Browse > Article
http://dx.doi.org/10.14407/jrpr.2021.00017

Study on Dual-Energy Signal and Noise of Double-Exposure X-Ray Imaging for High Conspicuity  

Song, Boram (Department of Radiology, Haeundae Paik Hospital, Inje University College of Medicine)
Kim, Changsoo (Department of Radiological Science, College of Health Sciences, Catholic University of Pusan)
Kim, Junwoo (Center for Advanced Medical Engineering Research, Pusan National University)
Publication Information
Journal of Radiation Protection and Research / v.46, no.4, 2021 , pp. 160-169 More about this Journal
Abstract
Background: Dual-energy X-ray images (DEI) can distinguish or improve materials of interest in a two-dimensional radiographic image, by combining two images obtained from separate low and high energies. The concepts of DEI performance describing the performance of double-exposure DEI systems in the Fourier domain been previously introduced, however, the performance of double-exposure DEI itself in terms of various parameters, has not been reported. Materials and Methods: To investigate the DEI performance, signal-difference-to-noise ratio, modulation transfer function, noise power spectrum, and noise equivalent quanta were used. Low- and high-energy were 60 and 130 kVp with 0.01-0.09 mGy, respectively. The energy-separation filter material and its thicknesses were tin (Sn) and 0.0-1.0 mm, respectively. Noise-reduction (NR) filtering used the Gaussian-filter NR, median-filter NR, and anti-correlated NR. Results and Discussion: DEI performance was affected by Sn-filter thickness, weighting factor, and dose allocation. All NR filtering successfully reduced noise, when compared with the dual-energy (DE) images without any NR filtering. Conclusion: The results indicated the significance of investigating, and evaluating suitable DEI performance, for DE images in chest radiography applications. Additionally, all the NR filtering methods were effective at reducing noise in the resultant DE images.
Keywords
Dual-Energy X-Ray Imaging; Modulation Transfer Function; Noise Power Spectrum; NEQ; Noise Reduction;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Shkumat NA, Siewerdsen JH, Dhanantwari AC, Williams DB, Richard S, Paul NS, et al. Optimization of image acquisition techniques for dual-energy imaging of the chest. Med Phys. 2007;34(10):3904-3915.   DOI
2 Richard S. Optimization of imaging performance and conspicuity in dual-energy x-ray radiography [dissertation]. Toronto, Canada: University of Toronto; 2008.
3 Bowman WA, Robar JL, Sattarivand M. Optimizing dual-energy x-ray parameters for the ExacTrac clinical stereoscopic imaging system to enhance soft-tissue imaging. Med Phys. 2017;44(3):823-831.   DOI
4 Jeon PH, Chung H, Kim D. Investigation of the effect of kV combinations on image quality for virtual monochromatic imaging using dual-energy CT: a phantom study. J Radiat Prot Res. 2018;43(1):1-9.   DOI
5 Richard S, Siewerdsen JH. Optimization of dual-energy imaging systems using generalized NEQ and imaging task. Med Phys. 2007;34(1):127-139.   DOI
6 Tucker DM, Barnes GT, Chakraborty DP. Semiempirical model for generating tungsten target x-ray spectra. Med Phys. 1991;18(2):211-218.   DOI
7 Kim DW, Park J, Kim J, Kim HK. Noise-reduction approaches to single-shot dual-energy imaging with a multilayer detector. J Instrum. 2019;14(1):C01021.   DOI
8 Ergun DL, Mistretta CA, Brown DE, Bystrianyk RT, Sze WK, Kelcz F, et al. Single-exposure dual-energy computed radiography: improved detection and processing. Radiology. 1990;174(1):243-249.   DOI
9 McCollough CH, Van Lysel MS, Peppler WW, Mistretta CA. A correlated noise reduction algorithm for dual-energy digital subtraction angiography. Med Phys. 1989;16(6):873-880.   DOI
10 Warp RJ, Dobbins JT. Quantitative evaluation of noise reduction strategies in dual-energy imaging. Med Phys. 2003;30(2):190-198.   DOI
11 Kim J, Kim DW, Kim SH, Yun S, Youn H, Jeon H, et al. Linear modeling of single-shot dual-energy x-ray imaging using a sandwich detector. J Instrum. 2017;12(1):C01029.   DOI
12 Friedman SN, Cunningham IA. Normalization of the modulation transfer function: the open-field approach. Med Phys. 2008;35(10):4443-4449.   DOI
13 Samei E, Flynn MJ, Reimann DA. A method for measuring the presampled MTF of digital radiographic systems using an edge test device. Med Phys. 1998;25(1):102-113.   DOI
14 Dobbins JT. Image quality metrics for digital systems. In: Beutel J, Kundel HL, Van Metter RL, editors. Handbook of medical imaging. Bellingham, WA: International Society for Optics and Photonics; 2000.
15 Siewerdsen JH, Jaffray DA. Three-dimensional NEQ transfer characteristics of volume CT using direct-and indirect-detection flat-panel imagers. In: Medical Imaging 2003: physics of medical imaging. Bellingham, WA: International Society for Optics and Photonics; 2003.
16 Richard S, Siewerdsen JH. Cascaded systems analysis of noise reduction algorithms in dual-energy imaging. Med Phys. 2008;35(2):586-601.   DOI
17 Kashani H, Varon CA, Paul NS, Gang GJ, Van Metter R, Yorkston J, et al. Diagnostic performance of a prototype dual-energy chest imaging system ROC analysis. Acad Radiol. 2010;17(3):298-308.   DOI
18 Martinez C, de Molina C, Desco M, Abella M. Optimization of a calibration phantom for quantitative radiography. Med Phys. 2021;48(3):1039-1053.   DOI