Abstract
To evaluate the performance of ultrasound imaging system, we investigated the change of spatial resolution according to changing sonic velocity change parameter provided by ultrasound equipment. Ultrasound phantom images were obtained using a 3.0 ~ 5.0 MHz convex transducer in an ultrasound diagnostic device used at a medical institution located at Iksan. N-365 multi-purpose ultrasound phantom was used to measure longitudinal distance measurement accuracy and longitudinal and transverse resolution. In the same manner, the sonic velocity of the ultrasound equipment was changed from 1580 m/sec to 1400 m/sec in six steps, and the full width at half maximum(FWHM) was measured using the image J program to determine whether the measured values were different. As a result, lateral resolution was measured from 1.91 mm to 5.3 mm according to the speed change, and the smallest FWHM was 1.91 mm at 1420 m/sec. The axial resolution was measured from 1.03 mm to 1.14 mm according to the speed change, and the smallest FWHM was 1.03 mm at 1400 m/sec. The slower the sound velocity of the ultrasound equipment, the shorter the length of longitudinal measurement.
Keywords
Ultrasound equipment; Quality control; Full width at half maximum; Ultrasound phantom;
Citations & Related Records
연도 인용수 순위
  • Reference
1 MM. Goodsitt, PL. Carson, S. Witt, DL. Hykes, JM. Jr. Kofler, "Realtime B-mode ultrasound quality control test procedures: Report of AAPM ultrasound task group No. 1," Medical Physics, Vol. 25, pp. 1385-1406, 1998.   DOI
2 EL. Madsen, "Quality assurance for grey-scale imaging," Ultrasound in Medicine and Biology, 26 Supplement 1:S48-S50, 2000.   DOI
3 P. N. Kim, J. W. Lim, H. C. Kim, Y. C. Yoon, D. J. Sung, M. H. Moon, J. S. Kim, J. C. Kim, "Quality Assessment of Ultrasonographic Equipment Using an ATS-539 Multipurpose Phantom," Journal of the Korean Society of Radiology , Vol. 58, No. 5, pp. 533-541, 2008.   DOI
4 N. J. Dudley, NM. Gibson, MJ. Fleckney, PD. Clark, "The effect of speed of sound in ultrasound test objects on lateral resolution," Ultrasound in Medicine and Biology, Vol. 28, No. 11, pp. 1561-1564, 2002.   DOI
5 M. Mokhtari-Dizaji, "Tissue-mimicking materials for teaching sonographers and evaluation of their specifications after three years," Ultrasound in Medicine and Biology, Vol. 27, No. 12, pp. 1713-1716, 2001.   DOI
6 J. E. Browne, K. V. Ramnarine, A. J. Watson, P. R. Hoskins, "Assessment of the acoustic properties of common tissue-mimicking test phantoms," Ultrasound in Medicine and Biology, Vol. 29, No. 7, pp. 1053-1060, 2003.   DOI
7 JM. Jr. Kofler, "Quality Assurance of Ultrasound Imagers: Procedures, Expectations, and Philosophies," AAPM 43rd Annual Meeting, 2001.
8 P. L. Carson, A. Fenster, "Evolution of ultrasound physics and the role of medical physicists and the AAPM and its journal in that evolution," Medical Physics, Vol. 36, pp. 411-428, 2009.   DOI
9 S. Larson, D. Pfeiffer, Ultrasound QC workshop. AAPM. 2007.
10 F. L. Zheng, QA/QC Ultrasound image. AAPM. 2003.
11 M. Rockville, "American Institute of Ultrasound in Medicine, Quality Assurance Manual for Gray-Scale Ultrasound Scanners," AIUM Publications, 1995.
12 J. E. Browne, A. J. Watson, N. M. Gibson, N. J. Dudley, A. T. Elliott, "Objective measurements of image quality," Ultrasound in Medicine and Biology, Vol. 30, No. 2, pp. 229-237, 2004.   DOI
13 J. Y. Meuwly, "Quality control in ultrasonography," Therapeutische Umschau, Vol. 54, No. 1, pp. 37-43, 1997.
14 S. Y. Chung, H. D. Kim, "Quality Assurance of Ultrasonic Diagnosis in Breast," Journal of Korean Society of Ultrasound in Medicine, Vol. 25, pp. 55-60, 2006.