Analysis of dose reduction of surrounding patients in Portable X-ray  

Choe, Deayeon (Pusan catholic University)
Ko, Seongjin (Pusan catholic University)
Kang, Sesik (Pusan catholic University)
Kim, Changsoo (Pusan catholic University)
Kim, Junghoon (Pusan catholic University)
Kim, Donghyun (Pusan catholic University)
Choe, Seokyoon (Pusan catholic University)
Publication Information
Journal of the Korean Society of Radiology / v.7, no.2, 2013 , pp. 113-120 More about this Journal
Abstract
Nowadays, the medical system towards patients changes into the medical services. As the human rights are improved and the capitalism is enlarged, the rights and needs of patients are gradually increasing. Also, based on this change, several systems in hospitals are revised according to the convenience and needs of patients. Thus, the cases of mobile portable among examinations are getting augmented. Because the number of mobile portable examinations in patient's room, intensive care unit, operating room and recovery room increases, neighboring patients are unnecessarily exposed to radiation so that the examination is legally regulated. Hospitals have to specify that "In case that the examination is taken out of the operating room, emergency room or intensive care units, the portable medical X-ray protective blocks should be set" in accordance with the standards of radiation protective facility in diagnostic radiological system. Some keep this regulation well, but mostly they do not keep. In this study, we shielded around the Collimator where the radiation is detected and then checked the change of dose regarding that of angles in portable tube and collimator before and after shielding. Moreover, we tried to figure out the effects of shielding on dose according to the distance change between patients' beds. As a result, the neighboring areas around the collimator are affected by the shielding. After shielding, the radiation is blocked 20% more than doing nothing. When doing the portable examination, the exposure doses are increased $0^{\circ}C$, $90^{\circ}C$ and $45^{\circ}C$ in order. At the time when the angle is set, the change of doses around the collimator decline after shielding. In addition, the exposure doses related to the distance of beds are less at 1m than 0.5m. In consideration of the shielding effects, putting the beds as far as possible is the best way to block the radiation, which is close to 100%. Next thing is shielding the collimator and its effect is about 20%, and it is more or less 10% by controlling the angles. When taking the portable examination, it is better to keep the patients and guardians far enough away to reduce the exposure doses. However, in case that the bed is fixed and the patient cannot move, it is suggested to shield around the collimator. Furthermore, $90^{\circ}C$ of collimator and tube is recommended. If it is not possible, the examination should be taken at $0^{\circ}C$ and $45^{\circ}C$ is better to be disallowed. The radiation-related workers should be aware of above results, and apply them to themselves in practice. Also, it is recommended to carry out researches and try hard to figure out the ways of reducing the exposure doses and shielding the radiation effectively.
Keywords
Portable; Collimator shielding; tube angle; bed distance;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 ICRP 80. Radiation Dose to patients from Radiopharmaceuticals(국제방사선방호위원회 간행물 53의 추록2) 서문., 1997
2 보건복지부고시 제 4조 6항, 진단용 방사선 발생장치의 안전관리에 관한 규칙 중 방사선 관계종사자의 선량한도, 2006
3 보건복지부고시 제 4조 3항, 진단용 방사선 발생장치의 안전관리에 관한 규칙 중 방사선 방어시설의 검사기준, 2006
4 Dose Ace Inc, Glass Dosimetry manual, 2011
5 Phantom Laboratory, USA, Rando phantom manual, 2009.
6 권덕문, 김성수, 김화곤 등, 의료영상정보학, 대학서림 pp. 15-16, 1996
7 박수성, 김건상, 이관세 등, 진단방사선원리, 대학서림 pp. 142-147, 1998
8 김태관, 노현아, 황민호 등, 투시조영 검사 시 조사야 변화에 따른 선량 및 화질에 관한 연구, 전국 방사선사 춘계학술대회집 pp. 146, 2011
9 정우경, 투시와 중재시술의 방사선 피폭과 저감화 방법, Korean Med Assoc J. Vol. 54, No12, pp. 1269-1276, 2011   DOI   ScienceOn
10 고신관, 강병삼, 임청환, 중재적 방사선 분야 방호용구 차폐효과, 대한방사선기술학회지 pp. 213-219, 2005   과학기술학회마을