A Study on Distortion and Dose of Images in Mobile Radiography  

Song, Hyeon-Seok (Department of Radiological Science, Hanseo University)
Lim, Cheong-Hwan (Department of Radiological Science, Hanseo University)
Jung, Hong-Ryang (Department of Radiological Science, Hanseo University)
Kim, Jong-Seong (Department of Radiology, Asan Medical Center)
Kim, Yeong-Ran (Department of Radiology, Asan Medical Center)
Jeong, Sung-Hun (Department of Radiological Science, Gimcheon University)
Publication Information
Journal of radiological science and technology / v.45, no.4, 2022 , pp. 305-312 More about this Journal
Abstract
The proportion and testing of portable radiation tests, which are limited at the request of the doctor, are gradually increasing only for patients in emergency situations and difficulties in moving. However, as there are many limiting factors compared to fixed devices, this study intends to measure and analyze the distortion of images according to the angle of the detector and the change in dose according to the position of the subject. For distortion experiments using a mobile radiation generator used in Hospital A, the SID was tilted by 110 cm, 14"×17" wireless FPD detector to 0°, -5°, -10°, -15°, -20°, and -35° to measure the change in area. The dose according to the location of the detector was analyzed on average by measuring the central dose at 110 cm of the SID and measuring the dose of 9 locations three times each. The analysis result of distortion by location according to the angle of the detector showed a statistically significant difference (f=58.74, p<0.000). Therefore, it can be seen that the angle of the detector and the tube is closely related to the distortion of the image. The difference in dose by location of the detector increased with respect to the center - pole, and decreased with the + pole. Tests using mobile radiation generators will require careful efforts by clinicians to position patients in the center of the detector for accurate diagnosis, and efforts will be made to level the angle between the mobile radiation generators and the detector.
Keywords
Mobile radiography; Detector; Angle; Image Distortion; Dose;
Citations & Related Records
Times Cited By KSCI : 5  (Citation Analysis)
연도 인용수 순위
1 Jacobi A, Chung M, Bernheim A, Eber C. Portable chest X-ray in coronavirus disease-19 (COVID-19): A pictorial review. 2020 Aug;64:35-42. DOI: https://doi.org/10.1016/j.clinimag.2020.04.001   DOI
2 Lee JS, Park HH. The Image Distortion Analysis of Levin-tube tip by Patient position and Incidence Angle when taking Mobile Chest AP Projection. Journal of the Korean Society of Radiology. 2015;9(7):467-71.   DOI
3 http://opendata.hira.or.kr/op/opc/olapDiagBhvIn fo.do-Item: Radiographic simple image (HIRA).
4 http://opendata.hira.or.kr/op/opc/olapDiagBhvInfo.do-Item: G2101 Chest [direct] 1 image (HIRA).
5 Im JK, Lee KS. Thoracic Radiology Science. Ilchokak Publishing; 2000:29-34.
6 Kang SS, Kwak BJ, Kwon SM, Kim NS, Kim SH, Kim CS, et al. Textbook of Radiographic Positioning and Clinical Diagnosis. 6th ed. Chung-Ku Publisher; 2019:670-726.
7 Joo YC, Lim CH, You IG, Jung HR, Lee SH. Adequacy of Source to Image Receptor Distance with Chest Postero-Anterior Projection in Digital Radiology System. Journal of Radiological Science and Technology. 2016;39(2):135-42.   DOI
8 Choi JG. Analysis of factors that affect dose increase and image deterioration in digital radiography [dissertation]. Dankook University; 2008.
9 Yoo YE, Lim CH, Ko JY. An Analysis of Factors That Affect Image Quality Deterioration in The Potable X-ray Examination on using Digital Wireless Detector. Journal of Radiological Science and Technology. 2014;37(2):93-100.
10 Shin JB, Kim SO, Kim SR. Study on the Distortion Factor of the Images of the Stitching Method Depending on the Angle of X-Ray Tube Using Phantom. Journal of Korean Society of Radiological Imaging Technology. 2017;14(1):35-45.
11 Cho PK. How to Improve Image Quality for the Chest PA and the Simple Abdomen X-ray Examinations. Journal of the Korean Society of Radiology. 2013;7(3):165-73.   DOI
12 Kim BG, Kang SI, Park SM, Lee MW, Kim SB. Assessment of Exposure Index based on Changes of Tube Angle in Portable X-ray Examination using Digital Wireless Detector. Journal of the Korean Society of Radiology. 2016;1:39-47.
13 Choi DY, Ko SJ, Kang SS, et al. Analysis of dose reduction of surrounding patients in Portable X-ray. Journal of the Korean Society of Radiology. 2013;7(2):113-20.   DOI
14 Kim JY, et al. The First Case of 2019 Novel Coronavirus Pneumonia Imported into Korea from Wuhan, China: Implication for Infection Prevention and Control Measures. Journal Korean Medical Science. 2020;35(5):e61.
15 Korea Disease Control and Prevention Agency. Coronavirus Infectious Disease-19 outbreak status in Korea; 2022. http://ncov.mohw.go.kr/bdBoardList_Real.do
16 De Moura J, Garcia LR, Vidal PFL, et al. Deep Convolutional Approaches for the Analysis of COVID-19 Using Chest X-Ray Images From Portable Devices. IEEE Access. 2020;8:195594-607.   DOI