The Analysis of Dose in a Rectum by Multipurpose Brachytherapy Phantom  

Huh, Hyun-Do (Department of Radiation Oncology, College of Medicine, Inha University)
Kim, Seong-Hoon (Department of Radiation Oncology, College of Medicine, Hanyang University)
Cho, Sam-Ju (Department of Medical Physics, Kyonggi University)
Lee, Suk (Department of Radiation Oncology College of Medicine, Korea University)
Shin, Dong-Oh (Department of Radiation Oncology, College of Medicine, KyungHee University)
Kwon, Soo-Il (Department of Medical Physics, Kyonggi University)
Kim, Hun-Jung (Department of Radiation Oncology, College of Medicine, Inha University)
Kim, Woo-Chul (Department of Radiation Oncology, College of Medicine, Inha University)
K. Loh John-J. (Department of Radiation Oncology, College of Medicine, Inha University)
Publication Information
Radiation Oncology Journal / v.23, no.4, 2005 , pp. 223-229 More about this Journal
Abstract
Purpose: In this work we designed and made MPBP(Multi Purpose Brachytherapy Phantom). The MPBP enables one to reproduce the same patient set-up in MPBP as the treatment of the patient and we tried to get an exact analysis of rectal doses in the phantom without need of in-vivo dosimetry. Materials and Methods: Dose measurements were tried at a point of rectum 1, the reference point of rectum, with a diode detector for 4 patients treated with tandem and ovoid for a brachytherapy of a cervix cancer. Total 20 times of rectal dose measurements were made with 5 times a patient. The set-up variation of the diode detector was analyzed. The same patient set-ups were reproduced in self-made MPBP and then rectal doses were measured with TLD. Results: The measurement results of the diode detector showed that the set-up variation of the diode detector was the maximum $11.25{\pm}0.95mm$ in the y-direction for Patient 1 and the maximum $9.90{\pm}4.50mm,\;20.85{\pm}4.50mm,\;and\;19.15{\pm}3.33mm$ in the z-direction for Patient 2, 3, and 4, respectively. Un analyzing the degree of variation in 3 directions the more variation was showed in the z-direction than x- and y-direction except Patient 1. The results of TLD measurements in MPBP showed the relative maximum error of 8.6% and 7.7% at a point of rectum 1 for Patient 1 and 4, respectively and 1.7% and 1.2% for Patient 2 and 3, respectively. The doses measured at R1 and R2 were higher than those calculated except R point of Patient 2. this can be thought to related to the algorithm of dose calculation, whcih corrects for air and water but is guessed not to consider the correction for the scattered rays, but by considering the self-error (${\pm}5%$) TLD has the relative error of values measured and calculated was analyzed to be in a good agreement within 15%. Conclusion: The reproducibility of dose measurements under the same condition as the treatment could be achieved owing to the self-made MPMP and the dose at the point of interest could be analyzed accurately. If a treatment is peformed after achieving dose optimization using the data obtained in the phantom, dose will be able to be minimized to important organs.
