Abstract
Because moire artifacts overlap with fine lesions and frequency bands, image processing software that removes moire artifacts can lead to loss of micro lesions. In this study, microscopic lesions such as microcalcification and microfracture were randomly formed on SD (Sprague Dawley) rats and image and optimized grid images were compared and analyzed using reference image and grid suppression software. The images were read by two consecutive radiologists using a McNemar's test. Among the 73 microcalcifications, in the 13 cases after grid suppression, the image of the optimized grid shows the loss of image in 3 cases, and the image after grid suppression shows statistically significant image loss (p=0.021). In all 53 fracture lines, there were 19 cases of image loss after the grid suppression, and only one case of the optimized grid showed no image loss. Therefore, the use of grid suppression software to remove moire artifacts should be carefully considered in the diagnosis of micro lesions.
Keywords
DR image; moire artifact; image processing software; grid suppression; optimized grid;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 SH Jeung, BH Han, HR Jung. Evaluation of Image Quality When Using Grid During Child Chest X-Ray Examination. Journal of Radiological Science and Technology, 2017;40(3):371-376   DOI
2 M. A. Thomas, A. H. Rowberg, S. G. Langer, Y. Kim, Iterative image enhancement of CR and DR images. Journal Digital Imaging, Sep. 2004; 17(3):189-195.   DOI
3 J. Wang and H. K. Huang, Film digitization aliasing artifacts caused by grid line patterns. IEEE Trans. Medical Imaging, June 1994; 13(2):375-385   DOI
4 D.S. Kim, S. Lee. Analysis on the Saturation of Grid Artifact and its Reduction in Digital Radiography Images Based on the Adaptive Filtering. Journal of the institute of electronics engineers of Korea, 2011;48(4),1-11
5 D.S. Kim. Grid Angle Optimization and Grid Artifact Reduction in Digital Radiography Images Based on the Modulation Model. Journal of the institute of electronics engineers of Korea, 2011;48(3):30-41.
6 HM Lee, J Yoon, HJ Kim. Effects of Contrast Improvement on High Voltage Rectification Type of X-ray Diagnostic Apparatus, Journal of Radiological Science and Technology, 2014;37(3):187-193
7 D.S. Kim, S. Lee. Grid artifact reduction for direct digital radiography detectors based on rotated stationary grids with homomorphic filtering. Med. Phys, 2013; 40(6):061905.   DOI
8 Motoaki Kato, Sachie Nishimura, Takahide Okamoto, et al. Clinical efficacy of image processing of grid detection and suppression (GDS) in computed radiography". Nihon Hoshasen Gijutsu Gakkai Zasshi. 2005; 61(8):1158-1169.   DOI
9 Kim, Dong-Sik. Artifact Reduction in Digital Radiography Images with the Stationary Grid Based on 1-Dimensional Filters. Journal of the Institute of Electronics Engineers of Korea SP, 2010; 47(5):117-126.
10 Ryoji Sasada, Masahiko Yamada, Shoji Hara, Hideya Takeo, Kazuo Shimura. Stationary grid pattern removal using 2D technique for moire-free radiographic image display, Proceedings Volume 5029, Medical Imaging 2003: Visualization, Image-Guided Procedures, and Display.
11 Chih-Yang Lin, Wen-Jeng Lee, Shyh-Jye Chen, et. al. A Study of Grid Artifacts Formation and Elimination in Computed Radiographic Images. Journal of Digital Imaging. 2006;19(4):351-361.   DOI
12 E. H. Mo, S .H. Park, S.H. Lee, et al. Loss of Fine Lesions after Removing Moire Artifacts in the DR Image, Journal of Convergence Information Technology(JCIT), 2016; 11(5):64-73
13 Carlos Ubeda, Eliseo Vano, Luciano Gonzalez, Patricia. Miranda. Influence of the antiscatter grid on dose and image quality in pediatric interventional cardiology X-ray systems. Catheterization and Cardiovascular Interventions, 2013; 82(1): 51-57.   DOI
14 Fritz, Shannon, and A. Kyle Jones. Guidelines for anti-scatter grid use in pediatric digital radiography. Pediatric radiology, 2014; 44(3):313-321.   DOI
15 Tapiovaara, Markku J., Michael Sandborg, and David R. Dance. A search for improved technique factors in pediatric fluoroscopy. Physics in medicine and biology, 1999; 44(2):537-559   DOI