. A. Dokhale, R. W. Silverman, K. S. Shah, M. A. McClish, R. Farrell, G. Entine and S. R. Cherry, "Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes," Physics in Medicine and Biology, Vol. 51, pp. 2131-2142, 2006.   DOI
4 S-J. Lee, C-H. Baek, "Three-dimensional cross point readout detector design for including depth information", Nuclear Instruments and Methods in Physics Research A, Vol. 887, pp. 13-17, 2018. doi: 10.1016/j.nima.2017.12.069   DOI
5 http://www.sensl.com/downloads/ds/PB-MatrixB.pdf
6 M. Ito, S. J. Hong, J. S. Lee, "Positron Emission Tomography (PET) Detectors with Depth-of-Interaction (DOI) Capability," Biomedical Engineering Letters, Vol. 1, No. 2, pp. 70-81, 2011.   DOI
7 F. Cayouette, D. Laurendeau, C. Moisan, "DETECT 2000: an improved Monte-Carlo simulator for the computer aided design of photon sensing devices," Proceedings of SPIE, Quebec, Vol. 4833, pp. 69-76, 2003.
8 http://www.furukawa-denshi.co.jp/cgi-bin/pdfdata/20140428162950.pdf
9 H. Peng, C. S. Levin, "Recent Developments in PET Instrumentation," Current Pharmaceutical Biotechnology, Vol. 11, No. 6, pp. 555-571, 2010.   DOI
10 R. S. Miyaoka, T. K. Lewellen, H. Yu, D. L. McDaniel, "Design of a Depth of Interaction (DOI) PET Detector Module," IEEE Transactions on Nuclear Science, Vol. 45, No. 3, pp. 1069-1073, 1998.   DOI