Changes of Radiation Dose and Image Quality Due to Additional Filtration Material in Computed Radiography  

Kwon, Soon-Mu (Department of Radiological Science, The Graduate School of Catholic University of Daegu)
Cho, Hyung-Wook (Department of Radiological Science, The Graduate School of Catholic University of Daegu)
Kang, Yeong-Han (Department of Radiology, Catholic University Hospital of Daegu School of Medicine)
Kim, Boo-Soon (Department of Radiologicic Technology, Daegu Health College)
Kim, Jung-Su (Department of Bio-convergence Engineering, Korea University Graduate School)
Publication Information
Journal of radiological science and technology / v.37, no.4, 2014 , pp. 239-246 More about this Journal
Abstract
Filter absorbs low-energy X-ray to increase the average energy and reduces patient exposure dose. This study investigates if the materials of Mo and W could be used for the digital imaging device CR by conducting image assessment and dose measurement of SNR, FOM and histogram. In addition, measurement of beam quality was conducted depending on the material of the filter, and at the same time, a proper combination of filters was examined depending on the change in tube voltage (kVp). In regard to entrance skin dose, Mo filter showed the dose reduction by 42~56%, compared to Cu filter. Moreover, Mo filter showed higher transmission dose by around 1.5 times than that of Cu filter. In image assessment, it was found that W was unsuitable to be used as a filter, whereas Mo could be used as a filter to reduce dose without decline in image quality at the tube voltage of 80 kVp or higher. As tube voltage increased, 2.0 mm Al+0.1 mm Mo almost had a similar histogram width to that of 2.0 mm Al+0.2 mm Cu. Therefore, Mo filter can be used at relatively high tube voltage of 80 kVp, 100 kVp and 120 kVp. The SNR of 2.0 mm Al+0.1 mm Mo did not show any significant difference from those of 2.0 mm Al+0.2 mm Cu and 2.0 mm Al+0.1 mm Cu. As a result, if Mo filter is used to replace Cu filter in general radiography, where 80 kVp or higher is used for digital radiation image, patient exposure dose can be reduced significantly without decline in image quality, compared to Cu filter. Therefore, it is believed that Mo filter can be applied to chest X-ray and high tube voltage X-ray in actual clinical practice.
Keywords
Mo filter; Cu filter; image assessment; patient exposure dose;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Sandborg M, Carlsson CA, Carlsson GA: Shaping x-ray spectra with filters in x-ray diagnostics, Med Biol Eng Comput, 32(4), 384-390, 1994   DOI
2 Behrman RH: The impact of increased Al filtration on x-ray tube loading and image quality in diagnostic radiology, Med Phys, 30(1), 9-78, 2003   DOI   ScienceOn
3 National Council on Radiation Protection and Measurements(NRCP): Medical X-ray and Gamma ray protection for energies up to 10 MeV, NRCP report No. 33, Washington, National Council on Radiation Protection and Measurements, 1968
4 RK Grewal, N Young, L Collins, N Karunaratne, R Sabharwal: Digital chest radiography image quality assessment with dose reduction, Australasian Physical & Engineering Sciences in Medicine, 35(1), 71-80, 2012   DOI
5 Martin C: The importance of radiation quality for optimisation in radiology, Biomed Imaging Interv J, 3(2), e38, 2007
6 Rossi RP, Harnisch B, Hendee WR: Reduction of radiation exposure in radiography of the chest, Radiology, 144(4), 909-914, 1982   DOI
7 Hamer OW, Sirlin CB, Strotzer M, Borisch I, Zorger N, Feuerbach S, Volk M: Chest radiography with a flat-panel detector: Image quality with dose reduction after copper filtration, Radiology, 237(2), 691-700, 2005   DOI   ScienceOn
8 Dance DR, Thilander AK, Sandborg M, Skinner CL, Castellano IA, Carlsson GA: Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study, BJR, 73, 1056-1067, 2000   DOI
9 Nishino TK, Wu X, Johnson RF Jr: Thickness of molybdenum filter and squared contrast-to-noise ratio per dose for digital mammography, AJR, 185(4), 960-963, 2005   DOI
10 Han Jun Yang, Sin Kwan Ko, Mi hwa Joo: Evaluation of MTF Image by Target/Filter combined of X-ray tube using mammography, Journal of Radiological Society and Technology, 30(2), 113-119, 2007
11 Villagran JE, Hobbs BB, Taylor KW: Reduction of patient exposure by use of heavy elements as radiation filters in diagnostic radiology, Radiology, 127(1), 249-254, 1978   DOI
12 Sandborg M, McVey G, Dance DR, Alm Carlsson G: Schemes for the optimization of chest radiography using a computer model of the patient and x-ray imaging system, Med Phys, 28(10), 2007-2019, 2001   DOI
13 Williams MB, Raghunathan P, More MJ etc: Optimization of exposure parameters in full field digital mammography, Med Phys, 35(6), 2414-2423, 2008   DOI   ScienceOn
14 Nam-gil Choi, Ho-Jin Seong, Joo-Seop Jeon, Youn-Hyun Kim and Dong-Ook Seong: A comparative study of image quality and radiation dose according to variable added filter and radiation exposure in diagnostic x-ray radiography, J Radiat Prot, 37(1), 25-34, 2012   DOI
15 Aichinger H, Dierker J, Joite-Barfuss S, Sabel M: Radiation exposure and Image Quality in x-ray Diagnostic Radiology, Heidelverg, Germany, Springer Verlag, 2004
16 Doyle P, Martin CJ, Gentle D: Application of contrast-to-noise ratio in optimizing beam quality of digital chest radiography: comparison of experimental measurements and theoretical simulations, Phys Med Biol, 51(11), 2953-2970, 2006   DOI