Verification of Indicator Rotation Correction Function of a Treatment Planning Program for Stereotactic Radiosurgery  

Chung, Hyun-Tai (Department of Neurosurgery, Seoul National University Hospital)
Lee, Re-Na (Department of Radiation Oncology, Ewha Womans University School of Medicine)
Publication Information
Journal of Radiation Protection and Research / v.33, no.2, 2008 , pp. 47-51 More about this Journal
Abstract
Objective: This study analyzed errors due to rotation or tilt of the magnetic resonance (MR) imaging indicator during image acquisition for a stereotactic radiosurgery. The error correction procedure of a commercially available stereotactic neurosurgery treatment planning program has been verified. Materials and Methods: Software virtual phantoms were built with stereotactic images generated by a commercial programming language, Interactive Data Language (version 5.5). The thickness of an image slice was 0.5 mm, pixel size was $0.5{\times}0.5mm$, field of view was 256 mm, and image resolution was $512{\times}512$. The images were generated under the DICOM 3.0 standard in order to be used with Leksell GammaPlan$^{(R)}$. For the verification of the rotation error correction function of Leksell GammaPlan$^{(R)}$, 45 measurement points were arranged in five axial planes. On each axial plane, there were nine measurement points along a square of length 100 mm. The center of the square was located on the z-axis and a measurement point was on the z-axis, too. Five axial planes were placed at z=-50.0, -30.0, 0.0, 30.0, 50.0 mm, respectively. The virtual phantom was rotated by $3^{\circ}$ around one of x, y, and z-axis. It was also rotated by $3^{\circ}$ around two axes of x, y, and z-axis, and rotated by $3^{\circ}$ along all three axes. The errors in the position of rotated measurement points were measured with Leksell GammaPlan$^{(R)}$ and the correction function was verified. Results: The image registration errors of the virtual phantom images was $0.1{\pm}0.1mm$ and it was within the requirement of stereotactic images. The maximum theoretical errors in position of measurement points were 2.6 mm for a rotation around one axis, 3.7 mm for a rotation around two axes, and 4.5 mm for a rotation around three axes. The measured errors in position was $0.1{\pm}0.1mm$ for a rotation around single axis, $0.2{\pm}0.2mm$ for double and triple axes. These small errors verified that the rotation error correction function of Leksell GammaPlan$^{(R)}$ is working fine. Conclusion: A virtual phantom was built to verify software functions of stereotactic neurosurgery treatment planning program. The error correction function of a commercial treatment planning program worked within nominal error range. The virtual phantom of this study can be applied in many other fields to verify various functions of treatment planning programs.
Keywords
Virtual phantom; Stereotactic images; Radiosurgery; Rotation correction;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Spetzqer V, Laborde G, Gilsbach JM. Frameless neuronavigation in modern neurosurgery. Minim Invasive Neurosurg. 1995; 38:163-166   DOI   ScienceOn
2 Schicho K, Figl M, Seemann R, Pretterklieber ML, Birkfellner W, Reichwein A, Wanschitz F, Kainberger F, Bergmann H, Wagner A, Ewers R. Comparison of laser surface scanning and fiducial marker-based registration in frameless stereotaxy. Technical note. J Neurosurg. 2007; 106:704-709   DOI   ScienceOn
3 Moerland MA, Beersma R, Bhagwandien R, Wijrdeman HK, Bakker CJG. Analysis and correction of geometric distortions in 1.5T magnetic resonance images for use in radiotherapy treatment planning. Phys. Med. Biol. 1995;40: 1651-1665   DOI   ScienceOn
4 Chung HT, Kim DG. Images for stereotactic neurosurgery. J Korean Soc. Stereotact and Funct. Neurosurg. 2005;1:1-9
5 Peters TM, Clark JA, Olivier A, Marchand EP, Mawko G, Dieumeqarde M, Dieumegarde M, Muresan LV, Ethier R. Integrated stereotaxic imaging with CT, MR imaging, and digital subtraction angiography. Radiology 1986;1:821-826
6 Guo WY: Application of MR in stereotactic radiosurgery. JMRI. 1998;8:415-420   DOI   ScienceOn
7 Park SW, Han MH, Kim DG, Chung HT, Song IC, Kim HD, Chang KH. Assessment of imaging distortion in magnetic resonance imaging for stereotactic radiosurgery: Through phantom study. J Korean Soc. Magn. Reson. Med. 2000;4:7-13
8 Maciunas, RJ, Galloway RL Jr, Latimer JW. The application accuracy of stereotactic frames. Neurosurgery 1994;35:682- 695   DOI   ScienceOn
9 Baldwin LN, Wachowicz K, Thomas SD, Rivest R, Fallone BG. Characterization, prediction, and correction of geometric distortion of 3T MR images. Med. Phys. 2007;34:388-399   DOI   ScienceOn
10 National Electrical Manufacturers Association: Digital imaging and communications in medicine (DICOM). PS3.1, Rosslyn: NEMA Standards Publication, 1993