Abstract
Used as an ingredient in the hospital for orthopedic prosthetic stainless and titanium metal the same size as on the MRI diagnostic value of imaging were compared. Center of images, background high band portion of the video signal is converted into a weighted intensity values Normal images and compared. The area of normal slice and also the distortion of images and cross-sectional imaging of a range of quantitative and sagittal planes were compared. As a result, the periphery high band signal intensity values of Stainless video phantom 2, pig bone 1.8, Titanium imaging of phantom 1.7 has higher value than Normal video pig bone 1.3 times the signal strength rivers. MRI distortion of the shape and the distortions of the range, if the cross-sectional area compared to Normal Slice Stainless case of phantom 65.8 %, pig bone 61.5 %, Titanium distortion phantom 23.1 %, pig bone 38.5 % of the range of community found. In this experiment, as a result, MRI was found to be Titanium more diagnostic value than the specimen with respect to the signal intensity weighted value and low distortion range, Stainless.
Keywords
MRI; Orthopedic; Image Distortion; Stainless; Titanium;
Citations & Related Records
연도 인용수 순위
  • Reference
1 홍현숙, 서원혁 : 자기공명 영상에서의 생체의학 이식 재료에 대한 인공물에 관한 연구, 고려논문집, 29(1), 195-200, 1992
2 이상복, 이삼열, 이준행 : MRI 영상에서 영역추출과 질환인식, 대한방사선기술학회지, 27(3), 19-24, 2004
3 Shellock FG, Morisoli S, Kanal E: MR procedures and biomedical implants, materials, and devices: 1993updates. Radiology, 189, 587-599, 1993
4 김형균, 최성대 : 정형보철용 금속이 자기공명영상에 미치는 영향 II, 한국 기계 가공학회지, 11(5), 115-120, 2012
5 The Korean Society of Magnetic Resonance Imaging : 자기공명영상학, 청구문화사, 16-17, 37-50, 55, 67-69, 107-110, 119-132, 2008
6 이성우 : 최신 자기공명기술학, 여문각, 288-289, 1998
7 이기섭 : MRI 기초물리, 대한방사선과학회 학술대회 논문집, 13.2-13.2, 1991
8 박용훈, 최용석, 김형태 등 : Magnet Depth type에 따른 Metallic susceptibility Artifact에 관한 고찰, 대한자기공명기술학회지, 19, 45-52, 2009
9 Shellock FG : MR imaging of metallic implants and materials: a complication of the literature. AJR, 151, 811-814, 1998