Treatment Planning Guideline of EBT Film-based Delivery Quality Assurance Using Statistical Process Control in Helical Tomotherapy  

Chang, Kyung Hwan (Department of Radiological Science, Far East University)
Publication Information
Journal of radiological science and technology / v.45, no.5, 2022 , pp. 439-448 More about this Journal
Abstract
The purpose of this study was to analyze the results from statistical process control (SPC) to recommend upper and lower control limits for planning parameters based on delivery quality assurance (DQA) results and establish our institutional guidelines regarding planning parameters for helical tomotherapy (HT). A total of 53 brain, 41 head and neck (H & N), and 51 pelvis cases who had passing or failing DQA measurements were selected. The absolute point dose difference (DD) and the global gamma passing rate (GPR) for all patients were analyzed. Control charts were used to evaluate upper and lower control limits (UCL and LCL) for all assessed treatment planning parameters. Treatment planning parameters were analyzed to provide its range for DQA pass cases. We confirmed that the probability of DQA failure was higher when the proportion of leaf open time (LOT) below 100 ms was greater than 30%. LOT and gantry period (GP) were significant predictor for DQA failure using the SPC method. We investigated the availability of the SPC statistic method to establish the local planning guideline based on DQA results for HT system. The guideline of each planning parameter in HT may assist in the prediction of DQA failure using the SPC statistic method in the future.
Keywords
Statistical Process Control; Delivery Quality Assurance; Helical Tomotherapy; Planning Guideline; Leaf Open Time;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Chang KH, Ji Y, Kwak J, Kim SW, Jeong C, Cho B, et al. Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations. Prog Med Phys. 2016;27(3):111-6.   DOI
2 Ham IS, Cho PK, Jung KK. Dosimetric Comparison of Setup Errors in Intensity Modulated Radiation Therapy with Deep Inspiration Breath Holding in Breast Cancer Radiation Therapy. Korean Society of Radiological Science. 2019;42(2):137-43.
3 Thiyagarajan R, Nambiraj A, Sinha SN, Yadav G, Kumar A, Subramani V, et al. Analyzing the performance of ArcCHECK diode array detector for VMAT plan. Reports of Practical Oncology & Radiotherapy. 2016;21(1):50-6.   DOI
4 Chen M, Chen Y, Chen Q, Lu W. Theoretical analysis of the thread effect in helical TomoTherapy. Medical Physics. 2011;38(11):5945-60.   DOI
5 Dong P, Lee P, Ruan D, Long T, Romeijn E, Low DA, et al. 4π noncoplanar stereotactic body radiation therapy for centrally located or larger lung tumors. Int J Radiat Oncol Biol Phys. 2013;86(3): 407-13.   DOI
6 Montgomery DC. Statistical quality control. New York: Wiley; 2009.
7 Prins J. Process or product monitoring and control. NIST/SEMATECH e-Handbook of Statistical Methods: National Institute of Standards & Technology, Gaithersburg; 2007.
8 Shimizu H, Sasaki K, Tachibana H, Tomita N, Makita C, Nakashima K, et al. Analysis of modulation factor to shorten the delivery time in helical tomotherapy. J Appl Clin Med Phys. 2017;18(3): 83-7.   DOI
9 Thomas SJ, Geater AR. Implications of leaf fluence opening factors on transfer of plans between matched helical tomotherapy machines. Biomedical Physics & Engineering Express. 2017;4(1):017001.   DOI
