A Study on Protection Performance of Radiation Protective Aprons classified by Manufacturers and Lead Equivalent using Over Tube Type Fluoroscopy  

Song, Jong-Nam (Dept. of Radiologic Kwangyang Health College, Technology)
Seol, Gwang-Wook (Dept. of Radiologic Kwangyang Health College, Technology)
Hong, Seong-Il (Dept. of Radiologic Kwangyang Health College, Technology)
Choi, Jeong-Gu (Dept. of Computer Information Kwangyang Health College)
Publication Information
Journal of the Korean Society of Radiology / v.5, no.3, 2011 , pp. 135-141 More about this Journal
Abstract
If protective performance of apron cannot be good, radiation exposure of an guardian or a patient, a person engaged in radiation related industry cannot rise. Therefore, It will be evaluated protection performance to radiation protection aprons by manufacturers and lead equivalent more than 0.25mm lead equivalent. And, will show in the direction of application to clinic. The new aprons by manufacturers(H, X, I, J company) and lead equivalent(0.50mmPb, 0.35mmPb, 0.25mmPb) measured transmitted dose rate and shielding rate, uniformity under fluoroscopy and general radiography using to fluoroscopy system and digital radiography system, x-ray multifunction meter. The shielding rate measurement results, 0.5mmPb apron was Shielding rate of apron of a I company(fluoroscopy : 97.96%) was the best under six companies, and shielding rate of apron of a J company(fluoroscopy : 96.25%) was worst. 0.35mmPb Apron was Shielding rate of a I company(fluoroscopy : 96.79%) was the best under the three companies, and shielding rate of an H company(fluoroscopy : 95.81%) was the worst. 0.25mmPb Apron was Shielding rate of X company apron(fluoroscopy : 90.908%) was better than H company apron(fluoroscopy : 88.82%) than two companies. The uniformity measurement results, 0.5mmPb Aprons of X company(fluoroscopy : 0.13) and I company(fluoroscopy : 0.19) was the best under the six companies, and J company apron(fluoroscopy : 0.45) was the worst. 0.35mmPb. Along a manufacturer and lead equivalent performance of apron protection is distinguished certainly. Therefore, a patient, guardian or a person engaged in radiation related industry shall enforce experiment of a lot of ways defined or evaluation so that the maximum reduces radiation exposure. Buy the apron that protective performance is good, It will be performed through experiment and evaluation.
Keywords
Lead Apron; Lead Equivalent; Shielding Rate; Uniformity;
Citations & Related Records
Times Cited By KSCI : 5  (Citation Analysis)
연도 인용수 순위
1 전자의료기기기준규격 : 식품의약품안전청고시 제2007-83호, 2007.
2 한국산업규격(KS) : X선 방호용품류의 납당량 시험 방법, A 4025, 1990.
3 박명환, 권덕문 : X선 및 감마선에 대한 apron의 차폐율 측정, 방사선기술과학, Vol. 30, No. 3, pp.245-250, 2007.
4 양한준, 전용석, 이근창, 이인자, 이선숙, 허준 : X선방어용 Apron의 성능평가, 대한방사선기술학회지, Vol. 16, No. 2, pp.73-79, 1993.
5 김영근, 장영일, 김정민 : 방사선 방호용 에어프런의 경량화와 차폐능 개선, 대한방사선 기술학회지, Vol. 26, No. 1, pp.45-50, 2003.
6 J.P. McCaffrey, B. Downton, H. Shen : The attenuation effects of lead and non-lead materials used in radiation shielding garments, 2006.
7 유정임, 구정완 : 방사선 방어용 앞치마를 착용하는 방사선 관련 종사자들의 근골격게 증상호소와 관련 요인, 대한산업의학회지, Vol. 16, No. 2, pp.166-177, 2004.
8 강세식, 권달관, 김경근 등 : 방사선 기기학, 청구문화사, 2011.
9 추성실 : 방사선의 의학적 이용에 따른 방사선 안전관리 현황, 대한방사선방어의학회지, Vol. 15, No. 1, pp.39-46, 1990.
10 김유현, 최종학, 김성수 등 : 진단방사선검사에서 환자피폭선량에 관한 연구, 방사선기술 과학, Vol. 28, No. 3, pp.241-248, 2005.
11 임봉식 : 한국에서 방사선 관련 종사자들의 개인피폭선량 실태에 관한 연구, 방사선기술 과학, Vol. 29, No,. 3, pp.185-195, 2006.
12 보건복지부 : 진단용방사선발생장치의 안전관리에 관한 규칙, 보건복지부령 제349호, 2006.
13 J.P. McCaffrey, H. Shen, B. Downton, E. Mainegra-Hing : Radiation attenuation by lead and nonlead materials used in radiation shielding garments, Medical Physics, Vol. 34, No. 2, pp.530-537, 2007.   DOI   ScienceOn