Browse > Article

Development of Spectroscopy Toolkit for Spectrum Measurement Experiments Using a CsI(Tl)/PIN Diode Detector  

Nam, Young-Mi (Korea Atomic Energy Research Institute)
Kim, Han-Soo (Korea Atomic Energy Research Institute)
Ha, Jang-Ho (Korea Atomic Energy Research Institute)
Lee, Jae-Hyung (Korea Atomic Energy Research Institute)
Publication Information
Journal of Radiation Protection and Research / v.35, no.2, 2010 , pp. 77-80 More about this Journal
Abstract
The spectroscopy toolkit has been developed and tested. The toolkit consists of a CsI(Tl)/PIN diode detector, integrated electronics, and a multi.channel.analyzer and its size was 40 cm(width) by 20 cm(length) by 6 cm(high). It is compact, very portable and simpler and cheaper compared to the conventional spectroscopy system. The gamma energy resolutions of the toolkit were 7.9% for the 660 keV of $^{137}Cs$ and 4.9% for 1,332 keV of $^{60}Co$ respectively. The linearity for gamma energies was good. When the energy spectrum of a ceramic sample containing $^{232}Th$ was measured with the spectroscopy toolkit for 20 minutes, there were significant peaks of the heavy metal. These results show that the resolution of the spectroscopy toolkit is sufficient to accumulate a quality spectrum in a few minutes by using weak, encapsulated commercial sources. Furthermore a toolkit experiment that how to measure energy spectra using the toolkit, and how to identify specific isotopes in a pottery piece, could be widely adopted for education and even for more sophisticated and higher level experiments.
Keywords
Spectroscopy toolkit; Nuclear education; Gamma spectrum; $^{137}Cs$; $^{60}Co$;
Citations & Related Records
연도 인용수 순위
  • Reference
1 UCS30 manual, www.spectrumtechniques.com.
2 Anjos RM, Facure A, Lima ELN, Gomes PRS, Santos MS, Brage JAP. Radioactivity teaching : Environmental consequences of the radiological accident in Goiania (Brazil). Am. J. Phys. 2001;69(3): 377‐381.   DOI   ScienceOn
3 Kim HS, Ha JH, Park SH. Cho SY, Kim YK. Fabrication and performance characteristics of a Cs(Tl)/PIN diode radiation detector for industrial applications. Appl. Radiat. Isotopes 2009:67:1463‐ 1465.   DOI   ScienceOn