Evaluation of Image Quality & Absorbed Dose using MCNPX Simulation in the Digital Radiography System  

An, Hyeon (Department of Radiology, Inje University Pusan Paik Hospital)
Lee, Dongyeon (Department of Radiation Oncology, Dongnam Institute of Radiological & Medical Sciences)
Ko, Sungjin (Department of Radiological Science, Catholic University of Pusan)
Kim, Changsoo (Department of Radiological Science, Catholic University of Pusan)
Publication Information
Journal of the Korean Society of Radiology / v.10, no.5, 2016 , pp. 327-335 More about this Journal
Abstract
The study is enforce to study image quality evaluation of condition provide the IEC and combination of clinical conditions each quality of radiation that image quality to assess the conditions provided to IEC in the clinical environment to conduct image quality assessment of the digital radiography system in the detector have environmental limits. First, image quality evaluation was evaluated by measuring the MTF, NPS using four quality of radiation and Using MCNPX simulation lastly DQE make a image quality evaluation after calculating the particle fluence to analyze spectrum quality of radiation. Second, Using MCNPX simulation of four quality of radiation was evaluated absorbed dose rate about electronic 1 per unit air, water, muscle, bone by using Radiation flux density and energy, mass-energy absorption coefficient of matter. Results of evaluation of image quality, MTF of four quality of radiation was satisfied diagnosis frequency domain 1.0 ~ 3.0 lp/mm of general X-ray that indicated 1.13 ~ 2.91 lp/mm spatial frequency. The NPS has added filter, spatial frequency 0.5 lp/mm at standard NPS showed a tendency to decrease after increase. Unused added filter, spatial frequency 0.5 lp/mm at standard NPS showed a certain NPS result value after decrease. DQE in 70 kVp / unuesd added filter(21 mm Al) / SID 150 cm that patial frequency 1.5 lp/mm at standard showed a tendency to decrease after certain value showed. Patial frequency in the rest quality of radiation was showed a tendency to decrease after increase. Results of evaluation of absorbed dose, air < water < muscle < bone in the order showed a tendency to increase. Based on the results of this study provide to basic data that present for the image quality evaluation method of a digital radiation imaging system in various the clinical condition.
Keywords
Digital Radiography System; MTF; NPS; DQE; Particle fluence; Absorbed dose;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 K. Doi, “Diagnostic imaging over the last 50 years: research and development in medical imaging science and technology,” Physics in Medicine and Biology, Vol. 51, No. 13, pp. 5-27, 2006.   DOI
2 D. L.Y. Lee, L. K. Cheung, B. G. Rodricks, G. F. P owell, "Improved imaging Performance of 14 ${\time}$ 17-inch Direct Radiography TM System using Se/TFT detector," Physics of Medical Imaging, Proceedings of SPIE, Vol. 3336, pp. 14-23, 1998.
3 J. P. Moy, "Signal-to-noise ratio and spatial resolution In x-ray electronic imagers: Is the MTF a relevant parameter?," Medical Physics, Vol. 71, No. 1, pp. 86-93, 2000.
4 C. E. Floyd, R. J. Warp, J. T. Dobbins, H. G. Chotas, A. H. Baydush, V. V. Rene, C. E. Ravin, “Imaging Characteristics of an Amorphous Silicon Flat-Panel Detector for Digital Chest Radiography,” Medical Physics, Vol. 218, No. 3, pp. 683-688, 2001.
5 ICRU Report 70, "Radiation Dosimetry, Image Quality in Chest Radiography," International Commission on Radiation Units and Measurements, Bethesda, 2003.
6 J. T. Bushberg, "The Essential Physics Of Medical Imaging, Third Edition," Lippincott Wiliams & Wilkins, pp. 70-95, 2012.
7 H. W. Cho, K. S. Chon, ''Change of MTF for Sampling Interval in Digital Detector," Journal of the Korean Society of Radiology, Vol. 8, No. 5, pp. 224-230, 2014.
8 E. Samei, N. T. Ranger, J. T. Dobbins III, Y. Chen, "Intercomparison of methods for image quality characterization," II. Noise power spectrum, Medical physics, Vol. 33, No. 5, pp. 1466-1475, 2006.   DOI
9 J. T. Dobbins III, D. L. Ergun, L. Rutz, D. A. Hinshaw, H. Blume, D. C. Clark, “DQE (f) of four generations of computed radiography acquisition devices,” Medical Physics, Vol. 22, No. 10, pp. 1581-1593, 1995.   DOI
10 E. Samei, M. J. Flynn, “An experimental comparison of detector performance for direct and indirect digital radiography systems,” Medical Physics, Vol. 30, No. 4, pp. 608-622, 2003.   DOI
11 IEC 62220-1, "Medical electrical equipment-Characteristics of digital X-ray imaging devices Part 1: determination of the detective quantum efficiency," 2003
12 IEC 61267, "Medical diagnostic X-ray equipment Rad iation conditions for use in the determination of Characteristics," 2005
13 E. Samei, N. T. Ranger, J. T. Dobbins III, Y. Chen, “Intercomparison of methods for image quality characterization I. Modulation transfer function,” Medical physics, Vol. 33, No. 5, pp. 1454-1465, 2006.   DOI
14 E. Samei, M. J. Flynn, “An experimental comparison of detector performance for direct and indirect digital radiography systems,” Medical physics, Vol. 30, No. 4, pp. 608-622, 2003.   DOI
15 K. T. Wook, "Performance evaluation of indirect digital radiography system," Department of Medical physics, Graduate School, Korea University, 2010.
16 H. Cember, T. E. Johnson, "Introduction to Health Physics," Fourth Edition 4th Edition, McGraw-Hill Professional Publishing, 182, 2011.
17 H. An, C. S. Kim, J. H. Kim, “Evaluation of Imaging Quality for Radiographic Positioning using IEC Radiation Quality in the Digital Radiography System,” The Korea Contents Association, Vol. 15, No. 7, pp. 289-299, 2015.
18 J. Y. Jung, H. S. Park, H. M. Cho, C. L. Lee, S. R. Nam, Y. J. Lee, H. J. Kim, "Imaging Characteristics of Computed Radiography Systems," Korean journal of medical physics, Vol. 19 No. 1 pp. 63-72, 2008.
19 M. Sandborg, C. A. Carlsson, G. A. Carlsson, “Shaping x-ray spectra with filters in x-ray diagnostic,” Medical & Biological engineering & Computing, Vol. 32, No. 4, pp. 384-390, 1994.   DOI
20 J. M. Boone, J. A. Seibert, “An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV,” Medical Physics, Vol. 24, No. 11, pp. 1661-1670, 1997.   DOI
21 K. S. Cho, "Development of the criteria for the performance evaluation of digital x-ray imager," Korea Advanced Institute of Science and Technology, Report, 2005.