Abstract
Panorama of dental radiation generators can observe the wide anatomical structures of oral and maxillofacial areas but there can be distortion of lengths, angles, or shapes. CBCT can diagnose 3D images and get the ones whose errors by superposition and interference are remarkably smaller between anatomical structures. But the quality of the images by movement of subjects can be lowered as it takes long to diagnose them. And if there are impermeable radiation objects like metal in mouths, impermeable radiation lines can radially appear with the objects as center. This study tries to analyze accuracy of panorama and CBCT and get useful anatomical information in dental treatment by comparing the length of wisdom teeth which were measured by Panorama and CBCT with the teeth which were actually extracted and analyzing distortion of the teeth. The test result could be found that Panorama is expanded by average 7.3% as the errors of Panorama and Digital Vernier Caliper range from 110.7% to 103.9%. The length of wisdom teeth which were measured in CBCT and Digital Vernier Caliper could be found that the error range is 1.3%. And the length of wisdom teeth which were measured in Panorama and Digital Vernier Caliper has found that the error range shows 7.3%. So it could be found that the images of CBCT is about 6% more exact than those of Panorama. It could be found that CBCT shows the more exact images than those of Panorama. But because the examination expenses of CBCT are higher than those of Panorama and exposure dose of CBCT is much more than that of Panorama, it is thought to find proper ways in examination.
Keywords
dental radiation; CBCT; panorama;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 임청환, 김승철, 정홍량 외, "치과 방사선 검사 시 노출 위험성에 따른 피폭선량 방어연구", 한국방사선학회지, Vol. 5, No. 5, pp238-239, 2011.   과학기술학회마을   DOI   ScienceOn
2 정천수, 모은희, 이기흔 외, "CBCT 영상에서 구치부의 하악관 형태에 다른 재구성 방법 연구", 한국방사선학회지, Vol. 5, No. 6, pp.370-374, 2011.   과학기술학회마을   DOI   ScienceOn
3 Stramotas S, Geenty JP, Petocz P, Darendeliler MA. Accuracy of linear a nd angula r measurements on panoramic radiographs taken at various pos itions in vitro. Eur J Orthod Vol, 24, pp.43-52, 2002.   DOI   ScienceOn
4 Choi YS, Hwang HE, Lss SR, "Clinical application of cone beam computed tomography in dental implant", J Korean Dent Assoc, Vol. 44, pp.172-179, 2006.   과학기술학회마을
5 Pinsky HM, Dydal S, Pinsky RW, Misch KA, Sarment DP, "Accuracy of three-dimensional measurements using cone-beam CT", Dentomaxillofac Radiol, Vol. 35, pp.410-416, 2006.   DOI   ScienceOn
6 Mah JK, Danforth RA, Bumann A, Hatcher D. Radiation absorbed in maxillofacial imaging with a new dental computed tomography device. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, Vol. 96, pp. 508-13, 2003.   DOI   ScienceOn
7 Tsiklakis K, Donta C, Gavala S, Karayianni K, Kamenopoulou V, Hourdakis CJ. Dose reduction in maxillofacial imaging using low dose Cone Beam CT. Eur J Radiol, Vol. 56, pp. 413-7, 2005.   DOI   ScienceOn
8 Halazonetis DJ. From 2-dimensional cephalograms to 3-dimensional computed tomography scans. Am J Orthod Dentofacial Orthop, Vol. 127, pp. 627-37, 2005.   DOI   ScienceOn
9 Huang J, Bumann A, Mah J. Three-dimensional radiographic analysis in orthodontics. J Clin Orthod, Vol. 39, pp. 421-8, 2005.
10 치위생과임상연구회, "치과임상 기초", 고문사, pp.216, 2002.
11 Lee JH, Han WJ, Choi YH, Kim EK. "Measurement of maxillary sinus volume and available alveolar bone height using computed tomography", Korean J Oral Maxillofac Radiol, Vol. 33, pp.35-41, 2003.   과학기술학회마을
12 전혜진, "CBCT를 이용한 상악동 중격의 진단 및 파노라마 방사선 사진의 정확도", 이화여자대학교 임상치의학대학원, 석사학위논문, pp.2-3, 2011.
13 Pinsky HM, Dydal S, Pinsky RW, Misch KA, Sarment DP, "Accuracy of three-dimensional measurements using cone-beam CT", Dentomaxillofac Radiol, Vol. 35, pp.410-416, 2006.   DOI   ScienceOn