Evaluation of Scatter Radiation in Digital Radiological Condition by using Photostimulated Luminescence (BaFBr:$Eu^{2+}$)  

Min, Jung-Whan (Department of Radiological Science, Shin-Gu University)
Han, Seong-Gyu (Department of Bio-convergence engineering, Graduate school, Korea University)
Kim, Jung-Min (Department of Radiological Science, Korea University)
Lee, Joo-Ah (Department of Oncology, Catholic University of Korea Incheon St.Mary's Hospital)
Kim, Ki-Won (Department of Radiology, Samsung Medical Center)
Jeong, Hoi-Woun (Department of Radiological Science, Beakseok Culture University)
Publication Information
Journal of radiological science and technology / v.37, no.2, 2014 , pp. 85-91 More about this Journal
Abstract
The purpose of this study is evaluated scatter radiation in digital radiological condition by using photo-stimulated luminescence (BaFBr:$Eu^{2+}$). Experiment condition changed kVp (from 50 kVp to 120 kVp), filed size (from $4{\times}4cm^2$ to $26{\times}26cm^2$) and phantom thickness (from 1 cm to 15 cm). This method was analysed ImageJ and characteristic curve of CR. This results was scatter radiation to primary radiation ratio increased from 50 kVp to 70 kVp, and it was fixed at over 80 kVp. The scatter radiation to primary radiation ratio are increased according to increasing the ratio of field size. Scatter radiation is also increased by increasing the phantom thickness.
Keywords
Scatter radiation; Field size; scatter radiation to primary radiation ratio;
Citations & Related Records
Times Cited By KSCI : 7  (Citation Analysis)
연도 인용수 순위
1 H. P. Chan and K. Doi : Physical characteristics of scattered radiation and the performance of antiscatter grids in diagnostic radiology, Radiographics, 2(3), 378-406, 1982   DOI
2 Jung-Min Kim, Jung-Whan Min, Hoi-Woun eong et al: The Noise Evaluation for Ragius 150 CR System. Journal of Radiological Science and Technology, 29(4), 237-240, 2006
3 Jung-Min Kim: Edge Profile Correction for Modulation Transfer Function on Digital X-ray Imaging System. Journal of Radiological Science and Technology, 30(3), 251-257, 2007
4 Jung-Min Kim, Ki-Sung Lee, You-Hyun Kim: Correction Method of Wiener Spectrum (WS) on Digital Medical Imaging Systems. Journal of Radiological Science and Technology, 32(1), 17-24, 2009
5 Hoi-Woun Jeong, Jung-Whan Min, Jung-Min Kim et al: Performance characteristic of a Cs I (TI) Flat panel detector radiography system. Journal of Radiological Science and Technology, 35(2), 109-117, 2012
6 Jung-Whan Min, Hoi-Woun Jeong, Jung-Min Kim et al: Comparison of Noise Power Spectrum Methodologies in Measurements by Using Megavoltage X-ray Energies. Journal of Korean Physics Society 60, 129-136, 2012   DOI   ScienceOn
7 Jung-Whan Min, Hoi-Woun Jeong, Jung-Min Kim et al: Performance of an Edge Block Used in a Configuration Detector: Image Quality Measurements. Journal of Korean Physics Society 64, 732-739, 2014   DOI   ScienceOn
8 International Electrotechnical Commission : Medical electrical equipment - Characteristics of digital X-ray imaging devices-Part 1 : determination of the detective quantum efficiency. IEC 62220-1, 2003
9 NIH, ImageJ : http://rsb.info.nih.gov/ij/
10 Simone Cardoso, Odair Goncalves, Helio Schechter : Evaluation of scatter-to-primary ratio in radiological conditions, Applied Radiation and Isotopes, 7, 544-548, 2009
11 Paul C. Johns and Martine J. Yaffe : Scattered radiation in fan beam imaging systems. Med. Phys., 9(2), 231-239, 1982   DOI   ScienceOn
12 Joon Huh, Chang-Kyun Kim, Jung-Min Kim et al: Influence of X-ray Tube Voltage and Object on X-ray Quality and Dose. Journal of Radiological Science and Technology, 7(1), 41-46, 1984
13 KFDA: A guide for general radiology of the patient dose recommend. 2012
14 Jung-Min Kim, Jung-Whan Min, Hoi-Woun Jeong et al: Correction Method of slit Modulation Transfer function on Digital Medical Imaging System. Journal of Radiological Science and Technology, 29(3), 133-139, 2006