Comparison of Dose and Quality of Copper and Nickel Additional Filter Plate in Diagnostic X-ray Generator  

Lee, Hyun-Kyung (Department of Radiological Science, Eulji University)
Go, Yu-Rim (Department of Radiological Science, Eulji University)
Park, Young-Kyeong (Department of Radiological Science, Eulji University)
Han, Dong-Kyoon (Department of Radiological Science, Eulji University)
Publication Information
Journal of the Korean Society of Radiology / v.11, no.6, 2017 , pp. 459-466 More about this Journal
Abstract
The purpose of this study was to evaluate the difference of dose and image quality according to the material of the additional filter plate by selecting copper and nickel. First, the absorbed dose was measured using a Rando phantom setting the additional filter plates of copper and nickel None, 0.1 mm, 0.2 mm, and 0.3 mm under 120 kVp, and 6.3 mAs. Second, We acquired image according to filter thickness of copper and nickel. by changing the tube voltage of 90 kVp, 100 kVp, 110 kVp, 120 kVp and exposure indexes of 400, 800 and 1600. Third, we obtained the SNR and CNR values using the Image J program and evaluated quantitatively and then evaluated image quality. As a result, Absorbed dose measurements showed that nickel was higher than copper, and the absorbed dose decreased as the thickness increased(p<0.05). Furthermore, Quantitative analysis of images showed no significant difference between the two images according to change the voltage and the exposure index(p>0.05). In conclusion, this study confirms that the nickel addition plate can maintain the current image quality while reducing the exposure dose compared to copper.
Keywords
Nickel; Copper; AEC; EI; Radiation dose; Image quality;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 J. H. Hwang, K. H. Lee, Y. B. Choi, B. S. Kang, "Effectiveness evaluation of the tube voltage measurement by using additional filter," Journal of radiological science and technology, Vol. 38, No. 4, pp. 355-363, 2015.   DOI
2 M. Ganten, B. Radeleff, A. Kampschulte, MD. Daniels, GW. Kauffmann, J. Hansmann, "Comparing image quality of flat-panel chest radiography with storage phosphor radiography and film-screen radiography," American Journal of Roentgenology(AJR), Vol. 181, No. 1, pp. 171-176, 2003.   DOI
3 R. T. M. Van Soldt, D. Zweers, L. van den Berg, J. Geleijns, J. T. Jansen, J. Zoetelief, "Survey of posteroanterior chest radiography in The Netherlands: patient dose and image quality," The British Journal of Radiology(BRIT J RADIOL), Vol. 76, No. 6, pp. 398-405, 2003.
4 C. E. Willis, "Strategies for dose reduction in ordinary radiographic examinations using CR and DR," Pediatric Radiology(PEDIATR RADIOL), Vol. 34, No. 3, pp. 196-200, 2004.   DOI
5 S. K. Jang, "DAP measurement and patient dose analysis for diagnostic digital radiography system and film-screen system," Korea University, 2012.
6 A. D. Wrixon, "New ICRP recommendations. Journal of radiological protection," Vol. 28, No. 2, pp. 161-168, 2008.   DOI
7 ICRP, "The evolution of system of radiological protection," International commission on radiological protection(ICRP), Vol. 23, No. 2, pp. 129-142, 2003.
8 H. L. Arkins, R. G. Fairchild, J. S. Robertson, D. Greenburg, et al, "Effect of absorption edge filters on diagnostic X-ray spectra," Radiology, Vol. 115, No. 2, pp. 431-437, 1975.   DOI
9 M. Y. Kim, "Study for Optimal exposure condition of General Chest Projection using Film, CR, DR," International University of Korea(IUK), 2012.
10 ICRP, "Radiological Protection and Safety in Medicine," Pergamon Press, Oxford, Vol. 73, No. 9, 1996.
11 L. J. Krof, W. J. Veldkamp, B. J. Mertens, J. P. van Delft, J. Geleijns, "Detection of simulated nodules on clinical radiographs: dose reduction at digital post eroanterior chest radiology," Radiology, Vol. 241, No. 2, pp. 392-398, 2006.   DOI
12 A. B. d. Gonzalez, S. Darby, "Risk of cancer from diagnostic X-rays:estimates for the UK and 14 other countries," Lancet31 Vol. 363, No. 9406 pp. 345-351, 2004.   DOI
13 J. Heo, C. G. Kim, I. J. Lee, H. S. Shin, J. C. Park, H. S. Kang, B. S. Ahn, "A study on the effect of rare earth screen filter. Journal of Radiological Science and Technology Vol. No. 1 11, pp. 17-23, 1988.
14 J. T. Dobbins, E. Samei, H. G. Chotas, et al, "Chest radiography:optimization of x-ray spectrum for cesium iodide-amorphous silicon flat-panel detector," Radiology, Vol. 226, No. 1, pp. 221-230, 2003.   DOI
15 W. H. Okka, B. S. Claude, S. Michael, et al, "Chest radiography with a flat-panel detector: Image quality with dose reduction after copper filtration," Radiology, Vol. 237, No. 2, pp. 691-700, 2005.   DOI
16 G. J. Yeon, "Flat-panel Digital Radiography of the Chest:comparison patient dose, image quality and DQE using copper Filter," Korea University, 2012.
17 J. S. Lee, S. J. Ko, S. S. Kang, J. H. Kim, H. H. Park, D. H. Kim, C. S. Kim, "Evaluation of Image Quality for Diagnostic Digital Chest Image Using Ion Chamber in the Total Mastectomy," The Journal of the Korea Contents Association(JKCA), Vol. 13, No. 3, pp. 204-210, 2013.   DOI
18 K. S. Kim, S. C. Kim, "Comparison of effective dose and Image Quality by additional in Digital Chest tomosynthesis," Vol. 38, No. 4, pp. 347-353, 2015.   DOI
19 S. H. Kim, The effect of x-ray filter materials in the skin dose and tube loading at x-ray examinations," Eulji University, 2015.
20 M. Y. Lee, "The effect of x-ray filter materials in the skin dose and tube loading at x-ray examinations," Korea University, 2005.
21 B. R. Park, "The Comparative Study of Image Quality and Radiation Dose with Changes of Tube Voltage and Current in Digital Chest Radiography, Kyung Hee University, 2007.
22 A. Madani, V. De Maertelaer, J. Zanen, P. A. Gevenois, "Pulmonary emphysema: radiation dose and section thickness at multidetector CT quantification-comparison with macroscopic and microscopic morphemtry," Radiology, Vol. 243, No. 1, pp. 250-257, 2007.   DOI
23 A. Waaijer A, M. Prokop, B. K. Velthuis, C. J. Bakker, G. A. De Kort, M. S. Van Leeuwen, "Circle of Willis at CT angiography: dose reduction and image quality-reducing tube voltage and increasing tube current settings, "Radiology, Vol. 242, No. 3, pp. 832-839, 2007.   DOI