Cross Calibration of Dual Energy X-ray Absorptiometry Equipment for Diagnosis of Osteoporosis: between Domestic Manufacturers and Global Manufacturers  

Kim, Jung-Su (Department of Radiologic Technology, Chungbuk Health & Science University)
Publication Information
Journal of the Korean Society of Radiology / v.12, no.7, 2018 , pp. 833-844 More about this Journal
Abstract
Dual energy X-ray absorptiometry is mainly used as an X-ray test method. For equipment manufactured GE and Hologic, cross-calibration analyses (CCA) of machines from the same manufacturer and between units from different manufacturers have been conducted, but the CCA of equipment manufactured in Korea are inadequate. Through CCA, we present a formula of the intersections between the Korean medical equipment company (KEC) with GE and Hologic manufactured DXA, and among the KEC DXA. The CCA was conducted for the European Spine Phantom on DXA from four KEC and three global medical equipment company (GEC) manufacturers. We compared bone mineral density (BMD) values and calculated the CCA equation by linear regression analysis. The standard-deviations (SD) of the BMD values were highest for the Dexxum T for the low, medium, and high spine, which were 0.030, 0.029, and 0.037, respectively. The smallest SD in the low and medium vertebrae were 0.005 and 0.004 for the Horizon Ci, respectively, and 0.005 for the Osteo Pro Max in the high vertebrae. Based on the intersection equations of the KEC DXA established in this study, CCA of various KEC DXA should be established for more accurate follow-up of BMD tests in clinical environments.
Keywords
Dual energy X-ray absorptiometry equipment; Dual energy X-ray absorptiometry; Cross calibration; Osteoporosis;
Citations & Related Records
연도 인용수 순위
  • Reference
1 J. Y. Yang, Y. M. Kim, "Correlation Analysis of BMD in Proximal Femur and Spine," Journal of the Korean Fracture Society, Vol. 16, No. 4, pp. 570-576, 2003.   DOI
2 R. B. Mazess, H. S. Barden, J. P. Bisek, J. Hanson, "Dual energy X-ray absorptiometry for total-body and regional bonemineral and soft-tissue composition," The American Journal of Clinical Nutrition, Vol. 51, No. 6, pp. 1106-1112, 1990.   DOI
3 P. Eiken, N. Kolthoff, O. Barenholdt, F. Hermansen, S. Pors Nielsen, "Switching from DXA pencil-beam to fan-beam. II: studies in vivo," Bone Vol. 1, No. 6, pp. 667-670, 1994.
4 H. W. Wahner, W. L. Dunn, M. L. Brown ML, B. L. Riggs, "Comparison of dual energy x-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine," Mayo Clinic Proceedings. Vol. 63, No. 11, pp. 1075-1084, 1988.   DOI
5 E. S. Orwoll, S. K. Oviatt, "Longitudinal precision of dual- energy x-ray absorptiometry in a multicenter study," Journal of Bone and Mineral Research, Vol. 6, No. 2, pp. 191-197, 1991.   DOI
6 J. A. Shepherd, Y. Lu, K. Wilson, T. Fuerst, H. Genant, T. N. Hangartner, C. Wilson, D. Hans, E. S. Leib, "Cross-calibration and minimum precision standards for dual energy X-ray absorptiometry: the 2005 ISCD Official Positions, the 2005 ISCD Official Positions," Journal of Clinical Densitometry, Vol. 9, No. 1, pp. 31-36, 2006.   DOI
7 Osteoprim spec sheet. Accessed at 29. November. 2018. http://orvosimuszer.optitechplus.hu/pdf/medonica/osteoprimacatalog.pdf
8 B. Abrahamsen, J. Gram, T. B. Hansen, H. Beck-Nielsen, "Cross calibration of QDR-2000 and QDR-1000 dual energy X-ray densitometers for bone mineral and soft-tissue measurements," Bone, Vol 16, No. 3, pp. 385-390, 1995.   DOI
9 B. Oldroyd, A. H. Smith, J. G. Truscott, "Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers - bone mineral density and body composition studies," European journal of clinical nutrition, Vol. 57, No. 8, pp. 977-987, 2003.   DOI
10 H. k. Genant, S. Grampp, C. C. Gluer, K. G. Faulkner, M. Jergas, K. Engelke, H. Satoshi, C. van Kuijk, "Universal standardization for dual x-ray absorptiometry: Patient and phantom cross-calibration results," Journal of bone and mineral research, Vol. 9, No. 10, pp. 1503-1514, 1994.   DOI
11 H. Hull, Q. He, J. Thornton, J. Fahad, A. Lynn, N. P, Richard, G. Dympna, "iDXA, Prodigy, and DPXL dual energy X-ray absorptiometry whole-body scans: a cross-calibration study," Journal of Clinical Densitometry. Vol. 12, No. 1, pp. 95-102, 2009.   DOI
12 The International Society for Clinical Densitometry web site. Accessed at 29. November. 2018. https://www.iscd.org/official-positions/official-positions/
13 W. A. Kalender, D. Felsenberg, H. K. Genant, M. Fischer, J. Dequeker, J. Reeve, "The European Spine Phantom a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT," European journal of radiology, Vol. 20, No. 2, pp. 83-92, 1995.   DOI
14 D. Pearson, S. A. Cawte, D. J. Green. A, "Comparison of Phantoms for Cross-Calibration of Lumbar Spine DXA," Osteoporosis International, Vol. 13, No. 12, pp. 948-954, 2002.   DOI
15 J. S Kim, Y. H Rho, I. J Lee, K. A. Kim, I. J. Lee, J. M. Kim, "Cross-Calibration of Domestic Devices and GE Lunar Prodigy Advance Dual Energy X-Ray Densitometer Devices for Bone Mineral Measurements," Journal of Radiation Industry. Vol. 11, No. 1, pp. 27-31, 2017.
16 N. A. Pocock, P. N. Sambrook, T. Nguyen, P. Kelly, J. Freund, J. A. Eisman, "Assessment of spinal and femoral bone density by Dual X-Ray absorptiometry: Comparison of lunar and hologic instruments," Journal of Bone and Mineral Research, Vol. 7, No. 9, pp. 1081-1084, 1992.   DOI