Keywords
Brachytherapy; MPBP(multi Purpose brachytherapy Phantom); MFA(multi function applicator); Diode detector; TLD;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Saarnak AE, Boersma M, Bunningen BNFM, Wolterink R, Steggerda MJ. Inter-observer variation in delineation of bladder and rectum contours for brachytherapy of cervical cancer. Radiother Oncol 2000;56:37-38   DOI   ScienceOn
2 Hellebust TP, Dale E, Skjonsberg A, Olsen DR. Inter fraction variations in rectum and bladder volumes and dose distributions during high dose rate brachytherapy treatment of the uterine cervix investigated by repetitive CT-examinations. Radiother Oncol 2001;60:273-280   DOI   ScienceOn
3 Gerstner NW, Wacher S, Reinstadler E, et al. Bladder and rectum dose defined from MRI base treatment planning for cervix cancer brachytherapy: comparison of dose-volume histograms for organ contours and organ wall, comparison with ICRU rectum and bladder reference point. Radiother Oncol 2003;68:269-276   DOI   ScienceOn
4 Alecu R, Alecu M. In-vivo rectal dose measurements with diodes to avoid misadministrations during intracavitary high dose rate brachytherapy for carcinoma of the cervix. Med phys 1999; 26:768-770   DOI   ScienceOn
5 Roeske JC, Mundt AJ, Halpern H, et al. Late rectal sequlae following definitive radiation therapy for carcinoma of the uterine cervix: a dosimetric analysis. Int J Radiat Oncol Biol phys 1997;37:351-358   DOI   ScienceOn
6 Chung EJ, Lee SH. Comparison between the calculated and measured dose in the rectum during high dose rate brachytherapy for uterine cervical carcinomas. J Korean Soc ther Radiol 2002;20:396-404
7 Sanini AS, Zhu TC. Temperature dependence of commercially available diode detectors. Med Phys 2002;29:622-630   DOI   ScienceOn
8 Ogino I, Kitamura T, Okamto N, et al. Late rectal complication following high dose rate intracavitary brachytherapy in cancer of the cervix. Int J Radiat Oncol Biol Phys 1995;32: 725-734   DOI   ScienceOn
9 ICRU Report 38: Dose and volume specification for reporting intracavitary therapy in gynecology. International Commission on Radiation Units and Measurements, Bethesda. MD 1985
10 AAPM. Task Group 40: Comprehensive QA for radiation oncology. Med phys 1994;21:581-618   DOI   ScienceOn
11 Cunningham DE, Stryker JA, Velkley DE, Chung CK. Routine clinical estimation of rectal, rectosigmoidal, and bladder dose from intracavitary brachytherapy in the treatment of carcinoma of the cervix. Int J Radiat Oncol Biol Phys 1981; 7:653-660   DOI
12 Eng TY, Fuller CD, Cavanaugh SX, Blough MM, Sadeghi A, Herman T. Significant rectal and bladder dose reduction via utilization of foley balloon catheters in high-dose-rate tandem and ovoid intracavitary brachytherapy of the uterine cervix. Int J Radiat Oncol Biol Phys 2004;59:174-178   DOI   ScienceOn
13 Crook JM, Esche BA, Chaplain G, Iose I, Sentenac I, Horiot JC. Dose-Volume analysis and the prevention of radiation sequelae in cervical cancer. Radiother Oncol 1987;8: 321-332   DOI   ScienceOn
14 Clark BG, Souhami L, Roman TN, Evans MDC, Pla C. Rectal complications in patients with carcinoma of the cervix treated with concomitant cisplatin and external beam irradiation with high dose rate brachytherapy: a dosimetric analysis. Int J Radit Oncol Biol Phys 1994;28:1243-1250   DOI
15 Shin KH, Huh SJ, Chie EK, et al. Analysis of correlation between rectal complications and rectal dose following high dose rate intracavitary radiotherapy in patients with uterine cervix cancer: In vivo Dosimetric Analysis. Radiation Medicine 1999;17: 289-293
16 elloski CE, Palmer M, Chronowski GM, Jhingran A, Horton J, Eifel PJ. Comparison between CT-based volumetric calculations and ICRU reference-point estimates of radiation dose delivered to bladder and rectum during intracavitary radiotherapy for cervical cancer. Int J Radiat Oncol Biol Phys 2005;62:131-137   DOI   ScienceOn
17 Waldhaausl C, Wambersie A, Potter R, Georg D. In-vivo dosimetry for gynecological brachytherapy: physical and clinical considerations. Radiother Oncol 2005;(in press):1-8
18 Joslin CA, Flynn A, Hall EJ. Computers in brachytherapy dosimetry. In:Arnold, Principles and Practice of Brachytherapy. 1st ed. New York, Oxford University press Inc. 2001:49-52
19 Scalchi P, Francescon P. Calibration of a MOSFET detection system for 6-MV In vivo dosimetry. Int J Radiat Oncol Biol Phys 1998;40:987-993   DOI   ScienceOn
20 Pai S, Reinstein LE, Gluckman G, Xu Z, Weiss T. The use of improved radiochromic film for in vivo quality assurance of high dose rate brachytherapy. Med phys 1998;25:1217-1221   DOI   ScienceOn
21 Kessaris N, Nori D. Rectal thermoluminescent dosimeter measurements using the microSelectron high dose rate afterloader in postoperative intravaginal radiation case. Int J Radiat Oncol Biol Phys 1992;26:341-345   DOI