10 Vaden R, Schauer N. Efficient planning for efficient treatments on tomotherapy. ASTRO; 2016.
11 Binny D, Lancaster CM, Trapp JV, Crowe SB. Statistical process control and verifying positional accuracy of a cobra motion couch using step-wedge quality assurance tool. J Appl Clin Med Phys. 2017;18(5):70-9.   DOI
12 Binny D, Aland T, Archibald-Heeren BR, Trapp JV, Kairn T, Crowe SB. A multi-institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control. J Appl Clin Med Phys. 2019;20(3):71-80.   DOI
13 Martisikova M, Ackermann B, Jakel O. Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams. Phys Med Biol. 2008;53(24): 7013-27.   DOI
14 Mezzenga E, D'Errico V, Sarnelli A, Strigari L, Menghi E, Marcocci F, et al. Preliminary Retrospective Analysis of Daily Tomotherapy Output Constancy Checks Using Statistical Process Control. PLoS One. 2016;11(2):e0147936.   DOI
15 Meyers SM, Balderson MJ, Letourneau D. Evaluation of Elekta Agility multi-leaf collimator performance using statistical process control tools. J Appl Clin Med Phys. 2019;20(7):100-8.   DOI
16 Westerly DC, Soisson E, Chen Q, Woch K, Schubert L, Olivera G, et al. Treatment planning to improve delivery accuracy and patient throughput in helical tomotherapy. Int J Radiat Oncol Biol Phys. 2009; 74(4):1290-7.   DOI
17 Chang KH, Lee YH, Park BH, Han MC, Kim J, et al. Statistical Analysis of Treatment Planning Parameters for Prediction of Delivery Quality Assurance Failure for Helical Tomotherapy. Technology in Cancer Research & Treatment. 2020;19:1-9.
18 Ko YS. The construction methodology of a reul-based expert system using CART-based decision tree method. The Journal of the Korea Institute of Electronic Communication Sciences. 2011;6(6):849-54.   DOI
19 Sanghangthum T, Suriyapee S, Srisatit S, Pawlicki T. Statistical process control analysis for patient-specific IMRT and VMAT QA. Journal of Radiation Research. 2013;54(3):546-52.   DOI
20 TomoTherapy® Treatment System StatRTTM Guide. TomoTherapy Incorporated; 2010(177)
21 Chang KH, Kim DW, Choi JH, Shin HB, Hong CS, Jung DM, et al. Dosimetric Comparison of Four Commercial Patient-Specific Quality Assurance Devices for Helical Tomotherapy. Journal of the Korean Physical Society. 2020;76(3):257-63.   DOI
22 Boyd R, Jeong K, Tome WA. Determining efficient helical IMRT modulation factor from the MLC leaf-open time distribution on precision treatment planning system. J Appl Clin Med Phys. 2019;20(5): 64-74.   DOI
23 Kim DG, Jung JH. Evaluating the Usefulness of Rice Bolus Phantom in Tomotherapy: Phantom Study. Korean Society of Radiological Science. 2021;44(6): 663-9.
24 Chang KH, Lee S, Jung H, Choo YW, Cao YJ, Shim JB, et al. Development of a 3D optical scanner for evaluating patient-specific dose distributions. Physica Medica. 2015;31(5):553-9.   DOI
25 Guckenberger M, Meyer J, Wilbert J, Baier K, Bratengeier K, Vordermark D, et al. Precision required for dose-escalated treatment of spinal metastases and implications for image-guided radiation therapy (IGRT). Radiother Oncol. 2007;84(1):56-63.   DOI
26 Breen SL, Moseley DJ, Zhang B, Sharpe MB. Statistical process control for IMRT dosimetric verification. Med Phys. 2008;35(10):4417-25.   DOI
27 Skorska M, Piotrowski T. Optimization of treatment planning parameters used in tomotherapy for prostate cancer patients. Phys Med. 2013;29(3): 273-85.   DOI
28 Binny D, Lancaster CM, Byrne M, Kairn T, Trapp JV, Crowe SB. Tomotherapy treatment site specific planning using statistical process control. Phys Med. 2018;53:32-9.   DOI
29 Zhao YL, Mackenzie M, Kirkby C, Fallone BG. Monte Carlo calculation of helical tomotherapy dose delivery. Med Phys. 2008;35(8):3491-500.   DOI