Browse > Article
http://dx.doi.org/10.7742/jksr.2018.12.5.669

Analysis of the Relationships Between ESD and DAP, and Image SNR·CNR According to the Frame Change of Cine Imaging in CAG : With Focus on 10 f/s and 15 f/s  

Jung, Myo-Young (Department of Radiology, Dongshin University)
Seo, Young-Hyun (Department of Radiology, Dongshin University)
Song, Jong-Nam (Department of Radiology, Dongshin University)
Han, Jae-Bok (Department of Radiology, Dongshin University)
Publication Information
Journal of the Korean Society of Radiology / v.12, no.5, 2018 , pp. 669-675 More about this Journal
Abstract
This study aimed to investigate the difference of X-ray exposure by comparing and analyzing entrance surface dose and absorbed dose according to the frame change in coronary angiography using an X-ray machine. Moreover, appropriate frame selection measures for examination, including the effect of frame change on the image quality, were sought by measuring and analyzing the SNR and CNR of the image through image J. The study was conducted on 30 patients (19 males and 11 females) who underwent CAG at this hospital from June 2017 to October 2017. In regard to the patients, their age range was 49-82 years (mean of $65{\pm}9$ years), body weight was 45-91 kg (mean of $67{\pm}8.9kg$), height was 150-179cm (mean of $165.1{\pm}8.9kg$), and BMI was 19.5-30.5(mean of $24.5{\pm}2.9$). For the entrance surface dose and absorbed dose, air kerma value and DAP were obtained and analyzed retrospectively. The SNR and CNR were measured and analyzed through imageJ, and the result values were derived by applying the values to the formula. As for the statistical analyses, the correlations between the entrance surface dose and absorbed dose, and between the SNR and CNR were analyzed by using the SPSS statistical program. The relationship between the entrance surface dose and absorbed dose was not statistically significant for both 10 f/s and 15 f/s (p>0.05). In terms of the relationship between the SNR and CNR, the SNR ($3.374{\pm}2.1297$) and CNR ($0.234{\pm}0.2249$) in 10 f/s were $1.43{\pm}0.4861$ and $0.132{\pm}0.0555$ lower, respectively, than the SNR ($4.929{\pm}2.8532$) and CNR ($0.391{\pm}0.3025$) in 15 f/s, which were not statistically significant (p>0.05). In the correlation analysis, statistically significant results were obtained among the BMI, air kerma, and DAP; between air kerma and DAP; and between SNR and CNR (p<0.001, p<0.001). In conclusion, there was no significant difference between the entrance surface dose and absorbed dose even when the images were taken by changing the frame from 10 f/s to 15 f/s at the time of the coronary angiography. SNR and CNR increased at 15 f/s than at 10 f/s, but they were not statistically significant. Therefore, this study suggests that the concern of the patient and practitioner regarding image quality degradation, as well as the problem of X-ray exposure caused by imaging at 10 f/s and 15 f/s, may be reduced.
Keywords
Entrance surface dose; absorbed dose; signal-to-noise ratio; contrast-to-noise ratio;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 T. Y. Yun, J. W. Son, K. J. Kim, and et al, "The analysis of scattered radiation distribution depends on changing tube angle in angiography suite", Cardio-Vascular Interventional Technology, Vol. 20, No. 1, pp. 55-63, 2017.
2 W. G. Jeong "Radiation exposure and its reduction in the fluoroscopic examination and fluoroscopy-guided interventional radiology", Journal of Korean Medical Association, Vol. 54, No. 12, pp. 1269-1276, 2011.   DOI
3 J. C. Gurley, S. E. Nissen, D. C. Booth, A. N. Demaria "Influence of operator-and patient-dependent variables on the suitability of automated quantitative coronary arterygraphy for routine clinical use", Journal of the American College of Cardiology, Vol. 19, No. 6, pp. 1237-1243, 1992.   DOI
4 Y. D. Kim, I. Park, S. T. Kim, and et al, "Interoserver Variability in the Assessment of Coronary Arteriogram -Comparison between Visual and Computer based Quantitative Estimation-", Radiographics, Vol. 23, No. 6, pp. 857-866, 1998.
5 L. Finci, B. Meirer, G. Steffenino, P. Roy, W. Rutishauser, "Radiation exposure during diagnostic catheterization and single- and double-vessel percutaneous transluminal coronary angioplasty", Journal of the American College of Cardiology, Vol. 60, No, 16, pp. 1401-1403, 1987.   DOI
6 S. M. Park, J. M. Park, S. Y. Lee, S. I. Gang, Y. B. Park, D. H. Han, "A Study on the Comparison of Deep Dose of Fluoroscopy and Cine Angiography in Heart Examination," Journal of the Korean Society of Cardio-Vascular Interventional Technology, Vol. 17, No. 1, pp. 145-150, 2014.
7 Y. H. Kang, P. G. Jo, "Reduction of Radiation Dose according to Geometric Parameters from Digital Coronary Angiography," Journal of the Korean Society of Radiology, Vol. 7, No. 4, pp. 277-284, 2013.   DOI
8 Ministry of Food and Drug Safety, "Radiation protection guidelines for patient dose reduction in interventional procedures," Radiation safety management series, No. 36, 2014.
9 J. W. Jaaco, D. L. Miler, "Measuring and monitoring radiation dose during fluoroscopically guided procedures, Tech Vasc Interv Tadiol, Vol. 13, No. 3, pp. 188-193, 2010.   DOI
10 O. J. Park, J. J. Kim, H. R. Ryu, H. W. Lee, "Delayed clinical symptom and treatment seek time in patients with acute myocardial infarction", Journal of Korean Academy of Nursingy, Vol. 30, No. 3, pp. 659-669, 1997.
11 J. I. Choi, D. G. Na, H. H. Kim, M. Y. Shin, K. J. An, J. Y. Lee, "Quality Control of Medical Imaging", Korean Journal of Radiology, Vol. 50, No. 5, pp. 317-331, 2004.   DOI
12 S. S. Kang, I. H. Go, H. G. Goo. et al, "Textbook of Vascular and Interventional Radiography", 2nd Ed.. The Korean Society of Medical Imaging Technology.. Korea, pp. 114-115, 2012.
13 Y. E. Kim, J. H. Kim, J. S. Kim, C. M. Kim, "Comparison of computed tomography (CT) dose and image quality - Dyna CT and Multi Detector CT center", Journal of the Korean Society of Cardio-Vascular Interventional Technology, Vol. 17, No. 1, pp. 163-172, 2014.
14 Y. E. Kim, J. S. Kim, J. H. Yun,, T. I. Kim, "A Study on the Evaluation of Dose and Quality of Image by C-Arms Angle and Copper Filter Thickness during Cerebral Angiography and Intervention", Journal of the Korean Society of Cardio-Vascular Interventional Technology, Vol. 20, No. 1, pp. 152-160, 2017.   DOI