Evaluation of Scatter Reduction Effect of the Aft-Multiple-Slit (AMS) System Using MC Simulation  

Chang, Jin-A (Department of Biomedical Engineering, The Catholic University of Korea School of Medicine)
Suh, Tae-Suk (Department of Biomedical Engineering, The Catholic University of Korea School of Medicine)
Jang, Doh-Yun (Department of Nuclear Engineering, Hanyang University)
Jang, Hong-Seok (Department of Radiation Oncology, Seoul St. Mary's Hospital, The Catholic University of Korea School of Medicine)
Kim, Si-Yong (Department of Radiation Oncology, Mayo Clinic)
Publication Information
Radiation Oncology Journal / v.28, no.4, 2010 , pp. 224-230 More about this Journal
Abstract
Purpose: We designed the aft-multiple-slit (AMS) system to reduce scatter in cone-beam computed tomography (CBCT). As a preliminary study, we performed a Monte Carlo N-Particle Transport Code (MCNP) simulation to verify the effectiveness of this system. Materials and Methods: The MCNPX code was used to build the AMS geometry. An AMS is an equi-angled arc to consider beam divergence. The scatter-reduced projection images were compared with the primary images only and the primary plus scatter radiation images with and without AMS to evaluate the effectiveness of scatter reduction. To obtain the full 2 dimensional (2D) projection image, the whole AMS system was moved to obtain closed septa of the AMS after the first image acquisition. Results: The primary radiation with and without AMS is identical to all the slit widths, but the profiles of the primary plus scattered radiation varied according to the slit widths in the 2D projection image. The average scatter reduction factors were 29%, 15%, 9%, and 8% when the slit widths were 5 mm, 10 mm, 15 mm, and 20 mm, respectively. Conclusion: We have evaluated the scatter reduction effect of the AMS in CBCT imaging using the Monte Carlo (MC) simulations. A preliminary study based on the MCNP simulations showed a mount of scatter reduction with the proposed system.
Keywords
Aft-multiple-slit (AMS); Scatter reduction; Cone-beam CT (BCT);
Citations & Related Records
연도 인용수 순위
  • Reference
1 Siewerdsen JH, Daly MJ, Bakhtiar B, et al. A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT. Med Phys 2006;33 187-197   DOI   ScienceOn
2 Neitzel U. Grids or air gaps for scatter reduction in digital radiography: a model calculation. Med Phys 1992;19:475-481   DOI   ScienceOn
3 Lui X, Shaw CC. Effects of anti-scatter grid on measurement of detector characteristics. Med Phys 2002;29:1342
4 Jing Z, Huda W, Walker JK. Scattered radiation in scanning slot mammography. Med Phys 1998;25:1111-1117   DOI   ScienceOn
5 Mainprize JG, Ford NL, Yin S, Tumer T, Yaffe MJ. A slot-scanned photodiode-array/CCD hybrid detector for digital mammography. Med Phys 2002;29:214-225   DOI   ScienceOn
6 Yang Y, Schreibmann E, Li T, Wang C, Xing L. Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation. Phys Med Biol 2007;52:685-705   DOI   ScienceOn
7 Ning R, Tang X, Conover D. X-ray scatter correction algorithm lor cone beam CT imaging. Med Phys 2004;31:1195-1202   DOI   ScienceOn
8 Morin O, Chen J, Aubin M, et al. Dose calculation using megavoltage cone-beam CT. Int J Radiat Oncol Biol Phys 2007;67:1201-1210   DOI   ScienceOn
9 Tkaczyk JE, Du Y, Walter D, Wu X, Li J, Toth T. Simulation of CT dose and contrast-to-noise as a function of bowtie shape. Proc SPIE 2004;5368:403-410
10 Mail N, Moseley DJ, Siewerdsen JH, Jaffray DA. The influence of bowtie filtration on cone-beam CT image quality. Med Phys 2009;36:22-32   DOI   ScienceOn
11 Shaw CC, Wang T, Gur D. Effectiveness of antiscatter grids in digital radiography. A phantom study. Invest Radiol 1994;29:636-642   DOI   ScienceOn
12 Samant SS, Xia J, Muyan-Ozcelik P, Owens JD. High performance computing for deformable image registration: towards a new paradigm in adaptive radiotherapy. Med Phys 2008;35:3546-3553   DOI   ScienceOn
13 Hawkins MA, Brock KK, Eccles C, Moseley D, Jaffray D, Dawson LA. Assessment of residual error in liver position using kV cone-beam computed tomography for liver cancer high-precision radiation therapy. Int J Radiat Oncol Biol Phys 2006;66:610-619   DOI   ScienceOn
14 Thilmann C, Nill S, Tucking T, et al. Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences. Radiat Oncol 2006;1:16   DOI
15 Langen KM, Meeks SL, Poole DO, et al. The use of megavoltage CT (MVCT) images for dose recomputations. Phys Med Bol 2005;50:4259-4276   DOI   ScienceOn
16 Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am 1984;1:612-619   DOI
17 Jaffray DA, Siewerdsen JH, Wong JW, Martinez AA. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 2002;53 1337-1349   DOI   ScienceOn
18 Samei E, Saunders RS, Lo JY, et al. Fundamental imaging characteristics of a slot-scan digital chest radiographic system. Med Phys 2004;31:2687-2698   DOI   ScienceOn