Adequacy of Source to Image Receptor Distance with Chest Postero-Anterior Projection in Digital Radiology System  

Joo, Young-Cheol (Dept. of Radiology, Samsung Medical Center)
Lim, Cheong-Hwan (Dept. of Radiology, Science, Hanseo University)
You, In-Gyu (Dept. of Radiology, Hallym University Hospital)
Jung, Hong-Ryang (Dept. of Radiology, Science, Hanseo University)
Lee, Sang-Ho (Dept. of Radiology, Science, Seonam University)
Publication Information
Journal of radiological science and technology / v.39, no.2, 2016 , pp. 135-142 More about this Journal
Abstract
The purpose of this study is to evaluate propriety of using SID 180cm at Chest PA examination and to find effect of geometrical cause to the image. XGEO-GC80, INNOVISION-SH, CXDI-40EG detector and a chest phantom designed self-production was used for this study. Images were acquired at SID 180cm with changing the factor OID as 0, 75 and 83mm and were analyzed by Centricity Radiography RA1000 PACS system. Statistical program was used the SPSS (Version 22.0, SPSS, Chicago, IL, USA), p-value(under 0.05) was considered to be statistically significant. In OID 0 mm was enlarged about 2.7~3.5 mm than the actual degree of the HS, BS of phantom in all equipments. Compared with the calculated magnification has been expanded 1.6~2.8% when viewed. The OID 75 mm with OID 83 mm was extended from the CS and BS 6~8 mm range. Compared to the calculated values, the measured values are expanded from 6.1 to 7.9%. CS and BS according to the OID change showed a statistically significant difference (p<0.05) among each group, the post-analysis only OID 0 mm group appeared as an independent group, 75 mm and 83 mm are separated in the same group It was. But had no statistically significant difference could change depending on the OID (p>0.05), post-mortem analysis showed, both in the same group. Heart sizes appears larger than actual size 6~8 mm at chest PA examination which is enlarged 6.1~7.9% more than the actual theoretical value. We can find magnification of the image because of the increase of the OID due to technical limitations between cover of standing detector and the image plate. so we suggest to have occurred between them when considering the need to adjust the equipment installed by the SID to match the characteristics of the equipment.
Keywords
Chest PA; CS; CTR;
Citations & Related Records
Times Cited By KSCI : 5  (Citation Analysis)
연도 인용수 순위
1 Van De Graaff: Human Anatomy, 6th ed, McGraw-Hill, 592, 2004
2 Daehun Kim, Seongjin Ko, Sesik Kang, Junghoon Kim, Changsoo Kim: Computer-Aided Diagnosis for Pulmonary Tuberclusis using Texture Features Analysis in Digital Chest Radiography, JKCA, 11(11), 187-193, 2011
3 J.H. Wouter, Veldhamp, J. M. Lucia: Dose and perceived image qualigy in chest radiography, Eropaean journaral of Radiologry, 72(2), 209-217, 2009
4 Boram Park, Dongwook Sung: A Comparative Study of Image Quality and Radiation Dose with Changes in Tube Voltage and Current for a Digital Chest Radiogray, J Korean Soc Radiol, 62, 131-137, 2010   DOI
5 Eunyoung Kim, Jueun Lim, Byunghoon Park et al.: The Importance of the Simple Chest X-ray Tend to Make Careless Interpretation: with a Case of Mediasinal Tumor, Tuberculosis and Respiratory Dis, 66(6), 482-485, 2009   DOI
6 Inja Lee, Changnam Kim, Changyeob Lee, Kyeyeon Park: Evaluation of Image According to Exposure Conditions using Contrast-Detail Phantion for Chest Digital Radiography, Korean Society of Radiological Science, 32(1), 25-32, 2009
7 Heesugn Ahn: Digital Radiography, Journal of the Korea Society for Nondestructive Testing, 33(1), 80-95, 2013
8 Marietta Garmer, Svenja P.Hennings, Horst J. Jager, et al.: Digital Radiograohy Versus Conventional Radiography in Chest imaging: Diagnostic Performance of a Large-Area Silicon Flat-Panel Detector in a Clincal CT-Controlled Study", AJR, 74(1), 75-80, 2000
9 Yangproducibility between conventional lateral cephalometric radiography and digital lateral cephalometric radiography, KOREA J ORTHOD, 32(2), 79-89, 2002
10 Byeongju An, A comparative study for resolution and density of chest imaging using film/screen, CR and DR, The Korean Society Radiology, 4(1), 25-30, 2010
11 Woonggi Jeon, Taeyun Kim, Sungjun Kim, Heungkuk Choi, Kwanggi Kim: Lung Segmentation Considering Global and Local Properties in Chest X-ray Images, Journal of Korea Multimedeia Society, 16(7), 829-840, 2013   DOI
12 www.pinterest.com
13 http://en.wikipedia.org/wiki/X-ray
14 Hamgyum Kim: The Finding on Cardiothracic Ratio in Simple Chest Radiography, Korean Society of Radiological Science, 27(4), 43-48, 2004
15 Seoghee Park, Yungil Lee, Jongwoo Kim: Comparative study in cardiac measurement between standard PA and supine AP views of the chest, J Korean Radil Soc, 16(2), 438-442, 1980   DOI
16 Soonmu Kwon, Changhee Park, Jeongkyu Park, Woonheung Song, Jaeeun Jung: The Effect of Source to Image-Receptor Distance(SID) on Radiation Dose for Digital Chest Radiography, J. Korean Soc. Radiol, 8(4), 203-210, 2014   DOI