Abstract
The study investigates the necessity of 3 dimensional dose distribution evaluation instead of point dose and 2 dimensional dose distribution evaluation. Treatment plans were generated on the RANDO phantom to measure the precise dose distribution of the treatment site 0.5, 1, 1.5, 2, 2.5, 3 cm with the prescribed dose; 1,200 cGy, 5 fractions. Gamma analysis (3%/3 mm, 2%/2 mm) of dose distribution was evaluated with gafchromic EBT2 film and ArcCHECK phantom. The average error of absolute dose was measured at $0.76{\pm}0.59%$ and $1.37{\pm}0.76%$ in cheese phantom and ArcCHECK phantom respectively. The average passing ratio for 3%/3 mm were $97.72{\pm}0.02%$ and $99.26{\pm}0.01%$ in gafchromic EBT2 film and ArcCHECK phantom respectively. The average passing ratio for 2%/2 mm were $94.21{\pm}0.02%$ and $93.02{\pm}0.01%$ in gafchromic EBT2 film and ArcCHECK phantom respectively. There was a more accurate dose distribution of 3D volume phantom than cheese phantom in patients DQA using tomotherapy. Therefor it should be evaluated simultaneously 3 dimensional dose evaluation on target and peripheral area in rotational radiotherapy such as tomotherapy.
Keywords
DQA; tomotherapy; 3D volume phantom; 3D dose evaluation;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Mack A., Scheib S.G., Major J., et al.: Precision dosimetry for narrow photon beams used in radiosurgery Determination of Gamma Knife output factors, Medical Physics, 29(9), 2080-2089, 2002   DOI
2 Choi T.J., Kim O.B.: Evaluation of the output dose of a linear accelerator photon beams by using the ionization chamber TM31010 series through TG-51 protocol to postal monitoring output of RPC for 5 years, Korean Journal of Medical Physics, 22(2), 92-98, 2011
3 Welsh J.S., Lock M., Harari P.M., et al.: Clinical implementation of adaptive helical tomotherapy: a unique approach to image-guided intensity modulated radiotherapy, Technology in cancer research & treatment, 5(5), 465-480, 2006   DOI
4 Cadman P., McNutt T., Bzdusek K.: Validation of physics improvements for IMRT with a commercial treatment-planning system, Journal of Applied Clinical Medical Physics, 6(2), 74-86, 2005   DOI
5 Mu G., Ludlum E., Xia P.: Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer, Physics in medicine & biology, 53(1), 77-88, 2008   DOI
6 Bayouth J. E., Wendt D., Morrill S. M.: MLC quality assurance techniques for IMRT applications, Medical Physics, 30(5), 743-750, 2003   DOI
7 LoSasso T., Chui C. S., Ling C. C.: Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode, Medical Physics, 28(11), 2209-2219, 2001   DOI
8 Ju S. G., Ahn Y. C., Huh S. J., Yeo I. J.: Film dosimetry for intensity modulated radiation therapy: dosimetric evaluation, Medical Physics, 29(3), 351-355, 2002   DOI
9 Herzen J., Todorovic M., Cremers F., et al.: Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine, Physics in medicine and biology, 52(4), 1197-1208, 2007   DOI
10 Daniel L, Misbah G, Di Yan, et al.: Evaluation of a 2D diode array for IMRT assurance. Radiation & Oncology, 70(1), 199-206, 2004   DOI
11 Young-Taek Oh, Haejin Kang, Miwha Kim et al.: Three-dimensional dosimetry using magnetic resonance imaging of polymer gel, The Journal of the Korean society for therapeutic radiology and oncology, 20(3), 264-273, 2002
12 Bedford J.L., Lee Y.K., Wai P., South C.P., Warrington A.P.: Evaluation of the Delta4 phantom for IMRT and VMAT verification, Physics in medicine and biology, 54(9), 167-176, 2009   DOI
13 Van Esch A., Clermont C., Devillers M., et al.: On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom, Medical Physics, 34(10), 3825-3837, 2007   DOI
14 H. Fakir, S. Gaede, M. Mulligan, J.Z. Chen.: Development of a novel ArcCHECK insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities, Medical Physics., 39(7), 4203-4208, 2012   DOI
15 Welsh J.S., Lock M., Harari P.M., Tome W.A., et al.: Clinical implementation of adaptive helical tomotherapy: a unique approach to image-guided intensity modulated radiotherapy, Technology in cancer research & treatment, 5(5), 465-480, 2006   DOI