A Study of Scattered Radiation Effect on Digital Radiography Imaging System  

Baek, Cheol-Ha (Department of Radiological Science, Dongseo University)
Publication Information
Journal of radiological science and technology / v.40, no.1, 2017 , pp. 71-78 More about this Journal
Abstract
Scattered radiation is inherent phenomenon of x-ray, which occurs to the subject (or patient). Therefore it cannot be avoidable but also interacts as serious noise factor because the only meaningful information on x-ray radiography is primary x-ray photons. The purpose of this study was to quantify scattered radiation for various shooting parameters and to verify the effect of anti-scatter grid. We employed beam stopper method to characterize scatter to primary ratio. To evaluate effect on the projection images calculated contrast to noise ratio of given shooting parameters. From the experiments, we identified the scattered radiation increases in thicker patient and smaller air gap. Moreover, scattered radiation degraded contrast to noise ratio of the projection images. We find out that the anti-scatter grid rejected scattered radiation effectively, however there were not fewer than 100% of scatter to primary ratio in some shooting parameters. The results demonstrate that the scattered radiation was serious problem of medical x-ray system, we confirmed that the scattered radiation was not considerable factor of dig ital radiog raphy.
Keywords
Digital radiography; Scatter rays; Contrast to noise ratio; Scatter to primary ratio;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Steven W. Smith, Robert A. Kruger, A signal processing model of diagnostic x-ray scatter, Medical Physics, 13(6), 831-835, 1986   DOI
2 P. C. Johns, M. Yaffe, Scattered radiation in fan beam imaging systems, Medical Physics, 9(2), 231-239, 1982   DOI
3 Min-Cheol Jeon, Hyun-Soo Lim, Man-Seok Han, The Study About Attenuation of Scatter Ray According to Distance Inverse Square Law at General Projection, Journal of Korean Society of Radiological Technology, 34(3), 183-188, 2011
4 H Khodjou-Chokami, M Sohrabpour, Design of linear anti-scatter grid geometry with optimum performance for screen-film and digital mammography systems, Physics in Medicine and Biology, 60(15), 5753-5765, 2015   DOI
5 Fritz. Shannon, Jones. A. Kyle, Guidelines for antiscatter grid use in pediatric digital radiography, Pediatric Radiology, 44(3), 313-321, 2014   DOI
6 Siewerdsen. J. H, Daly. M. J, Jaffray. D. A. et al, A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT, Medical Physics, 33(1), 187-197, 2006   DOI
7 Kentaro Naka, Aya Saito, Junji Morshita et al, Investigation of optimum anti-scatter grid selcetion for digital radiography: physical imaging properties and detectability of low-contrast signals, Radiological Physics and Technology, 6(1), 54-60, 2013   DOI
8 http://atom.kaeri.re.kr/cgi-bin/w3xcom.
9 N. Oberhofer, G. Compagnone, E. Moroder, Use of CNR as a Metric for Optimization in Digital Radiology, World congress on medical physics and biomedical enginnering, Germany, 25(2), 296-299, 2009
10 Hoo-Min Lee, Joon Yoon, Hyun-Ju Kim, Effects of Contrast Improvement on High Voltage Rectification Type of X-ray Diagnostic Apparatus, Journal of Korean Society of Radiological Technology, 34(3), 183-188, 2011
11 Ulrich Neitzel, "Grids or air gaps for scatter reduction in digital radiography: A model calculation," Medical Physics. 12(2), 475-481, 1985
12 Heang-Ping Chan, Kunio Doi, "Physical characteristics of scattered radiation in diagnostic radiology: Monte Carlo simulation studies," Medical Physics. 19(2), 152-165, 1992
13 John M. Boone, J. Anthony Seibert, Cha-Mei Tang, Steve M. Lane, Grid and slot scan scatter reduction in mammography: comparison by using Monte Carlo techniques. Radiology, 222(2), 519-527, 2002   DOI