Abstract
The purpose of this study is to evaluate the accuracy of beam delivery QA software using the MLC dynalog file, about the VMAT plan with AAPM TG-119 protocol. The Clinac iX with a built-in 120 MLC was used to acquire the MLC dynalog file be imported in MobiusFx(MFX). To establish VMAT plan, Oncentra RTP system was used target and organ structures were contoured in Im'RT phantom. For evaluation of dose distribution was evaluated by using gamma index, and the point dose was evaluated by using the CC13 ion chamber in Im'RT phantom. For the evaluation of point dose, the mean of relative error between measured and calculated value was $1.41{\pm}0.92%$(Target) and $0.89{\pm}0.86%$(OAR), the confidence limit were 3.21(96.79%, Target) and 2.58(97.42%, OAR). For the evaluation of dose distribution, in case of $Delta^{4PT}$, the average percentage of passing rate were $99.78{\pm}0.2%$(3%/3 mm), $96.86{\pm}1.76%$(2%/2 mm). In case of MFX, the average percentage of passing rate were $99.90{\pm}0.14%$(3%/3 mm), $97.98{\pm}1.97%$(2%/2 mm), the confidence limits(CL) were in case of $Delta^{4PT}$ 0.62(99.38%, 3%/3 mm), 6.6(93.4%, 2%/2 mm), in case of MFX, 0.38(99.62%, 3%/3 mm), 5.88(94.12%, 2%/2 mm). In this study, we performed VMAT QA method using dynamic MLC log file compare to binary diode array chamber. All analyzed results were satisfied with acceptance criteria based on TG-119 protocol.
Keywords
VMAT; MobiusFx(MFX); Dynalog file; Gamma index; $Delta^{4PT}$;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Chen Q, Chen M, Lu W: Ultrafast convolution/superposition using tabulated and exponential kernels on GPU. Med Phys, 38(3), 1150-1161, 2010   DOI
2 Oncentra Master Plan physics and algorithm documentation, Nucletron, Netherlands, 2011
3 MobiusFX ONE MINUTE IMRT QA, MOBIUS MEDICAL SYSTEM, Houston, TX, 2015
4 Chen Q, Chen M, Lu W: Ultra fast convolution/ superposition using tabulated and exponential kernels on GPU. Med Phys, 38, 1150-1161, 2010
5 $Delta^{4PT}Getting$ started, Scandidos, Sweden, 2010
6 Low DA, Dempsey JF: Evaluation of the gamma dose distribution comparison method. Med Phys, 30(9), 2455-2464, 2003   DOI
7 Francisco CH, Silvia VC: A Probability approach to the study on uncertainty effects on gamma index evaluations in radiation therapy. Computational and Mathematical Methods in Medicine, 2011(2011), 10, 2011
8 The Patient Specific QA of IMRT and VMAT Through tha AAPM Task Group Report 119. Journal of radiological science and technology, 35(3), 255-263, 2012
9 R. R. Reynolds, H. Johnston, A. Pompos, et al: INITIAL EXPERIENCE WITH VMAT PLAN AND DELIVERY VERIFICATION USING A DICOM-RT FRAMEWORK AND LINAC DELIVERY LOG FILES. AAPM 2014 Poster, 2014
10 Masi L, Casamassima F, Doro R, Francescon P: Quality assurance of volumetric modulated arc therapy: evaluation and comparison of different dosimetric systems. Med Phys, 38(2), 612-621, 2011   DOI
11 Ezzell GA, Burmeister JW, Dogan N, et al.: IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys, 36(11), 5359-5373, 2009   DOI
12 Young yih Han: Review on the Quality Assuarance for Intensity Modulated Radiation Therapy. PROGRESS in MEDICAL PHYSICS, 24(4), 213-219, 2013   DOI
13 Quan J, Lee L, Liu W, et al.: Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log files. Phys Med Biol, 55, 3597-3610, 2010   DOI