Abstract
This study was designed the band material in order to reduce the exposure pressure of the breast and the material was measured of Radiolucent and radiation properties with a radiation materials of PC, PMMA, Carbon. Also the image quality by image analysis to obtain the following results are below: Unfors Xi dosimetry using radiation transmittance when the results of 8.353mGy is measured after removal of the cuff, the PC 6.308mGy, PMMA 6.223mGy, Carbon 7.218mGy were measured respectively. Semi-layer PC 0.375mmAl, PMMA 0.370mmAl, with Carbon 0.360mmAl Carbon, PC, PMMA was higher radiation properties and transparency in order. InLight / OSL NanoDotTM dosimeter was used in the cuff and then removed by placing the dosimeter measured results center on 1.143mGy, at the edge 12.56mGy, on the central PC 8.990mGy, at the edge 10.291mGy, PMMA center on 8.391mGy, the edge 9.654mGy, on Carbon center 9.581mGy, 11.313 mGy were measured at the edge of Carbon, PC, PMMA showed a high permeability in order. Image Pixel average J is then removed from the cuff 976.655, PC 831.032, PMMA 819.069, Carbon 897.118 Carbon, PC, PMMA was measured by high order.
Keywords
Mammography; Compression paddle; PC; PMMA; CARBON;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Sung-Soo Kim: Status and countermeasures of patient exposure during mammography jounal of health science & medical technology, 31(1), 118-121, 2005
2 Cardenosa: 유방영상학(Breast Imaging), 한미의학, 2005
3 E klund GW, Cardenos A: The art of mammographic position, RCNA 30(1), 21-53, 1992
4 Dae-Cheol Kweon, Eun-Mi Lee, Sung-Man Hong, et al.: analysis of the compression force and thickness in the screening mammography korean institute of industrial engineer, 606-610, 2002
5 현춘수: 과학용어사전, 뉴턴코리아, 2010
6 http://starletzzang.blog.me/120165887326
7 CJ Hourdakis, ABoziari, E Koumbouli: The effect of a compression paddle on energy response, calibration and measurement with mammographic dosimeters using ionization chambers and solid-state detectors. Physics in Medicine and Biology, 54, 1047-1059, 2009   DOI
8 Gray JE: OSL : Dosimetry for the Twenty-First Century. First RCM on Development of Procedures for in-vivo Dosimetry in Radiotherapy. IAEA, Trieste, 2005
9 Yukihara EG, et al.: High-precision dosimetry for radiotherapy using the optically stimulated luminescence technique and thin dosimeters. Phys Med Biol, 50, 5619-5628, 2005   DOI
10 Aznar MC, et al.: Real-time optical-fibre luminescence dosimetry for radiotherapy, Physical characteristics and applications in photon beams. Phys Med Biol 49, 1655-1669, 2004   DOI
11 Andersen CE, Marckmann CJ, Aznar MC: An algorithm for real-time dosimetry in intensity-modulated radiation therapy using the radioluminescence signal from. Radiat Prot Dosim, 120, 7-13, 2006   DOI
12 Jung-Mi Kim, Su-Dong Jun, Keum-Moon Baek, et al.: Characteristic Evaluation of Optically Stimulated Luminescent Dosimeter (OSLD) for Dosimetry. Koean Society for Radiation Therapy, 22(2), 123-129, 2010
13 김근형외 5명: 플라스틱 이론과 응용. 문운당, 2010
14 과학나눔연구회: 다양한 형태와 천의 얼굴을 가진 탄소재료 탄소섬유. 일진사, 2010
15 Caroline E. Blane, MD, Mitchell M. Goodsitt, PhD, Jeffrey C. Grimm, et al.: New compression paddle for wire localization in mammography. Academic Radiology, 17(2), 142-145, 2010   DOI