The purpose of this study is computed tomography contrast agent at low concentrations and low tube voltage technique to evaluate the usefulness on the phantom image. By varying the degree of mixture by the contrast medium concentration it was inserted in phantom. It was taken by changing the tube voltage and tube current step by step, and to evaluate the dose and the CT value obtained from the phantom image. As a result, low-contrast, low tube voltage(300 mgI/ml, 100 kV) was reduced by an average 21%(CTDIvol; computed tomography dose indexvol) more standard condition(350 mgI/ml, 120 kV). SNR was increased at all depths of the phantom, respectively 1:10 and 1:20(by diluting a contrast agent and normal saline) 12.2(26%) 6.2(17%). CNR was increased at all depths of the phantom, respectively 1:10 and 1:20(by diluting a contrast agent and normal saline) 11.5(32%), 6.3(26%). Research work on the CT scan is necessary in a variety of studies on the low contrast concentration and low tube voltage techniques for dose reduction and reducing of side effects the contrast agent.
Keywords
Low contrast concentration; Low tube voltage; Attenuation coefficient;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Coles DR, Small MA, Negus IS: Computed Tomography Comparison of radiation doses from multislice Computed Tomography coronary and conventional diagnostic angiography. J Am Coll Cadiol, 47, 1840-1845, 2006   DOI
2 JH Kim, JJ Chung, KH Kim: Reduction of Contrast Volume in Head Angiography by MDCT. Journal of Korean Society of Computed Tomographic. Technology, 10(1), 118-121, 2008
3 Christiansen C: X-ray contrast media, an overview. Toxicology, 209, 185-187, 2005   DOI
4 Mulkens, TH, Bellinck P, Baeyaert M, et al: Use of an Automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation, Radiology, 237, 213-223, 2005   DOI
5 Yu L, Li H, Fletcher JG, et al: Automatic selection of tube potential for radiation dose reduction in CT: a general strategy. Med Phys, 37(1), 234-243, 2010   DOI
6 Caro JJ, Trindade E, McGregor M: The risks of death and of severe nonfatal reactions with high- vs low-osmolarity contrast media, a meta-analysis. AJR Am J Roentgenol 156, 825-832, 1991   DOI
7 Vogl Tj, Honold E, Wolf M, Mohajeri H, Hammerstingl R: Safety of iobitridol in the general population and at-risk patients. Eur Radiol, 16, 1288-1297 2006   DOI
8 Gomi T, Nagamoto M, Hasegawa M, Katoh A, Sugiyama M, Murata N, Kunihiro T, Kohda E: Are there any differences in acute adverse reactions among five low-osmolar non-ionic iodinated contrast media?. Eur Radiol, 20, 1631-1635, 2010   DOI
9 From AM, Bartholmai BJ, Williams AW, et al: Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc,83, 1095-1100, 2008   DOI
10 Marenzi G, Assanelli E, Marana I, et al: N-acetylcysteine and contrastinduced nephropathy in primary angioplasty. N Engl J Med, 354, 2773-2782, 2006   DOI
11 McCullough PA, Wolyn R, Rocher LL, et al: Acute renal failure after coronary intervention: incidence, risk factors, and relationship to mortality. Am J Med, 103, 368-375, 1997   DOI
12 Scanlon PJ, Faxon DP, Audet AM, et al: ACC/AHA guidelines for coronary angiography: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (Committee on Coronary Angiography) developed in collaboration with the Society for Cardiac Angiography and Interventions. J Am Coll Cardiol, 33, 1756-1824, 1999   DOI
13 Myung Jin Shin, Young Joo Cho: Management of adverse reaction to iodinated radiocontrast media. J Korea Med Assoc. 55(8), 779-780, 2012   DOI
14 Marenzi G, Assanelli E, Marana I, et al: N-acetylcysteine and contrast induced nephropathy in primary angioplasty. N Engl J Med, 354, 2773-2782, 2006   DOI
15 Le Jemtel TH, Padeletti M, Jelic S: Diagnostic and Therapeutic Challenges in Patients with Coexistent Chronic Obstructive Pulmonary Disease and Chronic heart Failure, J Am Coll Cardiol, 49, 171-182, 2007   DOI
16 Youngcheol Joo: The Study of Understanding of Waiting and Satisfaction of Examination in CT Appointed Patient: With a Focus Transfer Distance and Time Between Hospital and House, Journal of Korean Society of Computed Tomographic Technology, 11(2), 123-131, 2009
17 From AM, Bartholmai BJ, Williams AW, et al: Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc, 83, 1095-1100, 2008   DOI
18 McCullough PA, Wolyn R, Rocher LL, et al: Acute renal failure aftercoronary intervention: incidence, risk factors, and relationship to mortality. Am J Med, 103, 368-375, 1997   DOI
19 Scanlon PJ, Faxon DP, Audet AM, et al: ACC/AHA guidelines for coronary angiography: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (Committee on Coronary Angiography) developed in collaboration with the Society for Cardiac Angiography and Interventions. J Am Coll Cardiol, 33, 1756-1824, 1999   DOI
20 Lubicz B, Levivier M, Francois O, et al: Sixty-four-row multisection CT angiography for detection and evaluation of ruptured intracranial aneurysms: interobserver and intertechnique reproducibility. AJNR Am JNeuroradiol, 28, 1949-1955, 2007   DOI
21 McKinney AM, Palmer CS, Truwit CL, et al: Detection of aneurysms by 64-section multidetector CT angiography in patients acutely suspectedof having an intracranial aneurysm and comparison with digital subtractionand 3D rotational angiography. AJNR Am J Neuroradiol, 29, 594-602, 2008   DOI
22 Whal Lee, MD: MDCT Application of vascular system. The 2nd Seoul national university hospital MDCT symposium, 77-78, 2007
23 Francesco Papar o, Isabella Garello et al: CT of the abdomen: Degree and quality of enhancement obtained with two concentrations of the same iodinated contrast medium with fixed iodine delivery rate and total iodine load. European Journal of Radiology, 83, 1995-2000, 2014   DOI
24 M.M. Lell, G. Jost, J.G. Korporaal, A.H. Mahnken, T.G. Flohr, M. Uder, H. Pietsch: Optimizing contrast media injection protocols in state-of-the art computed tomographic angiography. Invest, Radiol, 50(3), 161-167, 2015   DOI
25 Fleischmann D. CT angiography: Injection and acquisition technique. Radiol Clin N Am, 48, 237-247, 2010   DOI
26 Huda W, Scalzetti EM, Levin G: Technique factors and image quality asfunctions of patient weight at abdominal CT. Radiology, 217, 430-435, 2000   DOI
27 Ertl-Wagner BB, Hoffmann RT, Bruning R, et al: Multiedetector row CT angiography of the brain at various kilovoltage settings. Radiology, 231, 528-535, 2004   DOI
28 Bahner ML, Bengel A, Brix G, et al: Improved vascular opacification incerebral computed tomography angiography with 80 kV. Invest Radiol, 40, 229-234, 2005   DOI
29 Casper Mihl, Madeleine Kok, Joachim E. Wildberger, Sibel Altintas et al: Coronary CT angiography using low concentrated contrast media injected with high flow rates. EJR, 84, 2155-2160, 2015   DOI
30 C. Mihl, J.E. Wildberger, T. Jurencak, M.J. Yanniello, E.C. Nijssen, J.F. Kalafut, G.Nalbantov, G. Muhlenbruch, F.F. Behrendt, M. Das. Intravascular enhancement with identical iodine delivery rate using different iodine contrast media in a circulation phantom. Invest, Radiol, 48(11), 813-818, 2013   DOI