Development of Dynamic Kidney Phantom System and its Evaluation of Usability of Application in Nuclear Medicine  

Park, Hoon-Hee (Dept. of Radiological Technology, Shingu College)
Lee, Juyoung (Dept. of Molecular Imaging, Philips Healthcare)
Kim, Sang-Wook (Dept. of Diagnostic Radiology, Korea Medical Institute)
Lyu, Kwang Yeul (Dept. of Radiological Technology, Shingu College)
Jin, Gye Hwan (Dept. of Radiological Science, Nambu University)
Publication Information
Journal of radiological science and technology / v.36, no.1, 2013 , pp. 49-55 More about this Journal
Abstract
Currently, commercially available phantom can reproduce and evaluate only a static situation, the study is incomplete research on phantom and system which is can confirmed functional situation in the kidney by time through dynamic phantom and blood flow velocity, various difference according to the amount of radioactive. Therefore, through this study, it has produced the dynamic kidney phantom to reproduce images through the dynamic flow of the kidney, it desires to evaluate the usefulness of nuclear medicine imaging. The production of the kidney phantom was fabricated based on the normal adult kidney, in order to reproduce the dynamic situation based on the fabricated kidney phantom, in this study, it was applied the volume pump that can adjust the speed of blood flow, so it can be integrated continuously radioactive isotopes in the kidney by using $^{99m}Tc$-pertechnate. Used the radioactive isotope was supplied through the two pump. It was confirmed the changes according to the infusion rate, radioactive isotopes and the different injection speeds on the left and right, analysis of the acquired images was done by drawn five times ROI in order to check the reproducibility of each on the front and rear of the kidney and bladder. Depending on the speed of injection, radioisotope was a lot of integrated and emissions up when adjusting the pressure of the pump as 30 stroke, it was the least integrated and emissions up when adjusting as 40 stroke. The integration of the left & right kidney was not reached in the amount of the highest when adjusting as 10 stroke. In the changes according to the amount of the radioactive isotope, 0.6 mCi(22.2 MBq), 0.8 mCi (29.6 MBq)was showed up similar tendency but, in the result of the different injection 0.8 mCi, it was showed up counts close to double of 0.6 mCi. In the result of the differently injection speed of the left & right kidney, as a result of different conditions that injection speed was 20 stroke through left kidney phantom, the injection speed was 30 stroke through right kidney phantom, it was enough difference in the resulting image can be easily distinguished with the naked eye. Through this study, the results showed that the dynamic kidney phantom system is able to similarly reproduce renogram in the actual clinical practice. Especially, the depicted over time for the flow to be excreted through the kidney into the bladder was adequately reproduce, it is expected to be utilized as basic data to check the quality of the dynamic images. In addition, it is considered to help in the field of functional imaging and quality control.
Keywords
Dynamic Renal Scintigraphy; Phantom Invention; Radiopharmaceuticals;
Citations & Related Records
연도 인용수 순위
  • Reference
1 C. Miyazaki, H. Harada, et al.: $^{99m}Tc$-DTPA dynamic SPECT and CT volumetry for measuring split renal function in live kidney donors, Ann Nucl Med, 24, 189-195, 2010   DOI
2 Ying-Chun Ma, Li Zuo, et al.: Comparison of $^{99m}Tc$-DTPA renal dynamic imaging with modified MDRD equation for glomerular filtration rate estimation in Chinese patients in different stages of chronic kidney disease, Nephrol Dial Transplant, 22, 417-423, 2007   DOI
3 R.D. Folks, E. V .Garcia, A. T. Taylor : Development and Prospective Evaluation of an Automated Software System for Quality Control of Quantitative $^{99m}Tc$-MAG3 Renal Studies, J Nucl Med Technol, 35, 57-33, 2007
4 J.O. Heikkinen : New Automated Physical Phantom for Renography, J Nucl Med, 45, 495-499, 2004
5 I. Karagoz, O. Erogul, et al.: A new dynamic renal phantom and its application to scintigraphic studies for pixel basis functional radionuclide imaging, 20, 473-479, 1998
6 A Celler, T Farncombe, et al.: Performance of the dynamic single photon emission computed tomography (dSPECT) method for decreasing or increasing activity changes, Phys Med Biol, 3525-3543, 2000
7 A.S. Houston, D.R. Whalley et al.: UKaudit and analysis of quantitative parameters obtained from gamma camera renography, Nuclear Medicine Communications, 22, 559-566, 2001   DOI   ScienceOn
8 J.O. Heikkiene, J.T. Kuikka, A.K.A. Ahonen, P.J. Rautio : Quality of dynamic radionuclide renal imaging:multicentre evaluation using a functional renal phantom, Nuclear Medicine Communications, 22, 987-995, 2001   DOI   ScienceOn