Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients  

Lee, Re-Na (Department on Radiation Oncology, College of Medicine, Ewha Womans University Mokdong Hospital)
Publication Information
Radiation Oncology Journal / v.23, no.3, 2005 , pp. 169-175 More about this Journal
Abstract
Purpose: Although computed tomography (CT) simulators are commonly used in radiation therapy department, many Institution still use conventional CT for treatments. In this study the setup errors that occur during simulation, CT scan (diagnostic CT scanner), and treatment were evaluated for the twenty one breast cancer patients. Materials and Methods: Errors were determined by calculating the differences in isocenter location, SSD, CLD, and locations of surgical clips implanted during surgery. The anatomic structures on simulation film and DRR image were compared to determine the movement of isocenter between simulation and CT scan. The isocetner point determined from the radio-opaque wires placed on patient's surface during CT scan was moved to new position if there was anatomic mismatch between the two images Results: In 7/21 patients, anatomic structures on DRR Image were different from the simulation Image thus new isocenter points were placed for treatment planning. The standard deviations of the diagnostic CT setup errors relative to the simulator setup in lateral, longitudinal, and anterior-posterior directions were 2.3, 1.6, and 1.6 mm, respectively. The average variation and standard deviation of SSD from AP field were 1.9 mm and 2.3 mm and from tangential fields were 2.8 mm and 3.7 mm. The variation of the CLD for the 21 patients ranged from 0 to 6 mm between simulation and DRR and 0 to 5 mm between simulation and treatment. The group systematic errors analyzed based on clip locations were 1.7 mm in lateral direction, 2.1 mm in AP direction, and 1.7 mm in SI direction. Conclusion: These results represent that there was no significant differences when SSD, CLD, clips' locations and isocenter locations were considered. Therefore, it is concluded that when a diagnostic CT scanner is used to acquire an image, the set-up variation is acceptable compared to using CT simulator for the treatment of breast cancer. However, the patient has to be positioned with care during CT scan in order to reduce the setup error between simulation and CT scan.
Keywords
Breast cancer; Setup error;
Citations & Related Records
연도 인용수 순위
  • Reference
1 ICRU Report 62: Prescribing, recording, and reporting Photon Beam Therapy. Inthernational Commission in Radiation Units and Mesurement, Bethesda, MD 1995
2 Haslam JJ, Lujan AE, Mundt AJ, et al. Setup errors in patients treated with intensity-modulated whole pelvic radiation therapy for gynecological malignancies. Med Dosim 2005;30: 36-42   DOI   ScienceOn
3 Alasti H, Petric MP, Catton CN, et al. Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate. Int J Radiat Oncol Biol Phys 2001;49:869-884   DOI   ScienceOn
4 Bijhold J, Lebesque JV, Hart AM, et al. Maximizing seup accuracy using portal images as applied to a conformal boost technique for prostate cancer. Radiother Oncol 1992; 24:261-271   DOI   ScienceOn
5 van Tienhoven G, Lanson JH, Cabeels D, et al. Accuracy in tangential breast set-up: a portal imaging study. Radiother Oncol 2005;22:317-322   DOI   ScienceOn
6 Valdagani R, Italia C. Early cancer irradiation after conservative surgery: quality control by portal localization films. Radiather Oncol 1991;22:341-343
7 Balter JM, Ten Haken RK, Lawrence TS, et al. Uncertainties in CT-based radiation therapy reatment planning associated with patient breating. Int J Radiat Oncol Biol Phys 1996;36:164-174   DOI   ScienceOn
8 Giraud P, De Rycke Y, Dubray B, et al. Conformal radiotherapy (CRT) planning for lung cancer: analysis of intrathoracic organ motion during extreme phases of breathing. Int J Radiat Oncol Biol Phys 2001;51:1081-1092   DOI   ScienceOn
9 Hanley J, Lumley MA, Mageras GS, et al. Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostate. Int J Radiat Oncol Biol Phys 1997;37:435-444   DOI   ScienceOn
10 Fein DA, McGee KP, Schultheiss TE, Fowble BL, Hanks GE. Intra- and interfractional reproducibility of tangential breast fields: a prospective on-line portal imaging study. Int J Radiat Oncol Biol Phys 1996;34:733-740   DOI   ScienceOn
11 Horst E, Micke O, Moustakis C, Schuck A, Schafer U, Willich NA. Conformal therapy for pancreatic cancer: variation of organ position due to gastrointerstinal distention-implications for treatment planning. Radiology 2002;222:681-686   DOI   ScienceOn
12 de Boer HC, van Sornsen de Koste JR, Senan S, et al. Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning. Int J Radiat Oncol Biol Phys 2001;49:857-868   DOI   ScienceOn
13 Astreinidou E, Bel A, Raajjmakers CP, et al. Adequate margins for random setup uncertainties in head-and-neck IMRT. Int J Radiat Oncol Biol Phys 2005;61:938-944   DOI   ScienceOn
14 Stroom JC, de Boer JC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys 1999;43:905-919   DOI   ScienceOn
15 Yan D, Wong J, Vicini F, et al. Adaptive modification of treatment planning to minimize the deleterious effects of treatment setup errors. Int J Radiat Oncol Biol Phys 1997;38: 197-206   DOI   ScienceOn
16 Smith RP, Bloch P, Harris EE, et al. Analysis of interfraction and intrafraction variation during tangential breast irradiation with an electronic portal imaging device. Int J Radiat Oncol Biol Phys 2005;62:373-378   DOI   ScienceOn
17 Kukolowicz PF, Debrowski A, Gut P, et al. Evaluation of set-up deviations during the irradiation of patients suffering from breast cancer treated with two different techniques. Radiother Oncol 2005;75:22-27   DOI   ScienceOn
18 Kitamura K, Shirato H, Seppenwoolde Y, et al. Threedimensional intrafractional movement of prostate measured during real-time tumor-tracking radiotherapy in supine and prone treatment position. Int J Radiat Oncol Biol Phys 2002; 53:1117-1123   DOI   ScienceOn